Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867101906), received between Jun 21, 2021 and Sep 23, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • Foreign Body In Patient2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Break

    It was reported that during the surgery, the tip of the drill broke off and remained inside the patient´s bone. The broken piece is located under the anchor in the bone. This was discovered during the computer tomography after the surgery. No further information was received.*** Update 15-Sep-2026 - Further information was received:The incident occurred during the shoulder surgery. The broken-off piece is intended to remain in the patient permanently and the broken piece is firmly embedded in the bone because it is secured by the anchor. The surgery was finished successfully without using a different device. It was not necessary to switch the surgical technique. No second surgery was performed, and no revision surgery is planned. The patient underwent one postoperative computer tomography scan of the shoulder.

    Patient problems: Foreign Body In Patient

    MAUDE report 50867635 ↗
  • MalfunctionApr 11, 2023Break; Break

    ON 03/15/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-3600D-1 DRILL BROKE. THIS OCCURRED DURING A LABRAL REPAIR ON (B)(6) 2023, WHEN HE WAS DRILLING INTO THE GLENOID THE END OF THE DRILL BIT BROKE OFF AND WAS STUCK IN THE GLENOID. THE PIECE OF THE DRILL WAS COUNTERSUNK INTO BONE SO IT HAD TO BE LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 16717569 ↗
  • MalfunctionJun 21, 2021Break

    IT WAS REPORTED THAT DURING A SHOULDER INSTABILITY REPAIR, WHEN THE SURGEON INSERTED THE DRILL BIT TO MAKE THE TUNNEL AND ENTER THE ANCHOR AND THE TIP BROKE. THE FRAGMENT WAS RETRIEVED AND THE CASE WAS COMPLETED BY USING A NEW AR-3600D-1.

    Patient problems: Insufficient Information

    MAUDE report 12032479 ↗