Arthrex, Inc. · Revers 39MM CA Adapter Assembly · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867230385), received between Aug 24, 2020 and Dec 11, 2025 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure4
  • Connection Problem1
  • Microbial Contamination of Device1

Patient problems

  • Insufficient Information4
  • No Clinical Signs, Symptoms or Conditions4
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING THE SHOULDER REPLACEMENT SURGERY, THE DEVICE CAN ONLY BE LOCKED INTO THE SUTURECUP WITH GREAT DIFFICULTY AND DOES NOT LOCK PROPERLY. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23776821 ↗
  • MalfunctionNov 5, 2025Positioning Failure; Positioning Failure

    ADDITIONAL INFORMATION WAS RECEIVED ON 11/03/2025: THE PATIENT UNDERWENT A REVISION REVERSE SHOULDER ARTHROPLASTY ON (B)(6) 2025 FOLLOWING A FALL-RELATED INCIDENT. DURING THE PROCEDURE, ATTEMPTS TO IMPLANT TWO COMPONENTS, AR-9502-39ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER 39 AND AR-9556-22RCA UNIVERS REVERS CA HUMERAL HEAD 56/22, WERE UNSUCCESSFUL AS THEY COULD NOT BE SECURED. THIS ISSUE EXTENDED THE SURGICAL TIME BY APPROXIMATELY 30 MINUTES. AS PART OF THE REVISION, ALL IMPLANTS FROM THE ORIGINAL REVERSE SHOULDER ARTHROPLASTY PERFORMED IN (B)(6) 2023 WERE EXPLANTED. THE ORIGINAL SURGERY WAS PERFORMED AT THE SAME FACILITY BY THE SAME SURGEON AND HAD NO REPORTED INTRAOPERATIVE COMPLICATIONS. THE ARTHREX PRODUCTS IMPLANTED IN THE ORIGINAL SURGERY INCLUDE, AR-9503M-03 ARTHREX UNIVERS REVERS HUMERAL INSERT MEDIUM / 39 / +3, AR-9502F-39CPC ARTHREX UNIVERS REVERS SUTURE CUP, 39 (NEUTRAL), AR-9562-28NL UNIVERS REVERS MODULAR GLENOID SYSTEM, PERIPHERAL SCREW, NON-LOCKING 4.5 X 28 MM, AR-9563-20 UNIVERS REVERS MODULAR GLENOID SYSTEM, PERIPHERAL SCREW, LOCKING 5.5 X 20 MM, AR-9563-28 UNIVERS REVERS MODULAR GLENOID SYSTEM, PERIPHERAL SCREW, LOCKING 5.5 X 28 MM, AR-9564-2439-LAT ARTHREX UNI

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23477670 ↗
  • MalfunctionNov 5, 2025Positioning Failure; Positioning Failure

    ADDITIONAL INFORMATION WAS RECEIVED ON 11/03/2025: THE PATIENT UNDERWENT A REVISION REVERSE SHOULDER ARTHROPLASTY ON (B)(6) 2025 FOLLOWING A FALL-RELATED INCIDENT. DURING THE PROCEDURE, ATTEMPTS TO IMPLANT TWO COMPONENTS, AR-9502-39ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER 39 AND AR-9556-22RCA UNIVERS REVERS CA HUMERAL HEAD 56/22, WERE UNSUCCESSFUL AS THEY COULD NOT BE SECURED. THIS ISSUE EXTENDED THE SURGICAL TIME BY APPROXIMATELY 30 MINUTES. AS PART OF THE REVISION, ALL IMPLANTS FROM THE ORIGINAL REVERSE SHOULDER ARTHROPLASTY PERFORMED IN (B)(6) 2023 WERE EXPLANTED. THE ORIGINAL SURGERY WAS PERFORMED AT THE SAME FACILITY BY THE SAME SURGEON AND HAD NO REPORTED INTRAOPERATIVE COMPLICATIONS. THE ARTHREX PRODUCTS IMPLANTED IN THE ORIGINAL SURGERY INCLUDE, AR-9503M-03 ARTHREX UNIVERS REVERS HUMERAL INSERT MEDIUM / 39 / +3, AR-9502F-39CPC ARTHREX UNIVERS REVERS SUTURE CUP, 39 (NEUTRAL), AR-9562-28NL UNIVERS REVERS MODULAR GLENOID SYSTEM, PERIPHERAL SCREW, NON-LOCKING 4.5 X 28 MM, AR-9563-20 UNIVERS REVERS MODULAR GLENOID SYSTEM, PERIPHERAL SCREW, LOCKING 5.5 X 20 MM, AR-9563-28 UNIVERS REVERS MODULAR GLENOID SYSTEM, PERIPHERAL SCREW, LOCKING 5.5 X 28 MM, AR-9564-2439-LAT ARTHREX UNI

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23477353 ↗
  • InjuryMar 6, 2025Positioning Failure; Positioning Failure

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT THE FOLLOWING ARTHREX PARTS (QTY. 2) OF AN AR-9502-42ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER 42, AR-9502-39ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER 39, AR-9556-22RCA UNIVERS REVERS CA HUMERAL HEAD 56/22, AR-9502F-42CPC ARTHREX UNIVERS REVERS SUTURE CUP, 42, AR-9502F-39CPC ARTHREX UNIVERS REVERS SUTURE CUP, 39 (NEUTRAL), AR-9501-10P ARTHREX UNIVERS REVERS HUMERAL STEM, SIZE 10 WERE UNSUCCESSFUL IN SEATING THE 42 ADAPTER INTO THE 42 CUP. A 39 ADAPTER AND A 39 CUP WERE ALSO ATTEMPTED BUT UNSUCCESSFUL IN PROPERLY SEATING. NOTHING BROKE INSIDE THE PATIENT. THE PROCEDURE WAS ABORTED. THIS WAS DISCOVERED DURING A REVISION REVERSE SHOULDER PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21536220 ↗
  • InjuryMay 10, 2024Connection Problem; Connection Problem

    ON 6/23/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT DURING A REVISION PROCEDURE, BOTH THE AR-9502-29ARCA, REVERS 39MM CA ADAPTER ASSEMBLY AND, AR-9550-29RCA, REVERS CA HUMERAL HEAD, 50/19, WERE HAVING ISSUES SEATING. THE SURGEON HAD THEM IN THE SUTURE CUP AND THEN AFTER TRIAL REDUCTION, THEY WERE AUGMENTING THE SUBS CAP AND DID AN INTERNAL ROTATION. THE ADAPTOR AND HEAD, BOTH, POPPED OUT OF THE CUP. THE HEAD BECAME STUCK ON THE POST AND THE ADAPTOR WOULD NOT SIT IN THE CUP. IN ORDER TO BE SAFE IT WAS AGREED THAT A NEW IMPLANT BE USED TO COMPLETE THE CASE. THE SURGEON HAD NO ISSUES AFTER SWAPPING OUT THE IMPLANTS, HOWEVER, IT DID ADD TIME TO THE CASE. THE REVISION WAS A CONVERSION FROM A REVERS TOTAL SHOULDER PROCEDURE TO A HEMI DUE TO THE MGS BEING LOOSE AND NOT HOLDING IN THE GLENOID. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 19290662 ↗
  • InjuryAug 24, 2020Microbial Contamination of Device

    AN ARTHREX EMPLOYEE HAS REPORTED THAT A PATIENT DEVELOPED AN INFECTION IN THE SURGICAL AREA. THE PROSTHESIS WAS EXCHANGED. IT WAS FURTHER REPORTED THAT THE CUSTOMER IS LOOKING FOR GERMS ON THE PROSTHESIS. NO FURTHER INFORMATION RECEIVED. UPDATE 30-JULY-2020: FURTHER INFORMATION WERE PROVIDED THAT THE INITIAL SURGERY WAS PERFORMED IN (B)(6) 2019. THE REPORTED PRODUCT WAS REPLACED DURING THE REVISION. THE INFECTION WAS IDENTIFIED ON (B)(6) AND THE REVISION WAS PERFORMED ON (B)(6) 2020.

    Patient problems: Unspecified Infection

    MAUDE report 10443157 ↗