Skip to content

Arthrex® AR-9501-10PRecalls & reported problems

Arthrex, Inc. · UNIVERS REVERS HUMERAL STEM, SIZE 10 · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867250420), received between Mar 6, 2025 and Apr 17, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1
  • Positioning Failure1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2025Material Separation; Material Separation

    IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT A 36 UNIVERS REVERS CA HUMERAL HEAD ADAPTER AND A CA HUMERAL HEAD DISENGAGED FROM AN AR-9502F-39CPC ARTHREX UNIVERS REVERS SUTURECUP AND HUMERAL STEM. THEN, THE 36 UNIVERS REVERS CA HUMERAL HEAD ADAPTER AND THE CA HUMERAL HEAD COULD NOT BE TAKEN APART. SIZING WAS THOUGHT TO BE AN ISSUE, SO THE SURGEON ATTEMPTED TO ENGAGE A 39 SUTURECUP AND AN AR-9502-39ARCA 39 HUMERAL HEAD ADAPTER ON THE BACK TABLE, BUT THEY KEPT COMING APART. THE CASE WAS COMPLETED USING A 36 HUMERAL HEAD ADAPTER. THIS WAS DISCOVERED DURING AN RTSA PROCEDURE ON (B)(6) 2025. THIS ISSUE ADDED AN ADDITIONAL HOUR AND A HALF TO THE CASE WHICH ALREADY NEEDED A TIME LIMIT DUE TO THE PATIENT'S HEALTH AND WAS AT A HIGHER RISK FOR SURGERY. AT THIS TIME, IT IS UNKNOWN IF ADDITIONAL ANESTHESIA WAS ADMINISTERED. ADDITIONAL INFORMATION WAS RECEIVED ON 3/31/2025: THE SALES REPRESENTATIVE PROVIDED THE PART NUMBER AR-9501-10P FOR THE ARTHREX UNIVERS REVERS HUMERAL STEM, AR-9550-19RCA FOR THE ARTHREX UNIVERS REVERS CA HUMERAL HEAD, AND AR-9556-22RCA UNIVERS REVERS CA HUMERAL HEAD. THE AR-9501-10P FOR THE ARTHREX UNIVERS REVERS HUMERAL STEM WAS NOT EXCHANGED AND WAS THE SAME

    Patient problems: Insufficient Information

    MAUDE report 21861130 ↗
  • InjuryMar 6, 2025Positioning Failure; Positioning Failure

    ON 02/14/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT THE FOLLOWING ARTHREX PARTS (QTY. 2) OF AN AR-9502-42ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER 42, AR-9502-39ARCA UNIVERS REVERS CA HUMERAL HEAD ADAPTER 39, AR-9556-22RCA UNIVERS REVERS CA HUMERAL HEAD 56/22, AR-9502F-42CPC ARTHREX UNIVERS REVERS SUTURE CUP, 42, AR-9502F-39CPC ARTHREX UNIVERS REVERS SUTURE CUP, 39 (NEUTRAL), AR-9501-10P ARTHREX UNIVERS REVERS HUMERAL STEM, SIZE 10 WERE UNSUCCESSFUL IN SEATING THE 42 ADAPTER INTO THE 42 CUP. A 39 ADAPTER AND A 39 CUP WERE ALSO ATTEMPTED BUT UNSUCCESSFUL IN PROPERLY SEATING. NOTHING BROKE INSIDE THE PATIENT. THE PROCEDURE WAS ABORTED. THIS WAS DISCOVERED DURING A REVISION REVERSE SHOULDER PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21536568 ↗

Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices

All PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices →

Sources and updates