Arthrex® AR-3600D-2HRecalls & reported problems

Arthrex, Inc. · 1.8mm DRILL BIT, KNOTLESS HIP FiberTak · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00888867298019), received between Feb 14, 2022 and Oct 23, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Difficult to Remove1

Patient problems

  • Insufficient Information6
  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2024Break; Break

    ON (B)(6)2024, A DISTRIBUTOR VIA (B)(4) REPORTED THAT AN AR-3600D-2H DRILL FOR A HIP FIBERTAK SUTURE ANCHOR BROKE DURING USE. THIS ISSUE OCCURRED WHEN THE SURGEON PUT IN THE FIRST ANCHOR OF A LABRAL REPAIR. THE TRAJECTORY OF THE DRILL & GUIDE WAS GOOD; WHEN THE SURGEON DRILLED TO THE GUIDE'S HARD STOP, THE DRILL TIP BROKE OFF IN THE BONE. THE SURGEON WAS ABLE TO BURR AROUND THE DRILL BIT & USE A CHONDRAL PICK TO CREATE SPACE AROUND THE DRILL BIT. THE SURGEON THEN GRASPED THE END OF THE DRILL BIT WITH A KINGFISHER & REMOVED THE BROKEN PIECE AND CONFIRMED. NO ADVERSE EFFECTS ON THE PATIENT WERE REPORTED, AND NO SECONDARY SURGERY WAS NEEDED. THIS WAS DISCOVERED DURING A HIP ARTHROSCOPY PROCEDURE WITH LABRAL REPAIR ON (B)(6)2024. ADDITIONAL INFORMATION WAS RECEIVED ON 10/08/2024: THERE WAS 8 MINUTES OF CASE DELAY WITH NO ADDED ANESTHESIA ADMINISTERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20513483 ↗
  • MalfunctionNov 29, 2023Break; Break

    ON 11/01/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3600D-2H DRILL BROKE. THIS OCCURRED DURING A HIP LABRAL REPAIR WHERE THE FRAGMENT WAS RETRIEVED AND THE CASE COMPLETED WITH NO FURTHER ISSUE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18229317 ↗
  • MalfunctionJul 3, 2023Break; Break

    ON 6/6/2023, IT WAS REPORTED BY A SALE REPRESENTATIVE VIA EMAIL THAT A HIP 1.8MM DRILL BROKE IN HALF. NO FURTHER INFORMATION HAS BEEN REPORTED. ADDITIONAL INFORMATION RECEIVED ON 6/6/2023: AN AR-3600D-2H 1.8MM DRILL FOR HIP FIBERTAK BROKE DURING REAMING INSIDE THE PATIENT. THIS WAS DISCOVERED DURING A LABRAL REPAIR PROCEDURE ON (B)(6) 2023. ALL THE BROKEN FRAGMENTS WERE RETRIEVED AND A NEW AR-3600D-2H 1.8MM DRILL FOR HIP FIBERTAK WAS USED TO COMPLETE THE CASE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17246525 ↗
  • MalfunctionFeb 13, 2023Break

    ON 01/25/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A AR-3600D-2H DRILL, FOR HIP FIBERTAKS TIP BROKE OFF. THIS OCCURRED DURING A HIP SCOPE. NO ADDITIONAL INFORMATION PROVIDED. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 16361748 ↗
  • MalfunctionFeb 13, 2023Break

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A AR-3600D-2H DRILL, FOR HIP FIBERTAKS TIP BROKE OFF. THIS OCCURRED DURING A HIP SCOPE. NO ADDITIONAL INFORMATION PROVIDED. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16361721 ↗
  • MalfunctionFeb 13, 2023Break

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A AR-3600D-2H DRILL, FOR HIP FIBERTAKS TIP BROKE OFF. THIS OCCURRED DURING A HIP SCOPE. NO ADDITIONAL INFORMATION PROVIDED. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 16361706 ↗
  • MalfunctionFeb 13, 2023Break; Difficult to Remove

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A AR-3600D-2H DRILL, FOR HIP FIBERTAKS TIP BROKE OFF. THIS OCCURRED DURING A HIP SCOPE ON (B)(6) 2023. THE TIP WAS UNABLE TO RETRIEVED. THE CASE CONTINUED USING THE KNOTLESS FIBERTAKS.

    Patient problems: Foreign Body In Patient

    MAUDE report 16361700 ↗
  • MalfunctionFeb 14, 2022Break

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-3600D-2H DRILL BROKE ABOUT 20MM INTO THE PATIENTS ACETABULUM. THIS WAS DISCOVERED DURING A HIP SCOPE PROCEDURE. ADDITIONAL INFORMATION RECEIVED 1/27/2022: THIS WAS DISCOVERED DURING A LABRAL REPAIR PROCEDURE. AFTER THE DRILL BIT BROKE, SURGEON DID NOT REMOVE THE BROKEN FRAGMENTS AND IT STILL REMAINS IN THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 13526263 ↗