Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867298026), received between May 15, 2024 and Jun 10, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Material Fragmentation1
  • Unintended Movement1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT (3) AR-3636H 1.8 KNOTLESS HIP FIBERTAKS, AND (2) AR-3600ND-2H DISPOSABLE DRILLS FAILED. THIS OCCURRED DURING AN ARTHROSCOPIC HIP LABRAL REPAIR ON (B)(6) 2023 WHEN DRILLING FOR THE ANCHORS IT FELT AS IF THE DEVICE WAS CHATTING UP AGAINST THE CURVED GUIDE. THE GUIDE WAS PULLED OUT OF THE HIP AND RAN THE DRILL THROUGH IT WITH THE SAME RESULT. THE DRILL GUIDE WAS SWITCHED OUT AND THE SAME ISSUE OCCURRED. AN ADDITIONAL DRILL BIT WAS TRIED AND THE ISSUE PERSISTED. A THIRD DRILL BIT WITH A DIFFERENT LOT NUMBER AND THE SAME DRILL GUIDE WAS TRIED AND THERE WAS NO CHATTER OF THE DRILL BIT. AFTER THE GUIDE AND DRILL BIT ISSUE WAS RESOLVED AND THEY BEGAN INSERTING THE ANCHORS, 3 OF THE ANCHORS PULLED OUT. THEY ALSO TRIED RELOADING EACH OF THESE ANCHORS AND REINSERTING AND THEY STILL PULLED OUT. IT WASN¿T UNTIL THEY OPENED AN ANCHOR OF A DIFFERENT LOT NUMBER THAT THE ANCHOR HELD IN THE BONE. THE PATIENT HAD AN UNUSUALLY HARD BONE. THERE WAS NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19497624 ↗
  • MalfunctionMay 24, 2024Unintended Movement; Unintended Movement

    ON (B)(6) 2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA SEMS-06550600 THAT AN AR-3600ND-2H DISPOSABLE DRILL TIP SKIPS AFTER DRILLING. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 19393437 ↗
  • MalfunctionMay 23, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 12/16/2021 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT THREE AR-3602H ANCHORS PULLED OUT OF THE BONE DURING A HIP ARTHROSCOPY. THE FIRST ONE PULLED OUT OF THE BONE WHEN IT WAS SET. IT WAS REPLACED AND THE SAME EVENT OCCURRED. A THIRD ANCHOR WAS USED BUT ONCE AGAIN PULLED OUT. THE ANCHORS WERE REPLACED WITH A NON ARTHREX PRODUCTS AND THE CASE WAS COMPLETED WITH NO FURTHER ISSUES. THE CASE INVOLVED A 32 YEAR OLD FEMALE AND, AS OF TODAY SHE IS FINE. THE REP IS ALSO RETURNING THE AR-3600ND-2H DRILL BIT TO BE INSPECTED ALSO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19380435 ↗
  • MalfunctionMay 15, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DRILL WAS SHEARING BITS OFF IN THE REUSABLE SLEEVE. THE SURGEON OPENED A NEW AR-3600ND-2H WITHOUT FURTHER ISSUE. ADDITIONAL INFORMATION PROVIDED: PROCEDURE BEING PERFORMED WAS A LABRAL REPAIR. NO FRAGMENTS BROKE OFF. THE CASE WAS COMPLETED BY USING A DIFFERENT ANCHOR SYSTEM AR-1938BC.

    Patient problems: Insufficient Information

    MAUDE report 19319080 ↗