Skip to content

Arthrex® AR-1588BTB-2JRecalls & reported problems

Arthrex, Inc. · TIGHTROPE ® II BTB WITH DEPLOYING SUTURE · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Arthrex TightRope BTB fixation device has 122 adverse event reports, with no deaths and 2 injuries. Most reports describe sutures breaking or sticking and failing to pass through the device during ligament surgery, and a few describe fixation coming loose after surgery. Reports fell from 25 in the previous year to 11 in the latest year. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
119
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 119 adverse event reports that list this exact device (DI 00888867319660), received between Nov 16, 2021 and Apr 11, 2026 — classified as 1 injury, 118 malfunctions.

The most frequently reported specific device problem is “Break” (39 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 5 reportsJan 2024: 7 reportsFeb 2024: 4 reportsMar 2024: 0 reportsApr 2024: 10 reportsMay 2024: 22 reportsJun 2024: 16 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 6 reportsMay 2025: 4 reportsJun 2025: 6 reportsJul 2025: 0 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break39
  • Failure to Advance21
  • Physical Resistance/Sticking14
  • Material Split, Cut or Torn12
  • Use of Device Problem6
  • Defective Device5
  • Device Dislodged or Dislocated5
  • Material Twisted/Bent4

Patient problems

  • No Clinical Signs, Symptoms or Conditions77
  • Insufficient Information44
  • Foreign Body In Patient3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Failure of Implant1

Recent adverse event reports

The 25 most recent of 119 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026Break

    IT WAS REPORTED THAT DURING THE RECONSTRUCTION OF THE ANTERIOR CRUCIATE LIGAMENT, THE TIGHTROPE SNAPPED WHILE BEING TIGHTENED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24846258 ↗
  • MalfunctionFeb 24, 2026Failure to Advance

    ON 06-FEB-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1588BTB-2J TIGHTROPE II BTB WITH DEPLOYING SUTURE EXPERIENCED AN ISSUE DURING USE. WHEN ATTEMPTING TO PASS THE SUTURE THROUGH THE TIGHTROPE MECHANISM, THE TAPE BALLED UP WITHIN THE BUTTON AND COULD NOT BE ADVANCED AS INTENDED. THE ISSUE OCCURRED IN A PROCEDURE ON (B)(6) 2026, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24429662 ↗
  • MalfunctionFeb 4, 2026Break

    ON 14-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1588BTB-2J TIGHTROPE II BTB SUTURE BROKE DURING TENSIONING AFTER IMPLANTATION. THIS WAS DISCOVERED DURING AN ACL RECONSTRUCTION WITH GRAFTLINK ALLOGRAFT PROCEDURE ON (B)(6) 2026, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 20-JAN-2026: ALL FRAGMENTS WERE RETRIEVED. THE SUTURE FAILED DURING FINAL TENSIONING, AFTER THE GRAFT HAD BEEN CYCLED THROUGH RANGE OF MOTION AND THE SURGEON PROCEEDED TO RE-TENSION FOR FINAL FIXATION. THE SAME PRODUCT PART NUMBER WAS OPENED AND USED TO COMPLETE THE PROCEDURE. THE SURGERY WAS EXTENDED BY APPROXIMATELY 30-45 MINUTES, AND THE PATIENT REQUIRED AN ADDITIONAL 30-45 MINUTES OF ANESTHESIA DUE TO THE DELAY OCCURRING AT THE FINAL STEP OF THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24258343 ↗
  • MalfunctionFeb 2, 2026Material Frayed

    ON 2ND FEBRUARY 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1588BTB-2J, TIGHTROPE® II BTB, WITH DEPLOYING SUTURE, THE WHITE TIGHTROPE SUTURE WAS NOT SHUTTLING THROUGH THE RT BUTTON. THE SUTURES WERE EXTREMELY FRAYED AND MADE PASSING VERY DIFFICULT. THIS WAS DETECTED DURING AN UNSPECIFIED PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS A NEW UNIT WAS OPENED.

    Patient problems: Insufficient Information

    MAUDE report 24232991 ↗
  • MalfunctionDec 11, 2025Break; Break

    ON (B)(6) 2025, A DISTRIBUTOR REPORTED VIA EMAIL THAT THREE TIGHTROPE DEVICES, AR-1588RT-2J TIGHTROPE II RT AND TWO AR-1588BTB-2J TIGHTROPE II BTB, WERE INVOLVED IN ISSUES DURING A HAMSTRING LCA RECONSTRUCTION PROCEDURE ON (B)(6) 2025. AN AR-1588RT-2J TIGHTROPE II RT (LOT 15412617) AND AN AR-1588BTB-2J TIGHTROPE II BTB (LOT 15446881) EXPERIENCED KNOTLESS MECHANISM FAILURE DURING GRAFT INSERTION INTO A 4 MM FEMORAL TUNNEL, PREVENTING COMPLETION OF THE PRIMARY FUNCTION DESPITE FOLLOWING THE RECOMMENDED TECHNIQUE AND WITHOUT EXCESSIVE FORCE. THE THIRD DEVICE, AR-1588BTB-2J TIGHTROPE II BTB (LOT 15462879), EXPERIENCED A SUTURE JAM IN THE KNOTLESS MECHANISM DURING CONVERSION STEP 3 OF THE BTB TIGHTROPE TECHNIQUE. THE STRAIGHT LIMB SUTURE (2) BECAME STUCK, PREVENTING SYSTEM CONVERSION. ADDITIONAL INFORMATION WAS RECEIVED ON 12/01/2025: FOR THE AR-1588RT-2J TIGHTROPE II RT (LOT 15412617) AND THE AR-1588BTB-2J TIGHTROPE II BTB (LOT 15446881), THE FAILURE OCCURRED AFTER THE BUTTON WAS FLIPPED WHILE PULLING THE WHITE SUTURES TO BRING THE GRAFT INTO THE TUNNEL. THE FINAL TENSION HAS NOT YET BEEN ACHIEVED. FOR THE THIRD DEVICE, AR-1588BTB-2J TIGHTROPE II BTB (LOT 15462879), THE STRAIGHT LIMB S

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23777592 ↗
  • MalfunctionDec 11, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON (B)(6) 2025, A DISTRIBUTOR REPORTED VIA EMAIL THAT THREE TIGHTROPE DEVICES, AR-1588RT-2J TIGHTROPE II RT AND TWO AR-1588BTB-2J TIGHTROPE II BTB, WERE INVOLVED IN ISSUES DURING A HAMSTRING LCA RECONSTRUCTION PROCEDURE ON (B)(6) 2025. AN AR-1588RT-2J TIGHTROPE II RT (LOT 15412617) AND AN AR-1588BTB-2J TIGHTROPE II BTB (LOT 15446881) EXPERIENCED KNOTLESS MECHANISM FAILURE DURING GRAFT INSERTION INTO A 4 MM FEMORAL TUNNEL, PREVENTING COMPLETION OF THE PRIMARY FUNCTION DESPITE FOLLOWING THE RECOMMENDED TECHNIQUE AND WITHOUT EXCESSIVE FORCE. THE THIRD DEVICE, AR-1588BTB-2J TIGHTROPE II BTB (LOT 15462879), EXPERIENCED A SUTURE JAM IN THE KNOTLESS MECHANISM DURING CONVERSION STEP 3 OF THE BTB TIGHTROPE TECHNIQUE. THE STRAIGHT LIMB SUTURE (2) BECAME STUCK, PREVENTING SYSTEM CONVERSION. ADDITIONAL INFORMATION WAS RECEIVED ON 12/01/2025: FOR THE AR-1588RT-2J TIGHTROPE II RT (LOT 15412617) AND THE AR-1588BTB-2J TIGHTROPE II BTB (LOT 15446881), THE FAILURE OCCURRED AFTER THE BUTTON WAS FLIPPED WHILE PULLING THE WHITE SUTURES TO BRING THE GRAFT INTO THE TUNNEL. THE FINAL TENSION HAS NOT YET BEEN ACHIEVED. FOR THE THIRD DEVICE, AR-1588BTB-2J TIGHTROPE II BTB (LOT 15462879), THE STRAIGHT LIMB S

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23777500 ↗
  • MalfunctionDec 9, 2025Positioning Failure; Positioning Failure

    ADDITIONAL INFORMATION: THIS ISSUE WAS DISCOVERED DURING AN ACLR PROCEDURE. IT IS UNCLEAR WHETHER THE SUSPENSION MECHANISM INITIALLY LOCKED AND THEN FAILED OR NEVER LOCKED, AS WELL AS WHETHER EXCESSIVE FORCE WAS APPLIED DURING GRAFT ADJUSTMENT OR MILKING MOTION. THERE WERE NO SIGNS OF DEVICE DAMAGE, SUCH AS FRAYED SUTURES. THE SURGERY WAS NOT DELAYED, AND NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED DURING OR FOLLOWING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23758816 ↗
  • MalfunctionNov 25, 2025Positioning Failure; Failure to Advance; Positioning Failure; Failure to Advance

    ON 10/29/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1588BTB-2J TIGHTROPE II BTB WOULDN¿T STAY CINCHED. THE IMPLANT WAS LEFT IN THE PATIENT. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL ON 11/11/2025: THE CASE WAS COMPLETED USING AN AR-4020C-09 FASTTHREAD BIOCOMPOSITE INTERFERENCE SCREW. THE SALES REPRESENTATIVE WAS UNABLE TO CONFIRM WHETHER THE CASE WAS DELAYED OR IF ADDITIONAL ANESTHESIA WAS REQUIRED.

    Patient problems: Foreign Body In Patient

    MAUDE report 23647046 ↗
  • MalfunctionAug 21, 2025Use of Device Problem; Use of Device Problem

    ON 08/05/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-1588BTB-2J CARD CONVERSION FAILED. THIS OCCURRED DURING AN ACL RECONSTRUCTION/ KNEE SCOPE THE BTB TIGHTROPE CARD CONVERSION FAILED. THE SURGEON PULLED THE WIRE LOOP THROUGH AND IT WAS NOT THREADED THROUGH THE METAL BUTTON. THEY OPENED A NEW BTB TIGHTROPE II TO COMPLETE THE CASE.

    Patient problems: Insufficient Information

    MAUDE report 22859894 ↗
  • MalfunctionAug 5, 2025Break; Break

    ON 17TH JULY 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA EMAIL THAT FOR AN AR-1588BTB-2J, TIGHTROPE® II BTB, WITH DEPLOYING SUTURE, THE NITINOL WIRE BROKE WHILE PULLING IN THE SUTURE. THIS WAS DETECTED DURING AN ACL RECON REVISION WITH BTB GRAFT ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, ACL DONE IN PATIENT. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22720582 ↗
  • MalfunctionAug 1, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    ON 14TH JULY 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN, AR-1588BTB-2J, TIGHTROPE® II BTB, WITH DEPLOYING SUTURE, THEY THREADED THE NO. 1 THREAD OF THE TIGHTROPE THROUGH THE BTB GRAFT, THREADED IT THROUGH ITSELF AND THEN THROUGH THE NITINOL SPLICING DEVICE, WHICH BROKE IN HALF WHEN DEPLOYED. THIS CAUSE SOME OF THE TIGHTROPE SUTURE TO SPLICE HALFWAY THROUGH ITSELF, WHICH THEN COULD NOT BE USED AS A TIGHTROPE, CAUSING IT TO BE FAULTY. THIS WAS DETECTED DURING USE IN AN ACL BTB PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS THEY OPENED ANOTHER TIGHTROPE BTB WITHOUT INCIDENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22687697 ↗
  • MalfunctionJun 30, 2025Break; Break

    ON 6/5/2025, IT WAS REPORTED VIA EMAIL BY A SALES REPRESENTATIVE THAT DURING AN ALLOGRAFT ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2025, TWO AR-1588BTB-2J BTB TIGHTROPES FAILED IN THE SAME MANNER. WHILE THE FINAL TENSION WAS BEING APPLIED ON THE TIBIAL SIDE, THE TIGHTROPE BROKE AT THE FINGER TRAP NEAR THE BUTTON. AFTER THE INITIAL FAILURE, A SECOND BTB TIGHTROPE FROM THE SAME LOT WAS USED, BUT IT ALSO FAILED AT THE SAME LOCATION. THE SURGICAL TEAM CONFIRMED THERE WAS NO METAL INTERFERENCE OR SHARP OBJECT PRESENT FROM PRIOR PROCEDURES THAT COULD HAVE CONTRIBUTED TO THE BREAKAGE. TO COMPLETE THE CASE, THE REMAINING SUTURE THROUGH THE GRAFT WAS TIED OVER ONE OF THE BUTTONS AND REINFORCED USING A SECONDARY FIXATION KIT. THE CASE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION RECEIVED ON 6/20/2025: ALL THE BROKEN FRAGMENTS WERE SUCCESSFULLY REMOVED FROM THE PATIENT. THE CASE WAS COMPLETED USING ANOTHER AR-1588BTB-2J BTB TIGHTROPE. THE CASE EXPERIENCED A DELAY OF FORTY-FIVE MINUTES, DURING WHICH ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22357107 ↗
  • MalfunctionJun 30, 2025Break; Break

    ON 6/5/2025, IT WAS REPORTED VIA EMAIL BY A SALES REPRESENTATIVE THAT DURING AN ALLOGRAFT ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2025, TWO AR-1588BTB-2J BTB TIGHTROPES FAILED IN THE SAME MANNER. WHILE THE FINAL TENSION WAS BEING APPLIED ON THE TIBIAL SIDE, THE TIGHTROPE BROKE AT THE FINGER TRAP NEAR THE BUTTON. AFTER THE INITIAL FAILURE, A SECOND BTB TIGHTROPE FROM THE SAME LOT WAS USED, BUT IT ALSO FAILED AT THE SAME LOCATION. THE SURGICAL TEAM CONFIRMED THERE WAS NO METAL INTERFERENCE OR SHARP OBJECT PRESENT FROM PRIOR PROCEDURES THAT COULD HAVE CONTRIBUTED TO THE BREAKAGE. TO COMPLETE THE CASE, THE REMAINING SUTURE THROUGH THE GRAFT WAS TIED OVER ONE OF THE BUTTONS AND REINFORCED USING A SECONDARY FIXATION KIT. THE CASE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION RECEIVED ON 6/20/2025: ALL THE BROKEN FRAGMENTS WERE SUCCESSFULLY REMOVED FROM THE PATIENT. THE CASE WAS COMPLETED USING ANOTHER AR-1588BTB-2J BTB TIGHTROPE. THE CASE EXPERIENCED A DELAY OF FORTY-FIVE MINUTES, DURING WHICH ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22357028 ↗
  • MalfunctionJun 26, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A KNEE SURGERY THE SUTURE TORE WHILE TIGHTENING. THE ENDOBUTTON COULD NOT BE USED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22336548 ↗
  • MalfunctionJun 26, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A KNEE SURGERY THE SUTURE TORE WHILE TIGHTENING. THE ENDOBUTTON COULD NOT BE USED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22336441 ↗
  • MalfunctionJun 4, 2025Break; Break

    ON 15TH MAY 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT 1 UNIT OF AR-1588RT-2J, TIGHTROPE® II RT, WITH DEPLOYING SUTURE AND 1 UNIT OF AR-1588BTB-2J, TIGHTROPE® II BTB, WITH DEPLOYING SUTURE, BROKE INTRA OP ONE AFTER ANOTHER. THIS WAS DETECTED DURING USE IN AN ACL PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING A NEW AR-1588BTB-2J. THERE IS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22143693 ↗
  • MalfunctionJun 4, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A BTB SURGERY THE SUTURE GOT TWISTED AND STUCK IN ITSELF WHEN APPLYING TO THE GRAFT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22143015 ↗
  • MalfunctionMay 28, 2025Break; Break

    IT WAS REPORTED THAT DURING A KNEE SURGERY, THE DEVICE BROKE DURING TIGHTENING. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22096316 ↗
  • MalfunctionMay 22, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION SURGERY SOMETHING SNAPPED ONCE THE GRAFT HAD BEEN PASSED AND BUTTON FLIPPED AND SEATED AND THE TOGGLE OF THE GRAFT WAS BROUGHT FURTHER INTO THE FEMORAL SOCKET. UPON ASSESSMENT, THE BUTTON HAD SNAPPED OFF AND THE TIGHTROPE LOOP WAS JUST ONE STRAND. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 22074936 ↗
  • MalfunctionMay 12, 2025Positioning Failure; Device Dislodged or Dislocated; Positioning Failure; Device Dislodged or Dislocated

    ADDITIONAL INFORMATION WAS RECEIVED ON 05/30/2025: THIS OCCURRED DURING AN ACL PROCEDURE ON (B)(6) 2025. ALL FAILED IMPLANTS WERE REMOVED FROM THE PATIENT. THERE WAS A 10-MINUTE CASE DELAY WITH NO ADDED ANESTHESIA ADMINISTERED. THERE WERE NO ADVERSE EFFECTS REPORTED ON OR AFTER THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22005479 ↗
  • MalfunctionMay 12, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 04/15/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1588BTB-2J TIGHTROPE II BTB ENCOUNTERED AN ISSUE DURING THE INTERNAL PASSING OF THE SUTURE DURING THE COMPLETION STEP FOR #3. THE SUTURE DID NOT WORK, BUNCHED UP, AND LOCKED UP. THE WIRE BROKE, AND THE COMPLETION SUTURE COULD NOT BE SPLICED THROUGH THE TIGHT ROPE DEVICE, MAKING IT UNUSABLE. THIS WAS DISCOVERED DURING A BTB ACL PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22001539 ↗
  • MalfunctionApr 24, 2025Break; Break

    ON 3/27/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO AR-1588BTB-2J TIGHTROPE II BTB NITINOL WIRES BROKE. THIS WAS DISCOVERED DURING AN ACL PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION REQUESTED ON 4/22/2025.

    Patient problems: Insufficient Information

    MAUDE report 21900518 ↗
  • MalfunctionApr 24, 2025Break; Break

    ON 3/27/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO AR-1588BTB-2J TIGHTROPE II BTB NITINOL WIRES BROKE. THIS WAS DISCOVERED DURING AN ACL PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION REQUESTED ON 4/22/2025.

    Patient problems: Insufficient Information

    MAUDE report 21900513 ↗
  • MalfunctionApr 23, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ACL SURGERY THE DEVICE BROKE WHILE PULLING IN THE PREDRILLED CHANNEL. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE PART NUMBER AR-4020C-09. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21890424 ↗
  • MalfunctionApr 23, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ACL SURGERY THE DEVICE BROKE WHILE PULLING IN THE PREDRILLED CHANNEL. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE PART NUMBER AR-4020C-09. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21890351 ↗

FDA clearance and classification

The FDA regulates product code MBI (fastener, fixation, nondegradable, soft tissue) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Arthrex® AR-1588BTB-2J

Has the Arthrex® AR-1588BTB-2J been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867319660), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-1588BTB-2J?

The FDA's MAUDE database lists 119 adverse event reports for this device, received between Nov 16, 2021 and Apr 11, 2026. In those reports, the most common specific device problems are “Break” (39), “Failure to Advance” (21) and “Physical Resistance/Sticking” (14); the most common patient problems are “Foreign Body In Patient” (3) and “Failure of Implant” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-1588BTB-2J?

None of the reports are classified as deaths, 1 as an injury and 118 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Arthrex® AR-1588BTB-2J MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Arthrex® AR-1588BTB-2J, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-1588BTB-2J). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page