Skip to content

Arthrex® AR-1588RT-2JRecalls & reported problems

Arthrex, Inc. · TIGHTROPE ® II RT WITH DEPLOYING SUTURE · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Arthrex TightRope II RT is a fastener used to hold soft tissue in place during surgery. Its 419 reports mostly describe the device breaking or its material splitting or tearing, and some describe it failing to position or advance during a procedure. The reports include no deaths and 8 injuries, and some mention pieces remaining inside the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
418
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 418 adverse event reports that list this exact device (DI 00888867319714), received between Jan 4, 2022 and Sep 17, 2026 — classified as 8 injuries, 410 malfunctions.

The most frequently reported specific device problem is “Break” (191 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 13 reportsJan 2024: 21 reportsFeb 2024: 10 reportsMar 2024: 3 reportsApr 2024: 17 reportsMay 2024: 21 reportsJun 2024: 27 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 6 reportsOct 2024: 6 reportsNov 2024: 4 reportsDec 2024: 5 reportsJan 2025: 11 reportsFeb 2025: 13 reportsMar 2025: 4 reportsApr 2025: 10 reportsMay 2025: 22 reportsJun 2025: 21 reportsJul 2025: 24 reportsAug 2025: 9 reportsSep 2025: 6 reportsOct 2025: 10 reportsNov 2025: 9 reportsDec 2025: 14 reportsJan 2026: 4 reportsFeb 2026: 20 reportsMar 2026: 21 reportsApr 2026: 8 reportsMay 2026: 5 reportsJun 2026: 2 reportsJul 2026: 6 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202627
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break191
  • Material Split, Cut or Torn158
  • Positioning Failure19
  • Device Dislodged or Dislocated13
  • Failure to Advance11
  • Material Frayed9
  • Use of Device Problem9
  • Defective Device8

Patient problems

  • No Clinical Signs, Symptoms or Conditions361
  • Insufficient Information43
  • Foreign Body In Patient25
  • Failure of Implant10
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Device Embedded In Tissue or Plaque1
  • Fibrosis1

Recent adverse event reports

The 25 most recent of 418 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Loosening of Implant Not Related to Bone-Ingrowth

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50801914 ↗
  • MalfunctionSep 8, 2026Break

    It was reported that during a surgery the suture broke during tensioning. Per complaint reporter there was no harm for patient, operator or third party. No further information received.Update 28-Aug-2026:Further information was provided that the issue occurred during an anterior cruciate ligament surgery. No parts broke off inside the patient. The surgery was finished successfully with a different device (interference screw) and different surgical technique to fixate with the screws. It was not necessary to do a second surgery.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50668953 ↗
  • MalfunctionAug 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50495594 ↗
  • MalfunctionJul 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50086979 ↗
  • MalfunctionJul 30, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50082724 ↗
  • MalfunctionJul 27, 2026Material Rupture

    Patient problems: Insufficient Information

    MAUDE report 50017313 ↗
  • MalfunctionJul 27, 2026Use of Device Problem

    Patient problems: Foreign Body In Patient

    MAUDE report 50009279 ↗
  • MalfunctionJul 27, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50009000 ↗
  • MalfunctionJul 27, 2026Break; Material Frayed

    Patient problems: Insufficient Information

    MAUDE report 50008638 ↗
  • MalfunctionJun 22, 2026Break

    IT WAS REPORTED THAT DURING THE RECONSTRUCTION OF THE ANTERIOR CRUCIATE LIGAMENT, THE LOOP OF THE TIGHTROPE FAILED AS THE SURGEON PULLED THE GRAFT INTO THE FEMORAL TUNNEL. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25576649 ↗
  • MalfunctionJun 17, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING THE RECONSTRUCTION OF THE ANTERIOR CRUCIATE LIGAMENT, THE LOOP SNAPPED AS THE SURGEON PULLED THE GRAFT INTO THE FEMORAL TUNNEL. THE DEFECTIVE TIGHTROPE WAS REMOVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25533510 ↗
  • MalfunctionMay 20, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE SUTURE TORE WHILE TIGHTENING THE IMPLANT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25230868 ↗
  • MalfunctionMay 19, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WITH LET THE TIGHTROPE TORE OUT OF THE BUTTON. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25217087 ↗
  • MalfunctionMay 8, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE SUTURES WERE TORN. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25123997 ↗
  • MalfunctionMay 8, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE SUTURES WERE TORN. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25123986 ↗
  • MalfunctionMay 4, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE SUTURE TORE WHILE INSERTING THE DEVICE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25073495 ↗
  • MalfunctionApr 15, 2026Material Rupture; Use of Device Problem

    IT WAS REPORTED THAT DURING THE RECONSTRUCTION OF THE ANTERIOR CRUCIATE LIGAMENT, THE TIGHTROPE LOCKING MECHANISM FAILED AND THE SUTURE RUPTURED WHEN THE TIGHTROPE WAS PULLED. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887636 ↗
  • MalfunctionApr 15, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE SUTURE OF THE DEVICE FAILED. THE FAILURE OCCURRED AT A CRITICAL MOMENT OF THE OPERATION, WHICH EXPOSED THE PATIENT TO AN EXTENDED PROCEDURE TIME AND THE NEED TO USE AN ADDITIONAL IMPLANT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885965 ↗
  • MalfunctionApr 10, 2026Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE FEMORAL TIGHTROPE FIXATION SYSTEM BROKE. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24843380 ↗
  • MalfunctionApr 8, 2026Positioning Failure

    IT WAS REPORTED THAT DURING THE RECONSTRUCTION OF THE ANTERIOR CRUCIATE LIGAMENT, THE SUTURE FAILED WHEN THE GRAFT WAS INSERTED INTO THE FEMORAL CANAL. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE FROM ANOTHER MANUFACTURER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817042 ↗
  • MalfunctionApr 8, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE SUTURE OF THE DEVICE TORE DURING TIGHTENING. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24816211 ↗
  • MalfunctionApr 8, 2026Positioning Failure

    IT WAS REPORTED THAT DURING A KNEE SURGERY WHILE TIGHTENING THE DEVICE AT THE TIBIAL END, THE SUTURE SECURING THE ENDO BUTTON CAME LOOSE. THE SURGEON RE-SUTURED IT, BUT THE ACL GRAFT REMAINED SLIGHTLY LOOSE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24815503 ↗
  • MalfunctionApr 8, 2026Positioning Failure

    IT WAS REPORTED THAT DURING A KNEE SURGERY THE BUTTON BECAME LOOSE WHEN THE SURGEON PULLED IN THE GRAFT. THE SURGEON HAD TO REPEAT THE PROCEDURE, RE-SEW THE GRAFT AND REATTACH THE BUTTON. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24815423 ↗
  • MalfunctionApr 8, 2026Positioning Failure

    IT WAS REPORTED THAT DURING A KNEE SURGERY THE TIGHTROPE BUTTONS CAME LOOSE WHEN PULLING AND RAISING THE GRAFT. THE SURGEON HAD TO REPEAT THE PROCEDURE, RE-SEW THE GRAFT, AND REATTACH A BUTTON. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24814816 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY A BREAKAGE OF THE TIGHTROPE DEVICE OCCURRED DURING TIGHTENING. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24743224 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page