Skip to content

Arthrex® AR-1288QT-100Recalls & reported problems

Arthrex, Inc. · ACL FiberTag® TightRope® Implant System · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867344716), received between Aug 18, 2023 and Aug 27, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Dull, Blunt1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2026Dull, Blunt

    On (b)(6) 2026, it was reported by a sales representative via email that an AR-1288QT-100 ACL FiberTag TightRope was associated with a dull cutting instrument during graft harvest. It was reported that the instrument felt dull when cutting the tendon, resulting in premature cutting of the tendon. The procedure was completed successfully with no pieces breaking off in the patient. This occurred during a Quad ACL Autograft Reconstruction procedure performed on (b)(6) 2026.

    Patient problems: Insufficient Information

    MAUDE report 50474875 ↗
  • MalfunctionOct 2, 2024Break; Break

    ON 09/24/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT THE HARVESTER FROM AN AR-1288QT-100 ACL FIBERTAG TIGHTROPE IMPLANT SYSTEM BROKE WHILE INITIATING THE GRAFT HARVEST. ALL THE BROKEN FRAGMENTS WERE REMOVED, AND THE CASE WAS COMPLETED USING AN AR-2386-10 QUADPRO HARVESTER. THE CASE WAS DELAYED ABOUT THIRTY SECONDS AND NO ADDITIONAL ANESTHESIA WAS NEEDED. THE PATIENT WAS NOT AFFECTED. THIS WAS DISCOVERED DURING A QUAD AUTO ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2024.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20353418 ↗
  • MalfunctionAug 18, 2023Material Separation; Material Separation

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-1288QT-100 ACL FIBERTAGS SUTURE DISASSOCIATED FROM THE NEEDLE. THIS OCCURRED DURING AN ACL RECONSTRUCTION ON (B)(6) 2023, NO FRAGMENTS ENTERED THE PATIENT AND THE WAS A SLIGHT DELAY WHILE SWITCHING TO ANOTHER AR-1288QT-100 TO COMPLETE THE CASE. THERE WAS NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17577716 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page