Arthrex® AR-8991-1Recalls & reported problems

Arthrex, Inc. · DX FiberTak Suture Anchor, #2 MTS w/ NDL · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867377301), received between Sep 29, 2023 and Mar 10, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Device Dislodged or Dislocated1
  • Positioning Failure1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient2
  • Failure of Implant1
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2026Break

    IT WAS REPORTED THAT DURING A SURGERY THE ANCHOR BROKE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION RECEIVED. UPDATE 04-MAR-2026: FURTHER INFORMATION WAS PROVIDED THAT THE REPORTED ISSUE OCCURRED DURING AN ANKLE STABILIZATION OF THE LEFT FIBULA. IT WAS FURTHER CLARIFIED THAT, CONTRARY TO THE INITIAL REPORT, THE ANCHOR HAD NOT FAILED. INSTEAD, IT WAS CONFIRMED THAT THE TIP OF THE DRIVER HAD BROKEN OFF, NOT THE ANCHOR. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24556960 ↗
  • MalfunctionNov 25, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED DURING A RECONSTRUCTION OF THE MEDIAL COLLATERAL LIGAMENT OF THE ANKLE THAT THE ANCHOR DID NOT UNFOLD AFTER BEING INSERTED INTO THE BONE TUNNEL AND WAS PULLED OUT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO DO A SECOND SURGERY BUT THE SURGEON USED A DIFFERENT SURGICAL TECHNIQUE, IN WHICH HE USED A TRANSOSSEOUS TUNNEL TO INSERT VICRYL SUTURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647045 ↗
  • MalfunctionNov 20, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 30TH OCTOBER 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8991-1 DX FIBERTAK SUTURE ANCHOR, #2 MTS W/ NDL WAS REPORTED TO HAVE BEEN PULLED OUT WITH MINIMAL FORCE ON TENSIONING, DESPITE THE IMPLANTS BEING INSERTED CORRECTLY. THIS WAS DETECTED DURING ARTHROBROSTRUM PROCEDURE ON (B)(6) 2024. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT HARM BY USING ADDITIONAL IMPLANTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20735279 ↗
  • MalfunctionSep 25, 2024Break; Use of Device Problem; Break; Use of Device Problem

    ON 18 SEPTEMBER 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8991-1 DX FIBERTAK SUTURE ANCHOR, #2 MTS W/ NDL WAS REPORTED TO HAVE FAILED DURING USE. THIS OCCURRED ON 03 SEPTEMBER 2024 IN CASTLECRAG PRIVATE HOSPITAL DURING AN ARTHROBROSTRUM. THE CASE WAS COMPLETED SUCCESSFULLY USING ADDITIONAL ANCHORS. THERE WAS CASE INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 20300256 ↗
  • MalfunctionFeb 16, 2024Break; Break

    IT WAS REPORTED THAT DURING AN ANKLE INSTABILITY ARTHROSCOPY, AFTER ANCHOR INSERTION AND AFTER PASSING THE BLUE SUTURE THROUGH THE TISSUE THE SURGEON PASSED THE WORKING BLUE SUTURE THROUGH THE BLACK AND WHITE FIBERLINK LOOP AND DOUBLE IT OVER AT THE SECOND PURPLE MARKING. AFTER PULLING THE BLACK AND WHITE SUTURE OUT OF THE PATIENT THE END OF THE BLUE FIBERLINK SUTURE BROKE AND SURGEON COULDN¿T TENSION THE SUTURE OR EVEN TAKE THE ANCHOR OUT OF THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 18722249 ↗
  • MalfunctionJan 4, 2024Break; Break

    IT WAS REPORTED THAT DURING AN INSTABILITY REPAIR SURGERY AFTER THE ANCHOR WAS INTRODUCED, A SMALL PART OF METAL FROM THE ANCHOR DRIVER BROKE OFF INSIDE THE PATIENT. THE BROKEN PIECE REMAINS INSIDE THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 18443992 ↗
  • MalfunctionSep 29, 2023Break; Break

    IT WAS REPORTED THAT DURING A BROSTROM SURGERY THE TAPES SNAPPED OFF. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 17845160 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →