Skip to content

Arthrex® AR-1288RTIB-FC3Recalls & reported problems

Arthrex, Inc. · RT IB TightRope w FlipCutter III Drill · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867379978), received between Nov 7, 2023 and Oct 27, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Defective Device1
  • Detachment of Device or Device Component1
  • Material Twisted/Bent1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2025Break; Break

    ON 10/14/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT THE FLIPCUTTER III FROM AN (B)(6) TIGHTROPE II RT WITH FIBERTAPE IB, FLIPCUTTER III, AND FIBERSTICK IMPLANT SYSTEM BROKE AND WAS UNABLE TO BE USED. THIS WAS DISCOVERED DURING A PROCEDURE ON AR-1288RTIB-FC3 2025. ADDITIONAL INFORMATION WAS RECEIVED ON 10/20/2025: DURING THE PROCEDURE, THE FLIPCUTTER BROKE WHILE RETRO DRILLING INSIDE THE PATIENT. FORTUNATELY, ALL FRAGMENTS WERE SUCCESSFULLY RETRIEVED WITHOUT COMPLICATION. TO COMPLETE THE CASE, AN INDIVIDUAL AR-1240FF FLIPCUTTER III WAS USED. THE INCIDENT DID NOT RESULT IN ANY SURGICAL DELAY, AND NO ADDITIONAL ANESTHESIA WAS REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 23393366 ↗
  • MalfunctionSep 19, 2025Break; Break

    ON 08/29/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT THE BLACK GEAR HOUSING OF A FLIPCUTTER III, PART OF THE AR-1288RTIB-FC3 TIGHTROPE II RT-IB IMPLANT SYSTEM, BROKE DURING BACK-REAMING. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 09/08/2025: DURING AN ACL RECONSTRUCTION PROCEDURE ON 08/29/2025, THE BLACK BULBOUS COMPONENT OF A FLIPCUTTER III, PART OF THE AR-1288RTIB-FC3 TIGHTROPE II RT-IB IMPLANT SYSTEM, SPLIT IN HALF. THIS DID NOT OCCUR OVER AN OPEN INCISION, AND NO FRAGMENTS WERE REPORTED TO HAVE ENTERED THE PATIENT. THE BROKEN PIECE WAS REMOVED FROM THE SURGICAL FIELD. THE CASE WAS COMPLETED USING A NEW AR-1204FF FLIPCUTTER III DRILL. THERE WAS A BRIEF CASE DELAY OF APPROXIMATELY TWO MINUTES, AND IT IS UNCLEAR WHETHER ADDITIONAL ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23096800 ↗
  • MalfunctionJan 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-1288RTIB-FC3 FLIPCUTTER DRILL SEPERATED, SURGEON USED ANOTHER DRILL TO COMPLETE PROCEDURE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18477304 ↗
  • MalfunctionJan 8, 2024Material Twisted/Bent; Material Twisted/Bent

    ON 12/11/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-1288RTIB-FC3 IMPLANT SYSTEM TIGHTROPE® II RT HAD SOME KIND OF TWIST IN THE WHITE SUTURE THAT COULD NOT ALLOW THE SUTURE TO PASS THROUGH THE BUTTON TO BRING THE GRAFT INTO THE SOCKET. THIS OCCURRED DURING AN ACL LIGAMENT RECONSTRUCTION ON (B)(6) 2023 WHERE THE CASE WAS COMPLETED BY REMOVING THE DEVICE AND PLACING THE GRAFT ON ANOTHER TIGHTROPE.

    Patient problems: Insufficient Information

    MAUDE report 18464212 ↗
  • MalfunctionNov 7, 2023Defective Device; Defective Device

    ON (B)(6) 2023 IT WAS REPORTED BY A SALES REP, VIA PHONE, THAT DURING AN ACL RECONSTRUCTION PROCEDURE, THE BLADE ON THE FLIPCUTTER 3 (THAT IS A COMPONENT OF THE AR-1288RTIB-FC3. RT IB TIGHTROPE W FLIPCUTTER III DRILL), DID NOT DEPLOY. THE REP WAS ABLE TO QUICKLY REPLACE THE FLIPCUTTER WITH ANOTHER AND THE CASE WAS COMPLETED WITHOUT FURTHER INCIDENT.

    Patient problems: Insufficient Information

    MAUDE report 18086922 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →

Sources and updates