Arthrex® AR-3740SPRecalls & reported problems

Arthrex, Inc. · Double Loaded Knotless FiberTak, Knee · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Arthrex Double Loaded Knotless FiberTak knee anchor is a soft-tissue fixation implant. Reports describe anchors breaking or dislodging during procedures, sometimes leaving small fragments behind; 22 mention foreign body. There are 93 adverse event reports, 2 injuries, 91 malfunctions, no deaths, and no recalls name this device identifier. Last 12 months had 36 reports, compared with 26 the prior year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
91
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 91 adverse event reports that list this exact device (DI 00888867428041), received between Sep 11, 2023 and Jul 31, 2026 — classified as 2 injuries, 89 malfunctions.

The most frequently reported specific device problem is “Break” (39 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 7 reportsMar 2024: 5 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 4 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 5 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 6 reportsJan 2026: 5 reportsFeb 2026: 14 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break39
  • Device Dislodged or Dislocated21
  • Use of Device Problem6
  • Positioning Failure5
  • Failure to Advance4
  • Material Split, Cut or Torn3
  • Material Twisted/Bent3
  • Patient Device Interaction Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • Insufficient Information26
  • Foreign Body In Patient22
  • Failure of Implant2
  • Hypersensitivity/Allergic reaction1
  • Pain1

Recent adverse event reports

The 25 most recent of 91 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2026Break

    Patient problems: Foreign Body In Patient

    MAUDE report 50105285 ↗
  • MalfunctionJul 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50084720 ↗
  • MalfunctionJul 27, 2026Break

    It was reported that during a lateral extra-articular tenodesis surgery the tip of the inserter broke off due to an issue with trajector despite guidance. The failure was noticed by the Arthrex employee who was present during the surgery. This failure was forwarded to the surgeon, but the surgeon decided to finish the surgery without retrieving the tip. The surgery was finished successfully with the same device. It was not necessary to switch the surgical technique or do a second surgery.\\Update 16- Jul-2026:\Further information was received on 15-Jul-2026 that the tip was not retrieved as it was only a small piece. No revision surgery is planned.

    Patient problems: Foreign Body In Patient

    MAUDE report 50017823 ↗
  • MalfunctionMay 12, 2026Break

    ON 24TH APR 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT AR-3740SP DOUBLE LOADED KNOTLESS FIBERTAK, KNEE BATCH 15437960 BROKE DURING INSERTION. THIS WAS DETECTED DURING LATERAL EXTRA-ARTICULAR TENODESIS PROCEDURE ON THE SAME DAY. A BROKEN FRAGMENT OF APPROXIMATELY 1 MM FROM THE IMPLANT HOLDER FORK REMAINED IN THE PATIENT DESPITE ATTEMPTS TO REMOVE IT USING AVAILABLE INSTRUMENTS. ADDITIONAL INFORMATION RECEIVED ON 5TH MAY 2026 INDICATES THAT AN AR-3710 DRILL SOCKET, BATCH NUMBER 15549638, WAS ALSO USED BY THE SURGEON DURING THIS SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 25148455 ↗
  • MalfunctionMar 12, 2026Material Frayed; Device Dislodged or Dislocated

    ON 16-FEB-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHOR EXPERIENCED AN ISSUE DURING USE. WHILE PASSING THE ANCHOR THROUGH THE GUIDE INSIDE THE PATIENT, THE ANCHOR WAS PULLED OUT, AND THE SUTURES WERE REATTACHED TO THE ANCHOR. WHEN THE DEVICE WAS REINSERTED THROUGH THE GUIDE, THE LOOP END AT THE BOTTOM OF THE ANCHOR WAS FOUND TO BE FRAYED AND DAMAGED. ALL COMPONENTS WERE REMOVED AND REPLACED WITH ANOTHER DEVICE OF THE SAME PART NUMBER. THE EVENT RESULTED IN LESS THAN A FIVE-MINUTE CASE DELAY, WITH UNKNOWN ADDITIONAL ANESTHESIA TIME. THIS ISSUE OCCURRED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2026, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24580413 ↗
  • MalfunctionMar 5, 2026Material Twisted/Bent

    ON 12-FEB-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT TWO AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHORS EXPERIENCED ISSUES DURING A LET PROCEDURE PERFORMED ON (B)(6) 2026. ONE ANCHOR BROKE AT THE SELF-PUNCHING PORTION AT THE TIP OF THE INSERTER, AND THE BROKEN PIECES WERE NOT RETRIEVED. A SECOND ANCHOR WAS FOUND TO BE BENT. ALTHOUGH THE LOT NUMBERS FOR THE TWO ANCHORS ARE KNOWN, IT IS UNKNOWN WHICH LOT CORRESPONDS TO EACH AFFECTED DEVICE. A REPLACEMENT AR-3740SP ANCHOR WAS USED TO COMPLETE THE PROCEDURE. NO SIGNIFICANT PROCEDURAL DELAY OCCURRED, AND NO ADDITIONAL ANESTHESIA WAS REQUIRED. ADDITIONAL INFORMATION WAS RECEIVED ON 18-FEB-2026: THE SECOND ANCHOR WAS OBSERVED BENDING ON THE INSERTER. THERE WAS NO CONFIRMATION REGARDING WHICH LOT NUMBER THE RETAINED FRAGMENT ORIGINATED FROM. ALL THREE ANCHORS REMAIN IMPLANTED IN THE PATIENT.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 24519061 ↗
  • MalfunctionMar 5, 2026Break

    ON 12-FEB-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT TWO AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHORS EXPERIENCED ISSUES DURING A LET PROCEDURE PERFORMED ON (B)(6) 2026. ONE ANCHOR BROKE AT THE SELF-PUNCHING PORTION AT THE TIP OF THE INSERTER, AND THE BROKEN PIECES WERE NOT RETRIEVED. A SECOND ANCHOR WAS FOUND TO BE BENT. ALTHOUGH THE LOT NUMBERS FOR THE TWO ANCHORS ARE KNOWN, IT IS UNKNOWN WHICH LOT CORRESPONDS TO EACH AFFECTED DEVICE. A REPLACEMENT AR-3740SP ANCHOR WAS USED TO COMPLETE THE PROCEDURE. NO SIGNIFICANT PROCEDURAL DELAY OCCURRED, AND NO ADDITIONAL ANESTHESIA WAS REQUIRED. ADDITIONAL INFORMATION WAS RECEIVED ON 18-FEB-2026: THE SECOND ANCHOR WAS OBSERVED BENDING ON THE INSERTER. THERE WAS NO CONFIRMATION REGARDING WHICH LOT NUMBER THE RETAINED FRAGMENT ORIGINATED FROM. ALL THREE ANCHORS REMAIN IMPLANTED IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24519059 ↗
  • MalfunctionMar 3, 2026Positioning Failure

    ON 11-FEB-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHOR WAS FOUND TO CONTAIN ONLY ONE FIXED PRECONSTRUCTED KNOTLESS LOOP. THE BLUE SUTURE WAS NOT SECURED WITHIN THE ANCHOR AND WAS FREELY SLIDING THROUGH THE MECHANISM. THE ISSUE WAS IDENTIFIED DURING A LET PROCEDURE ON (B)(6) 2026. NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 17-FEB-2026: THE ISSUE WAS DISCOVERED DURING IMPLANTATION. THE CASE WAS COMPLETED USING AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK. THE SURGEON ELECTED TO CONTINUE USING THE IMPLANTED ANCHOR AND TIED IN THE SUTURE RATHER THAN UTILIZING THE KNOTLESS MECHANISM. NO SURGICAL DELAY OCCURRED, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24497307 ↗
  • MalfunctionFeb 23, 2026Break

    ON 05-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA EMAIL THAT AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK EXPERIENCED AN ISSUE DURING USE. THE FORK-LIKE TIP BROKE UPON FIRST CONTACT WITH THE PATIENT¿S BONE. THIS ISSUE WAS DISCOVERED DURING A LET PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 11-FEB-2025: ALL FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED WITH AN AR-2324KBCC BIOCOMPOSITE KNOTLESS SWIVELOCK ANCHOR. THE CASE WAS DELAYED FIVE MINUTES, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Insufficient Information

    MAUDE report 24417555 ↗
  • MalfunctionFeb 18, 2026Positioning Failure

    ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT (2) AR-3770SP HYBRID KNEE FIBERTAK® ANCHORS AND (2) AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK® ANCHORS BACKED OUT. THIS OCCURRED DURING AN MPFL RECONSTRUCTION ON (B)(6) 2026 WHEN ALL (4) DEVICES BACKED OUT OF THE PATIENT. THE PATIENT HAD VERY GOOD BONE QUALITY, WHICH WAS PREPPED USING AN AR-3712-29 2.9MM DRILL FOR KNEE FIBERTAKS. THESE FAILURES RESULTED IN A 15-MINUTE DELAY WHICH NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE PROCEDURE WAS COMPLETED USING AN AR-2324KBCC BIOCOMPOSITE KNOTLESS SWIVELOCK® ANCHOR, AND AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK® ANCHOR. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24376533 ↗
  • MalfunctionFeb 18, 2026Positioning Failure

    ON 30-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT (2) AR-3770SP HYBRID KNEE FIBERTAK® ANCHORS AND (2) AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK® ANCHORS BACKED OUT. THIS OCCURRED DURING AN MPFL RECONSTRUCTION ON 30-JAN-2026 WHEN ALL (4) DEVICES BACKED OUT OF THE PATIENT. THE PATIENT HAD VERY GOOD BONE QUALITY, WHICH WAS PREPPED USING AN AR-3712-29 2.9MM DRILL FOR KNEE FIBERTAKS. THESE FAILURES RESULTED IN A 15-MINUTE DELAY WHICH NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE PROCEDURE WAS COMPLETED USING AN AR-2324KBCC BIOCOMPOSITE KNOTLESS SWIVELOCK® ANCHOR, AND AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK® ANCHOR. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24376528 ↗
  • MalfunctionFeb 16, 2026Break

    IT WAS REPORTED THAT DURING A LATERAL EXTRA-ARTICULAR TENODESIS SURGERY IN THE KNEE THE TIP OF THE INSERTER BROKE OFF. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (SWIVELOCK). IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24345406 ↗
  • MalfunctionFeb 10, 2026Break

    ON (B)(6) 2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT THREE AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHORS WERE INVOLVED IN AN INTRAOPERATIVE ISSUE DURING AN ACL RECONSTRUCTION WITH LET PROCEDURE ON (B)(6) 2026. DURING THE INSERTION OF THE FIRST ANCHOR, THE INSERTER BROKE, PREVENTING SUCCESSFUL PLACEMENT. A SECOND ANCHOR WAS THEN ATTEMPTED; HOWEVER, IT DID NOT SECURE WITHIN THE BONE TUNNEL AND SUBSEQUENTLY FELL OUT. A THIRD ANCHOR WAS OPENED FOR USE, BUT ITS INSERTER ALSO BROKE DURING INSERTION. DESPITE THIS, THE THIRD ANCHOR WAS SUCCESSFULLY IMPLANTED, AND THE SURGERY WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. THE LOCATION OF THE BROKEN FRAGMENT OF THE INSERTER REMAINS UNKNOWN. ADDITIONAL INFORMATION WAS RECEIVED ON 26-JAN-2026: THE THREE AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHORS WERE PLACED IN THE PATELLA SOCKET. FOR THE FIRST AND THIRD ANCHOR COMPLAINTS, THE INSERTER TIPS BROKE DURING USE, AND THE BROKEN FRAGMENTS WERE RETRIEVED FROM THE PATIENT IN BOTH INSTANCES. THE PROCEDURE EXPERIENCED A BRIEF DELAY OF APPROXIMATELY 15 MINUTES. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24308442 ↗
  • MalfunctionFeb 10, 2026Break

    ON 20-JAN-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT THREE AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHORS WERE INVOLVED IN AN INTRAOPERATIVE ISSUE DURING AN ACL RECONSTRUCTION WITH LET PROCEDURE ON (B)(6) 2026. DURING THE INSERTION OF THE FIRST ANCHOR, THE INSERTER BROKE, PREVENTING SUCCESSFUL PLACEMENT. A SECOND ANCHOR WAS THEN ATTEMPTED; HOWEVER, IT DID NOT SECURE WITHIN THE BONE TUNNEL AND SUBSEQUENTLY FELL OUT. A THIRD ANCHOR WAS OPENED FOR USE, BUT ITS INSERTER ALSO BROKE DURING INSERTION. DESPITE THIS, THE THIRD ANCHOR WAS SUCCESSFULLY IMPLANTED, AND THE SURGERY WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. THE LOCATION OF THE BROKEN FRAGMENT OF THE INSERTER REMAINS UNKNOWN. ADDITIONAL INFORMATION WAS RECEIVED ON 26-JAN-2026: THE THREE AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHORS WERE PLACED IN THE PATELLA SOCKET. FOR THE FIRST AND THIRD ANCHOR COMPLAINTS, THE INSERTER TIPS BROKE DURING USE, AND THE BROKEN FRAGMENTS WERE RETRIEVED FROM THE PATIENT IN BOTH INSTANCES. THE PROCEDURE EXPERIENCED A BRIEF DELAY OF APPROXIMATELY 15 MINUTES. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24308438 ↗
  • MalfunctionFeb 10, 2026Failure to Advance

    ON 20-JAN-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT THREE AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHORS WERE INVOLVED IN AN INTRAOPERATIVE ISSUE DURING AN ACL RECONSTRUCTION WITH LET PROCEDURE ON (B)(6) 2026. DURING THE INSERTION OF THE FIRST ANCHOR, THE INSERTER BROKE, PREVENTING SUCCESSFUL PLACEMENT. A SECOND ANCHOR WAS THEN ATTEMPTED; HOWEVER, IT DID NOT SECURE WITHIN THE BONE TUNNEL AND SUBSEQUENTLY FELL OUT. A THIRD ANCHOR WAS OPENED FOR USE, BUT ITS INSERTER ALSO BROKE DURING INSERTION. DESPITE THIS, THE THIRD ANCHOR WAS SUCCESSFULLY IMPLANTED, AND THE SURGERY WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. THE LOCATION OF THE BROKEN FRAGMENT OF THE INSERTER REMAINS UNKNOWN. ADDITIONAL INFORMATION WAS RECEIVED ON 26-JAN-2026: THE THREE AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHORS WERE PLACED IN THE PATELLA SOCKET. FOR THE FIRST AND THIRD ANCHOR COMPLAINTS, THE INSERTER TIPS BROKE DURING USE, AND THE BROKEN FRAGMENTS WERE RETRIEVED FROM THE PATIENT IN BOTH INSTANCES. THE PROCEDURE EXPERIENCED A BRIEF DELAY OF APPROXIMATELY 15 MINUTES. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24308437 ↗
  • MalfunctionFeb 10, 2026Break

    ON 22-JAN-2026, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT WHEN THE SUTURES OF AN AR-3740SP SELF-PUNCHING 2.6 MM DOUBLE LOADED KNOTLESS KNEE FIBERTAK WERE PULLED, THE ANCHOR BROKE. THIS WAS DISCOVERED DURING USE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24306436 ↗
  • MalfunctionFeb 10, 2026Break

    ON 14-JAN-2026, A MONTHLY MAUDE DATABASE REVIEW IDENTIFIED A REPORT INVOLVING AN AR-3740SP KNEE DOUBLE LOADED KNOTLESS FIBERTAK ANCHOR THAT HAD NOT BEEN PREVIOUSLY REPORTED TO ARTHREX. ACCORDING TO THE MAUDE NARRATIVE, DURING SURGERY, THE SURGEON INSERTED THE SELF-PUNCHING ANCHOR USING THE INSERTER. UPON REMOVAL OF THE INSERTER FROM THE KNEE, THE TIP WAS NOTED TO BE MISSING. RADIOGRAPHS IDENTIFIED A SMALL FRAGMENT OF THE INSERTER, APPROXIMATELY 1¿2 MM IN SIZE, RETAINED IN THE PATIENT. THE SURGEON EXPLAINED THAT ATTEMPTING TO RETRIEVE THE RETAINED FOREIGN BODY COULD CAUSE ADDITIONAL DAMAGE AND THEREFORE RECOMMENDED LEAVING IT IN PLACE, AS IT IS NOT EXPECTED TO HARM THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 29-JAN-2026: THIS WAS DISCOVERED DURING A LEFT KNEE REVISION ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION AND LATERAL EXTRA-ARTICULAR TENODESIS. THE FIBERTAK ANCHOR WAS SUCCESSFULLY IMPLANTED, AND NO SURGICAL DELAYS WERE REPORTED. THE PROCEDURE WAS PERFORMED UNDER GENERAL ANESTHESIA WITH THE ADDITION OF A NERVE BLOCK. ADDITIONAL INFORMATION WAS RECEIVED ON 5-FEB-2026: ACCORDING TO THE FACILITY REPRESENTATIVE, THE AR-3740SP KNEE DOUBLE LOADED KNOTLESS FIBERTAK ANCHOR BROKE D

    Patient problems: Foreign Body In Patient

    MAUDE report 24304494 ↗
  • MalfunctionFeb 6, 2026Break

    IT WAS REPORTED THAT DURING A LATERAL EXTRA-ARTICULAR TENODESIS SURGERY IN THE KNEE THE TIP OF THE INSERTER BROKE OFF ON INSERTION. THE BROKEN PART REMAINED IN THE PATIENT, SINCE IT WAS SECURELY EMBEDDED IN THE BONE. THE ANCHOR OF THE INSERTER WAS STILL DEPLOYED AND WAS USED SUCCESSFULLY TO FINISH THE SURGERY. THE SURGEON ALSO NOTED THAT THE PATIENT WASN'T STABLE AS HE MALLETED THE ANCHOR IN, THE KNEE MOVED WHICH PUT THE FORK TIPS UNDER EXCESSIVE FORCE RESULTING IN BREAKAGE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 24280809 ↗
  • MalfunctionFeb 6, 2026Device Dislodged or Dislocated

    ON 16-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHORS PULLED OUT DURING A LET KNEE PROCEDURE PERFORMED ON (B)(6) 2026. THE SURGICAL TEAM RELOCATED THE FIXATION SITES AND SUCCESSFULLY IMPLANTED AN ADDITIONAL AR-3740SP ANCHOR, WHICH HELD AS INTENDED. NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED ON 20-JAN-2026. ADDITIONAL INFORMATION WAS RECEIVED ON 22-JAN-2026: THE PROCEDURE WAS DELAYED APPROXIMATELY 5¿10 MINUTES WHILE ATTEMPTS WERE MADE TO PLACE ADDITIONAL ANCHORS. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24278355 ↗
  • MalfunctionFeb 6, 2026Device Dislodged or Dislocated

    ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHORS PULLED OUT DURING A LET KNEE PROCEDURE PERFORMED ON (B)(6) 2026. THE SURGICAL TEAM RELOCATED THE FIXATION SITES AND SUCCESSFULLY IMPLANTED AN ADDITIONAL AR-3740SP ANCHOR, WHICH HELD AS INTENDED. NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED ON 20-JAN-2026. ADDITIONAL INFORMATION WAS RECEIVED ON 22-JAN-2026: THE PROCEDURE WAS DELAYED APPROXIMATELY 5¿10 MINUTES WHILE ATTEMPTS WERE MADE TO PLACE ADDITIONAL ANCHORS. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24278145 ↗
  • MalfunctionFeb 4, 2026Break

    ON 14-JAN-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHOR WAS INSERTED, AND THE TIP OF THE INSERTER BROKE OFF AND REMAINED INSIDE THE BONE. THE CASE WAS COMPLETED WITH THE PRODUCT IN PLACE. THERE WAS NO SIGNIFICANT PROCEDURAL DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE ISSUE WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2026.

    Patient problems: Foreign Body In Patient

    MAUDE report 24257218 ↗
  • MalfunctionFeb 4, 2026Device Dislodged or Dislocated

    ON (B)(6) 2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHOR PULLED OUT WHILE TENSION WAS BEING APPLIED TO THE LOOP AFTER THE ANCHOR WAS PLACED INTO BONE. THE ANCHOR WAS REMOVED, AND THE PROCEDURE WAS COMPLETED USING A REPLACEMENT AR-3740SP ANCHOR. THERE WAS NO SIGNIFICANT PROCEDURAL DELAY, NO ADDITIONAL ANESTHESIA WAS REQUIRED, AND NO ADVERSE EFFECTS WERE REPORTED. THE ISSUE OCCURRED DURING A PROCEDURE ON (B)(6) 2026, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24257190 ↗
  • MalfunctionJan 13, 2026Break

    ON 23-DEC-2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT THE ANCHOR SHAFT TIP BROKE ON AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHOR AFTER INSERTION INTO THE PILOT HOLE. THE BROKEN FRAGMENT COULD NOT BE RETRIEVED. THE ANCHOR REMAINED FUNCTIONAL AND WAS USED TO COMPLETE THE PROCEDURE WITHOUT ADDITIONAL ISSUES. NO SIGNIFICANT SURGICAL DELAY OCCURRED, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE EVENT WAS IDENTIFIED DURING THE PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION WAS RECEIVED ON 05-JAN-2026: THIS ISSUE OCCURRED DURING A LATERAL EXTRA-ARTICULAR TENODESIS.

    Patient problems: Foreign Body In Patient

    MAUDE report 24051826 ↗
  • MalfunctionJan 12, 2026Detachment of Device or Device Component; Device Dislodged or Dislocated

    ON 26-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-3740SP DOUBLE-LOADED KNOTLESS FIBERTAK ANCHOR WAS INSERTED INTO THE LATERAL FEMORAL CONDYLE. THE TUNNEL WAS DRILLED, AND THE ANCHOR WAS INSERTED IN STANDARD FASHION. UPON REMOVAL OF THE INSERTER, ONE OF THE SELF-PUNCHING METAL PRONGS WAS MISSING AND REMAINED IN THE LATERAL FEMORAL CONDYLE. THE SURGEON ATTEMPTED TO SET THE ANCHOR, WHICH SUBSEQUENTLY PULLED OUT, WHILE THE METAL PRONG REMAINED IN THE BONE. THE SURGEON ELECTED TO DRILL A NEW TUNNEL, AND A SECOND AR-3740SP WAS OPENED AND IMPLANTED WITHOUT ISSUE. THIS WAS DISCOVERED DURING A LET PROCEDURE ON (B)(6) 2025. THE CASE WAS COMPLETED SUCCESSFULLY, AND ALL BROKEN FRAGMENTS WERE REMOVED. ADDITIONAL INFORMATION WAS RECEIVED ON 30-DEC-2025: THE METAL PRONG WAS NOT REMOVED FROM THE PATIENT. AN AR-3740SP DOUBLE-LOADED KNOTLESS KNEE FIBERTAK ANCHOR WAS USED TO COMPLETE THE CASE. THE CASE WAS DELAYED FIVE MINUTES. IT IS UNKNOWN IF ADDITIONAL ANESTHESIA WAS ADMINISTERED. THE BONE QUALITY WAS ASSESSED AS NORMAL.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 24040054 ↗
  • MalfunctionJan 12, 2026Break

    ON 26-DEC-2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-3740SP DOUBLE LOADED KNOTLESS KNEE FIBERTAK ANCHOR EXPERIENCED AN ISSUE DURING A PROCEDURE PERFORMED ON (B)(6) 2025. AFTER INSERTING THE ANCHOR INTO THE PILOT HOLE, THE TIP OF THE ANCHOR SHAFT WAS FOUND TO BE BROKEN. THE BROKEN PIECE COULD NOT BE RETRIEVED AND REMAINS IN THE PATIENT. THE ANCHOR FUNCTIONED WITHOUT FURTHER ISSUES AND WAS USED AS IS, ALLOWING THE CASE TO BE COMPLETED. THERE WAS NO SIGNIFICANT DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. ADDITIONAL INFORMATION WAS RECEIVED ON 05-JAN-2026: THIS ISSUE OCCURRED DURING A LATERAL EXTRA-ARTICULAR TENODESIS.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 24039954 ↗

FDA clearance and classification

The FDA regulates product code MBI (fastener, fixation, nondegradable, soft tissue) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Arthrex® AR-3740SP

Has the Arthrex® AR-3740SP been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867428041), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-3740SP?

The FDA's MAUDE database lists 91 adverse event reports for this device, received between Sep 11, 2023 and Jul 31, 2026. In those reports, the most common specific device problems are “Break” (39), “Device Dislodged or Dislocated” (21) and “Use of Device Problem” (6); the most common patient problems are “Foreign Body In Patient” (22), “Failure of Implant” (2) and “Hypersensitivity/Allergic reaction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-3740SP?

None of the reports are classified as deaths, 2 as injuries and 89 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Arthrex® AR-3740SP MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the Arthrex® AR-3740SP, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-3740SP). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →