Arthrex® AR-3637Recalls & reported problems

Arthrex, Inc. · Knotless 1.8 FiberTak Implant System · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867459403), received between Oct 30, 2025 and Feb 11, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5

Patient problems

  • Foreign Body In Patient4
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2026Break

    ON 23-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.

    Patient problems: Foreign Body In Patient

    MAUDE report 24316437 ↗
  • MalfunctionFeb 11, 2026Break

    ON 23-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.

    Patient problems: Foreign Body In Patient

    MAUDE report 24316387 ↗
  • MalfunctionFeb 11, 2026Break

    ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.

    Patient problems: Foreign Body In Patient

    MAUDE report 24316259 ↗
  • MalfunctionFeb 11, 2026Break

    ON 23-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.

    Patient problems: Foreign Body In Patient

    MAUDE report 24316189 ↗
  • MalfunctionOct 30, 2025Break; Break

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-3637 KNOTLESS 1.8 FIBERTAK IMPLANT SYSTEM'S ANCHOR TIP SNAPPED IN HALF DURING INSERTION. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. AN AR-3636 KNOTLESS 1.8 FIBERTAK SOFT ANCHOR WAS USED TO COMPLETE THE CASE. THERE WAS NO CASE DELAY, AND IT IS UNCERTAIN IF ADDED ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED. THIS WAS DISCOVERED DURING A LABRUM REPAIR PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23425451 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →