Arthrex® AR-3637Recalls & reported problems
Arthrex, Inc. · Knotless 1.8 FiberTak Implant System · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867459403), received between Oct 30, 2025 and Feb 11, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
Patient problems
- Foreign Body In Patient4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 11, 2026Break
ON 23-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.
Patient problems: Foreign Body In Patient
MAUDE report 24316437 ↗MalfunctionFeb 11, 2026Break
ON 23-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.
Patient problems: Foreign Body In Patient
MAUDE report 24316387 ↗MalfunctionFeb 11, 2026Break
ON (B)(6) 2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.
Patient problems: Foreign Body In Patient
MAUDE report 24316259 ↗MalfunctionFeb 11, 2026Break
ON 23-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT TWO AR-3637 1.8 KNOTLESS FIBERTAK CONVERSION SUTURE SETS BROKE. THIS WAS DISCOVERED DURING A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2026. ADDITIONAL INFORMATION WAS RECEIVED ON 30-JAN-2026: DURING THE PROCEDURE, ISSUES WERE ENCOUNTERED WITH FOUR AR-3637 1.8 KNOTLESS FIBERTAK KITS. IN TWO OF THE KITS, ONE REPAIR SUTURE FROM ONE ANCHOR IN EACH KIT BROKE WHILE THE SURGEON ATTEMPTED TO SHUTTLE THE SUTURE THROUGH THE KNOTLESS MECHANISM. IN THE REMAINING TWO KITS, TWO REPAIR SUTURES FROM TWO ANCHORS BROKE IN THE SAME MANNER. ALL ANCHORS REMAINED IN PLACE, AND ALL SUTURE FRAGMENTS WERE RETRIEVED. THE CASE WAS COMPLETED USING AR-2323KBCC 4.75 MM KNOTLESS SWIVELOCK ANCHORS IN PLACE OF THE AFFECTED DEVICES. THE EVENT RESULTED IN AN ESTIMATED 5¿10-MINUTE SURGICAL DELAY. IT IS UNKNOWN WHETHER THE PATIENT REQUIRED ADDITIONAL ANESTHESIA.
Patient problems: Foreign Body In Patient
MAUDE report 24316189 ↗MalfunctionOct 30, 2025Break; Break
ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA PHONE THAT AN AR-3637 KNOTLESS 1.8 FIBERTAK IMPLANT SYSTEM'S ANCHOR TIP SNAPPED IN HALF DURING INSERTION. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. AN AR-3636 KNOTLESS 1.8 FIBERTAK SOFT ANCHOR WAS USED TO COMPLETE THE CASE. THERE WAS NO CASE DELAY, AND IT IS UNCERTAIN IF ADDED ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED. THIS WAS DISCOVERED DURING A LABRUM REPAIR PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23425451 ↗
Other Fastener, fixation, nondegradable, soft tissue devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,686 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,398 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE932 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE721 reports
- RIGIDLOOP 232447Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE633 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue624 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE609 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE460 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page