Skip to content

Arthrex® AR-1288QAI-100Recalls & reported problems

Arthrex, Inc. · All-Inside QuadLink Kit, 10mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867473553), received between Jun 24, 2025 and Mar 12, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Positioning Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2026Break

    ON (B)(6), A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT THE TIGHTROPE CONSTRUCT ON THE FEMORAL SIDE FROM AR-1288QAI-100 ALL-INSIDE QUADLINK KIT BROKE DURING THE PROCEDURE. THE CASE WAS CONTINUED USING A REPLACEMENT AR-1588RTT2-IB FIBERTAG TIGHTROPE II IMPLANT, WHICH ALSO BROKE. A SECOND AR-1588RTT2-IB FIBERTAG TIGHTROPE II IMPLANT FROM THE SAME LOT WAS THEN OPENED AND FUNCTIONED WITHOUT ISSUE. THE EVENT OCCURRED INTRAOPERATIVELY ON (B)(6), WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6): IT WAS REPORTED THAT AN AR-1588RTT2-IB FIBERTAG TIGHTROPE II IMPLANT DEVICE BROKE WHILE PULLING THE TIGHTROPE THROUGH THE FEMORAL TUNNEL AND TENSIONING IT DOWN DURING AN ACL RECONSTRUCTION PROCEDURE. THE BREAK OCCURRED INSIDE THE PATIENT. BOTH IMPLANTS THAT BROKE WERE REMOVED, AND ANOTHER TIGHTROPE WAS IMPLANTED ON THE FEMORAL SIDE. THE FEMORAL SIDE HAD TO BE REDONE TWICE, RESULTING IN AN APPROXIMATE 15-MINUTE SURGICAL DELAY. NO ADDITIONAL ANESTHESIA INFORMATION WAS REPORTED, OTHER THAN THE DELAY TIME. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING OR FOLLOWING THE PROCEDURE DUE TO THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24578951 ↗
  • MalfunctionJan 12, 2026Positioning Failure

    ON (B)(6) 2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-1288QAI-100 ALL-INSIDE QUADLINK KIT (10 MM) FAILED TO TENSION PROPERLY. THE TIGHTROPE MECHANISM APPEARED TO ¿GIVE OUT¿ WHILE ATTEMPTING TO SEAT THE GRAFT IN THE FEMORAL TUNNEL AND COULD NOT SUSPEND THE GRAFT USING THE TENSIONING TAILS. THE AUTOGRAFT WAS THEN REMOVED, AND THE DEFECTIVE TIGHTROPE WAS CUT OUT AND REPLACED WITH AN AR-1588BTB-IB TIGHTROPE II BTB, WHICH FUNCTIONED AS INTENDED. THE CASE WAS DELAYED, AND ADDITIONAL ANESTHESIA WAS LIKELY ADMINISTERED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS ISSUE OCCURRED DURING AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 29-DEC-2025: THE PROCEDURE WAS DELAYED BY APPROXIMATELY 30 MINUTES, AND THE AMOUNT OF ADDITIONAL ANESTHESIA ADMINISTERED IS UNKNOWN. THE ISSUE WITH THE TIGHTROPE IMPLANT WAS FIRST IDENTIFIED WHEN THE BUTTON WAS CONFIRMED TO BE FLIPPED, AND THE GRAFT WAS BEING TENSIONED ON THE FEMORAL SIDE. AT THAT POINT, THE IMPLANT FAILED. THERE WAS NO FRAYING OF THE TENSIONING TAILS DURING THE EVENT. THE CONSTANT UNLOADING OCCURRED SUDDENLY. NO FRAGMENTS REMAINED INSIDE THE PATIENT; ALL COM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24036176 ↗
  • MalfunctionDec 8, 2025Break; Break

    ON 11/17/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-1288QAI-100 QUAD LINKS FLIP CUTTER BROKE DURING USE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT. THE BROKEN PIECE WAS RETRIEVED WITHOUT ISSUES. ANOTHER KIT WAS OPENED, AND THE FLIP CUTTER WAS USED WITHOUT ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23742088 ↗
  • MalfunctionJun 24, 2025Break; Positioning Failure; Break; Positioning Failure

    ON 06/03/2025, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT THE FIBERTAG TIGHTROPE II WITH INTERNALBRACE, INCLUDED IN AN AR-1288QAI-100 ALL-INSIDE QUADLINK KIT, DID NOT PROPERLY TENSION THE GRAFT INTO THE FEMORAL TUNNEL. THE ISSUE WAS IDENTIFIED DURING AN ACLR PROCEDURE ON (B)(6) 2025. THE CASE WAS COMPLETED SUCCESSFULLY, AND NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22317743 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page