Skip to content

LEEP PRECISION™ Integrated System LP-10-120Recalls & reported problems

Coopersurgical, Inc. · LEEP PRECISION™ Integrated System (120 VAC) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00888937014228), received between May 26, 2022 and Aug 20, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (9 reports).

4 other recalls cover devices cleared under the same FDA submission (K963348); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem9
  • Output Problem3
  • Loss of Power1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Electric Shock1
  • Pain1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2026Electrical /Electronic Property Problem

    It was reported that the device gave an R/F Leakage error and didn't cauterize as it should. Unit to be returned for repair. No patient harm. (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50348310 ↗
  • MalfunctionSep 19, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23097162 ↗
  • MalfunctionJun 16, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT WHILE ATTEMPTING TO ENTER COAG MODE, THE DEVICE GIVES A "R/F LEAKAGE ERROR." DEVICE TO BE RETURNED FOR SERVICE. NO PATIENT HARM REPORTED. NO ADDITIONAL INFORMATION AVAILABLE. LP-10-120 LEEP 2025-06-0000090.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22212308 ↗
  • MalfunctionApr 2, 2025Loss of Power; Loss of Power

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21752114 ↗
  • MalfunctionAug 22, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20036673 ↗
  • MalfunctionAug 21, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED VIA MEDWATCH, THAT THE PATIENT WAS SCHEDULED TO UNDERGO A LEEP (LOOP ELECTROSURGICAL EXCISION PROCEDURE) ON (B)(6) 2024. THE LEEP UNIT WAS FOUND TO NOT FUNCTION PROPERLY WHEN TURNED ON. THERE WAS A PROBLEM WITH THE FOOT PEDAL CONTROL. BEFORE THE PROCEDURE WAS TO BEGIN, THE PATIENT REPORTED FEELING PAIN AND THE PROVIDER STATED THAT THE UNIT WAS ALREADY HOT. PATIENT DECLINED PROCEDURE DUE TO THE FEELING OF PAIN. PROVIDER INFORMED NURSE THAT THE DEVICE DID ARC TO THE PATIENT WITHOUT USING FOOT PEDAL. PATIENT WAS GIVEN IBUPROFEN FOR PAIN TREATMENT. PATIENT VITALS WERE STABLE AND SHE WAS DISCHARGED. ATTEMPTS FOR ADDITIONAL INFORMATION WERE MADE, BUT NO RESPONSE PROVIDED. LP-10-120 LEEP (B)(4).

    Patient problems: Pain

    MAUDE report 20016062 ↗
  • MalfunctionOct 24, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17999294 ↗
  • MalfunctionOct 13, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17931148 ↗
  • MalfunctionSep 25, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17808681 ↗
  • MalfunctionMay 22, 2023Output Problem; Output Problem

    AS PER THE DETAILS RECEIVED ON INCOMING E-COMPLAINT (B)(4): "WE HAVE A MEDICAL INCIDENT WHERE BOTH THE DOCTOR AND PATIENT WERE SHOCKED DURING A PROCEDURE.".

    Patient problems: Electric Shock

    MAUDE report 16973030 ↗
  • MalfunctionNov 4, 2022Output Problem; Output Problem

    GENERATOR IS OPERATING INTERMITTENTLY 1216677-2022-00294 LEEP PRECISION INTG. SYS. LP-10-120 E-COMPLAINT-(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15735663 ↗
  • MalfunctionMay 26, 2022Electrical /Electronic Property Problem; Nonstandard Device; Output Problem; Electrical /Electronic Property Problem

    ADDITIONAL COMPLAINT IN REF TO (B)(4) SECOND PATIENT INCIDENT. REPORT FORWARDED BY CUSTOMER SERVICE REPRESENTATIVE TRUMBULL. I GOT A CALL FROM (B)(6) MEDICAL CENTER, (B)(4). THEIR ACCOUNT NUMBER IS (B)(6). THEY HAVE THE LEEP PRECISION. THE UNIT IS NOT WORKING CORRECTLY AND TWO PATIENTS WERE BURNED. THE SERIAL NUMBER IS (B)(4). IT TURNS OUT THAT IS PART OF THE RECALL. LEEP PRECISION INTG SYS LP-10-120 (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions; Superficial (First Degree) Burn

    MAUDE report 14515747 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K963348. One submission often covers many models, so check the product and lot details in each record.

  • Z-0258-2022 · Sep 29, 2021 · Open, Classified
    LEEP PRECISION Integrated System, 220V CooperSurgical part numbers: LP-10-220
  • Z-0257-2022 · Sep 29, 2021 · Open, Classified
    LEEP PRECISION Integrated System, 120V CooperSurgical part number: LP-10-120
  • Z-2057-2019 · Jun 18, 2019 · Terminated
    LEEP Precision Integrated System 120V & LEEP Precision Generator, Model Nos. LP-10-120 & LP-20-120
  • Z-1840-2019 · Apr 23, 2019 · Terminated
    LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Ge

Other Electrocautery, Gynecologic (And Accessories) devices

All Electrocautery, Gynecologic (And Accessories) devices →

Sources and updates