ENSEAL X1 NSLX120LRecalls & reported problems
Ethicon Endo-Surgery, LLC · ENSEAL 20 cm shaft, Large Jaw · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon ENSEAL X1 is an electrosurgical instrument used to seal and cut tissue during surgery. Of its 170 reports, 45 are filed as injuries and 2 as deaths; many describe jaws that will not open or close, sometimes during hysterectomies, and some describe bleeding or a piece coming off the jaws. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 172
- Deaths reported
- 2
- Injuries reported
- 45
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 172 adverse event reports that list this exact device (DI 10705036015024), received between Nov 21, 2017 and Sep 23, 2026 — classified as 2 deaths, 45 injuries, 125 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (58 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close58
- Adverse Event Without Identified Device or Use Problem18
- Device Displays Incorrect Message16
- Appropriate Device Problem Term/Code Not Available14
- Activation, Positioning or Separation Problem9
- Break7
- Mechanical Problem6
- Device Operates Differently Than Expected5
Patient problems
- No Consequences Or Impact To Patient62
- No Clinical Signs, Symptoms or Conditions56
- Hemorrhage/Blood Loss/Bleeding12
- No Code Available12
- Blood Loss10
- Burn(s)6
- Hematoma6
- Burn, Thermal4
Recent adverse event reports
The 25 most recent of 172 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Difficult to Open or Close
It was reported that during an unknown procedure, the handles were pressed to close and they got stuck without the possibility of being able to open. No patient consequences reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868800 ↗MalfunctionSep 11, 2026Structural Problem
It was reported that during a Laparotomy procedure the device was opened and the handle broke off. Opened a new device to continue operation. No patient consequences reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50733473 ↗MalfunctionMay 20, 2026Difficult to Open or Close
IT WAS REPORTED THAT DURING A HYSTERECTOMY PROCEDURE, THE DEVICE CLAMPED DOWN AND WOULD NOT RELEASE, RESULTING IN THE JAW NOT OPENING. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25233331 ↗MalfunctionMar 30, 2026Detachment of Device or Device Component
IT WAS REPORTED DEVICE ¿MIS FIRED¿ AND A PIECE CAME OFF THE JAWS DURING AN OPEN TOTAL ABDOMINAL HYSTERECTOMY. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24735165 ↗MalfunctionFeb 9, 2026Break
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE'S HANDLE LATCH BROKEN. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24291697 ↗MalfunctionJan 26, 2026Difficult to Open or Close
IT WAS REPORTED THAT DURING A VET PROCEDURE, ON FIRST USE, THE TRIGGER DROPPED OUT AND CAME LOOSE, PREVENTING IT FROM CLOSING. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24177189 ↗MalfunctionDec 15, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN EXPLORATORY LAPAROTOMY THE HANDLE WHEN THEY OPEN AND CLOSE IT GET KEEP STOCK AND WOULDN'T OPEN. SO, IF CLOSE THE JAW OFF THE DEVICE, SEAL AND CUT THEN IT WOULDN'T OPEN BACK. THEY SAID IT BASICALLY DOING THAT AS SOON IT CAME OUT OF THE BOX. THEY OPEN A NEW DEVICE AND THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23804820 ↗MalfunctionDec 8, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT, DURING AN UNKNOWN SURGERY, THE HANDLE DID NOT OPEN WHEN THE DEVICE WAS USED. THERE WAS NO EFFECT ON THE PATIENT. ANOTHER COMPETITIVE DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23744796 ↗MalfunctionDec 8, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT, DURING AN UNKNOWN SURGERY, THE HANDLE DID NOT OPEN WHEN THE DEVICE WAS USED. ANOTHER DEVICE WAS USED. THERE WAS NO EFFECT ON THE PATIENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23744708 ↗MalfunctionNov 24, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN THE SURGEON WAS ACTIVATING IT THE HANDLE WOULDN'T UNLOCK. NO ADVERSE IMPACT PATIENT/USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23631563 ↗MalfunctionNov 24, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN THE SURGEON WAS ACTIVATING IT THE HANDLE WOULDN'T UNLOCK. ONE OF THEM IS IN A COMPLETE LOCK BUT HE WAS ABLE TO PULL IT OFF THE PATIENT. NO ADVERSE IMPACT PATIENT/USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23631496 ↗MalfunctionNov 20, 2025Difficult to Open or Close; No Apparent Adverse Event; Difficult to Open or Close; No Apparent Adverse Event
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE CLAMP WAS USED AND REMAINED ATTACHED TO THE TISSUE, UNABLE TO OPEN IT. WE HAD TO TEAR THE FABRIC TO RELEASE IT. THE CLAMP WAS NOT UNLOCKED. THERE ARE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23607376 ↗MalfunctionNov 20, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT, DURING A TOTAL CYSTECTOMY, THE JAWS DID NOT OPEN EVEN WHEN THE RATCHET WAS RELEASED AFTER SEALING AND CUTTING THE TISSUE. THERE WAS NO TISSUE DAMAGE WHEN THE JAWS WERE FORCIBLY OPENED AND CLOSED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23604944 ↗MalfunctionNov 10, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, BEFORE USED ON THE PATIENT, THE PACKAGE WAS FOUND DAMAGED. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509856 ↗MalfunctionAug 6, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TRIGGER THAT TRIGGERS COAGULATION STUCK AND DID NOT RETURN TO THE INITIAL POSITION, MAKING IT IMPOSSIBLE TO USE THE DEVICE. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22728787 ↗MalfunctionJun 9, 2025Difficult to Open or Close; No Apparent Adverse Event; Difficult to Open or Close; No Apparent Adverse Event
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE LOCKED UP. THE JAW WOULD NOT OPEN, AND IT WAS TEARING THE TISSUE ON THE REMOVAL. THEY WHERE ABLE TO OPEN AND REPLACE IT WITH A NEW ONE. NO ADVERSE IMPACT TO PATIENT/USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22168149 ↗MalfunctionApr 21, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING BOWEL RESECTION PROCEDURE THE CUT BUTTON WAS NOT WORKING. NO PATIENT HARM REPORTED. THERE WAS NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875511 ↗MalfunctionFeb 3, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED BY THE DISTRIBUTOR THAT DURING AN UNKNOWN PROCEDURE THE DEVICE: THE JAW REMAINS CLOSED AFTER THE CAUTERY CYCLE. THE SURGEON COULD NOT OPEN IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21296603 ↗MalfunctionJan 31, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING A HYSTERECTOMY THE DEVICE LOCKED WITHIN THE PROCEDURE AND COULD NOT BE OPENED. ANOTHER DEVICE WAS OPENED AND USED TO CONTINUE THE PROCEDURE. THE SURGERY WAS DELAYED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21283290 ↗MalfunctionDec 9, 2024Self-Activation or Keying; Self-Activation or Keying
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THAT THE DEVICE WAS ACTIVATING BY ITSELF. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878262 ↗MalfunctionNov 19, 2024
SURGEON WAS USING THE DEVICE. THE DEVICE WAS NOT CLOSING AND GIVING ERROR MESSAGE. DEVICE WAS REMOVED AND REPLACED WITH NEW PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20722860 ↗MalfunctionNov 13, 2024Difficult to Open or Close; No Apparent Adverse Event; Difficult to Open or Close; No Apparent Adverse Event
IT WAS REPORTED THAT DURING AN EXPLORATORY LAPAROTOMY MASS DISSECTION AND TOTAL ABDOMINAL HYSTERECTOMY THAT WHILE DISSECTING THE IP LIGAMENT DEVICE WAS CLOSED TO SEAL AND CUT BUT THE JAWS WOULD NOT OPEN. TROUBLE SHOOTING STEPS WERE TAKEN BUT THE JAWS WOULD NOT OPEN. DEVICE WAS REMOVED BY DISSECTED AROUND IT IS USING A COMPETITOR'S DEVICE. COMPETITORS DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20675440 ↗MalfunctionOct 18, 2024Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THEY WERE UNABLE TO OPEN JAWS OF INSTRUMENT AND REQUESTED TROUBLESHOOTING IN OPENING JAW OF PRODUCT. WENT OVER STEPS OF USE WITHOUT JAW OPENING. INSTRUMENT WAS REMOVED OFF OF PATIENT PRIOR TO CALL AND ANOTHER INSTRUMENT WAS USED TO FINISH CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20482184 ↗MalfunctionSep 30, 2024Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT DURING A HYSTERECTOMY THERE WAS TISSUE ADHESION ON THE JAWS. NO PATIENT CONSEQUENCES. THIS IS A VET PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20332804 ↗MalfunctionAug 14, 2024Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE DEVICE "WOULD NOT RELEASE". THERE WAS NO PATIENT CONSEQUENCES REPORTED. SURGICAL DELAY UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19985435 ↗
Other ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES) devices
- Enseal NSLX137CEthicon Endo-Surgery, LLC · ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES)3 deaths344 reports
- Fischer Cone Biopsy Excisor 900-157Coopersurgical, Inc. · Electrocautery, Gynecologic (And Accessories)1 recall0 reports
- LEEP PRECISION™ Generator LP-20-120Coopersurgical, Inc. · Electrocautery, Gynecologic (And Accessories)45 reports
- Harmonic NSLX137SEthicon Endo-Surgery, LLC · ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES)1 death19 reports
- LEEP PRECISION™ Integrated System LP-10-120Coopersurgical, Inc. · Electrocautery, Gynecologic (And Accessories)12 reports
- Fischer Cone Biopsy Excisor 900-151Coopersurgical, Inc. · Electrocautery, Gynecologic (And Accessories)10 reports
- AESCULAP PM438RAesculap AG · ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES)7 reports
- Valleylab E1564Covidien LP · ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES)6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page