Skip to content

AESCULAP PM438RRecalls & reported problems

Aesculap AG · MARYLAND ADTEC BIPOLAR Grasping Forceps, jaw inserts, curved, (310 mm), diam. 5 mm, fenestrated, double action, reusable · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04038653322862), received between Nov 3, 2020 and Dec 17, 2021 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Material Integrity Problem3
  • Break2
  • Material Fragmentation2
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 17, 2021Material Fragmentation; Material Fragmentation

    INVOLVED COMPONENTS. PM973R - INSULATED OUTER TUBE 5/5MM 310MM - LOT UNKNOWN. PM450R - HANDLE FOR BIPOLAR INSTRUMENTS - LOT UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 13027644 ↗
  • MalfunctionDec 10, 2021Break; Break

    CORRECTION / ADDITIONAL INFORMATION: PATIENT INJURY HAD OCCURRED AND PATIENT HARM WAS REPORTED AS "THE FRAGMENTS REAMINED IN SITU". ADDED INVOLVED COMPONENTS. THE FRAGMENT REMAINED IN SITU. INVOLVED COMPONENTS: PM450R - HANDLE FOR BIPOLAR INSTRUMENTS - LOT UNKNOWN. PM973R - INSULATED OUTER TUBE 5/5MM 310MM - LOT UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12975779 ↗
  • MalfunctionJul 13, 2021Break; Break

    IT WAS REPORTED TO AESCULAP AG THAT A JAW INS.BIP.MARYLAND DISS.FEN.5/310MM (PART # PM438R) WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE JAW BROKE. A BRANCH FELL INTO THE SITUS BUT COULD BE RECOVERED. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE XC (B)(4). INVOLVED COMPONENTS: PM450R - HANDLE FOR BIPOLAR INSTRUMENTS - LOT UNKNOWN. PM973R - INSULATED OUTER TUBE 5/5MM 310MM - LOT UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 12159283 ↗
  • MalfunctionMay 11, 2021Fracture; Material Fragmentation; Fracture; Material Fragmentation

    INVOLVED COMPONENTS: PM600201 - METAL INNER TUBE 31CM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11806408 ↗
  • MalfunctionNov 3, 2020Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A MARYLAND FIXATION FORCEPS JAW INSERT. ACCORDING TO THE COMPLAINANT, THE CERAMIC PART OF THE JAW CAME OFF DURING SURGERY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT HAS NOT YET BEEN RECEIVED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10782215 ↗
  • MalfunctionNov 3, 2020Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A MARYLAND FIXATION FORCEPS JAW INSERT. ACCORDING TO THE COMPLAINANT, THE CERAMIC PART OF THE JAW CAME OFF DURING SURGERY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10781980 ↗
  • MalfunctionNov 3, 2020Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A MARYLAND FIXATION FORCEPS JAW INSERT. ACCORDING TO THE COMPLAINANT, THE CERAMIC PART OF THE JAW CAME OFF DURING SURGERY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT HAS NOT YET BEEN RECEIVED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED. THE MALFUNCTION IS FILED UNDER AAG REFERENCE: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10781976 ↗

FDA clearance and classification

The FDA regulates product code HGI (electrocautery, gynecologic (and accessories)) under 21 CFR 884.4120 as a Class II device. 21 CFR 884.4120 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AESCULAP PM438R

Has the AESCULAP PM438R been recalled?

No FDA recall record in our data names this exact device identifier (DI 04038653322862), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AESCULAP PM438R?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 3, 2020 and Dec 17, 2021. In those reports, the most common specific device problems are “Material Integrity Problem” (3), “Break” (2) and “Material Fragmentation” (2); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AESCULAP PM438R?

None of the reports are classified as deaths, 1 as an injury and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the AESCULAP PM438R, and is it still sold?

The FDA's device database lists Aesculap AG as the company responsible for it (brand AESCULAP, model PM438R). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES) devices

All ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page