NexGen® 5881-26Recalls & reported problems
Zimmer, Inc. · Class 1
All models of NexGen (198)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00889024199958), received between Jun 21, 2022 and Mar 16, 2026 — classified as 1 injury, 13 malfunctions.
The most frequently reported specific device problem is “Fracture” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture14
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Foreign Body In Patient1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 16, 2026Fracture
IT WAS REPORTED THAT A SCREWDRIVER FRACTURED WHILE TIGHTENING. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24606163 ↗MalfunctionFeb 25, 2026Fracture
IT WAS REPORTED DURING KNEE ARTHROPLASTY THAT THE HEX SCREW DRIVER HEAD FRACTURED WHILE TORQUING THE BOLT. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24445298 ↗MalfunctionJan 6, 2026Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23981022 ↗MalfunctionFeb 18, 2025Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF A SCREWDRIVER FRACTURED DURING SURGERY. THE FRACTURED PIECE WAS RECOVERED BY THE SURGEON. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21399653 ↗MalfunctionAug 7, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Foreign Body In Patient
MAUDE report 19927755 ↗MalfunctionMar 6, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18848773 ↗MalfunctionFeb 27, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18789808 ↗MalfunctionDec 18, 2023Fracture; Fracture
NO ADDITIONAL INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18348277 ↗MalfunctionDec 5, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18266291 ↗MalfunctionOct 31, 2023Fracture; Fracture
IT WAS REPORTED THE BROKEN SCREW DRIVER WAS FOUND IN THE LOANER KIT BY THE WAREHOUSE TEAM DURING CHECK IN, AFTER THE PRODUCRT WAS RECEIVED. THE CUSTOMER'S STERILIZATION SERVICE CONFIRMED BY TELEPHONE THE LOANER KIT WAS USED FOR A SURGERY, THAT THE INSTRUMENT WAS RETURNED BACK TO ZIMMER BIOMET IN THE KIT, BUT NO IMPACT ON PATIENT OR NO INCIDENT DURING THE SURGERY WAS REPORTED TO THE SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18038388 ↗MalfunctionOct 31, 2023Fracture; Fracture
IT WAS REPORTED THE BROKEN SCREW DRIVER WAS FOUND IN THE LOANER KIT BY THE WAREHOUSE TEAM DURING CHECK IN, AFTER THE PRODUCT WAS RECEIVED. THE CUSTOMER'S STERILIZATION SERVICE CONFIRMED BY TELEPHONE THE LOANER KIT WAS USED FOR A SURGERY, THAT THE INSTRUMENT WAS RETURNED BACK TO ZIMMER BIOMET IN THE KIT, BUT NO IMPACT ON PATIENT OR NO INCIDENT DURING THE SURGERY WAS REPORTED TO THE SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18038385 ↗MalfunctionAug 29, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A KNEE SURGERY, THE SURGEON WAS USING THE SCREWDRIVER WITH ITS TIP TO UNSCREW THE HINGE SCREW FROM THE PROSTHESIS IN A COUNTER-CLOCKWISE DIRECTION. AFTER THREE-FOURTHS OF A TURN, THE SCREWDRIVER JAMMED AND BROKE OFF INTO THE HINGE SCREW. IT WAS THEN NOT POSSIBLE TO REMOVE THE PROSTHESIS WITHOUT REMOVING THE HINGE. THIS RESULTED IN THE USE OF A LARGE NUMBER OF COMPLIMENTARY BOXES LEADING TO A RISK OF LACK OF MATERIAL FOR THE OTHER PATIENTS FOR THE DAY. IT ALSO INCREASED OPERATING TIME BY 2 HOURS. A DRILL BIT WAS USED TO REMOVE THE HINGE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17640953 ↗InjuryAug 29, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A REVISION KNEE PROCEDURE, THE HEX DRIVER FRACTURED WHILE ATTEMPTING TO DECOUPLE THE KNEE PROSTHESIS. THE PROCEDURE WAS UNABLE TO BE COMPLETED AT THAT TIME. DUE DILIGENCE IS COMPLETE AS MULTIPLE ATTEMPTS WERE MADE; HOWEVER, NO FURTHER INFORMATION OR PRODUCT HAS BEEN RECEIVED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15314666 ↗MalfunctionJun 21, 2022Fracture; Fracture
IT WAS REPORTED THAT THE TORQUE WENCH TIP BROKE OFF WHEN TRYING TO LOOSEN THE PEG SO THE SURGEON HAD TO USE A BURR TO CUT OFF THE HINGE POST TO REMOVE THE WHOLE FEMUR. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14754731 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page