Quick summary

This Zimmer NexGen femoral impactor is a manual surgical tool used during knee replacement procedures to seat the implant. FDA records show 25 malfunction reports and no deaths or injuries. The reports describe the impactor pad or block breaking in half during surgery, typically without fragments entering the patient and without surgical delays. No recalls name this device. Report volume has been lower in the most recent 12 months (4 reports) than the prior period (10 reports).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00889024201668), received between Oct 28, 2020 and Jul 13, 2026 — classified as 24 malfunctions.

The most frequently reported specific device problem is “Fracture” (23 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 5 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture23
  • Break1
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 13, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25800057 ↗
  • MalfunctionDec 10, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A KNEE ARTHROPLASTY, AFTER PLACEMENT OF THE DEFINITIVE IMPLANT, THE SURGEON IMPACTED TO ADJUST THE IMPLANT AND THE FEMORAL IMPACTOR BROKE. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. THE IMPLANT REMAINED INTACT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23769411 ↗
  • MalfunctionDec 2, 2025Fracture; Fracture

    IT WAS REPORTED THAT AN IMPACTOR PAD BROKE DURING SURGERY. THE SURGICAL TECHNIQUE WAS UTILIZED; THERE WAS NO SURGICAL DELAY. THERE WERE NO FOREIGN BODIES RETAINED. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23693379 ↗
  • MalfunctionOct 31, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23437588 ↗
  • MalfunctionOct 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A TOTAL KNEE PROCEDURE, THE FEMORAL IMPACTOR BLOCK FRACTURED. THERE WAS NO PATIENT IMPACT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. BACKUP INSTRUMENTATION WAS USED TO COMPLETE THE PROCEDURE WITHOUT FURTHER INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23405600 ↗
  • MalfunctionOct 28, 2025Crack; Fracture; Crack; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23405518 ↗
  • MalfunctionOct 17, 2025Fracture; Fracture

    IT WAS REPORTED THE FEMORAL IMPACTOR BROKE DURING IMPACTION. NO PATIENT HARM OR IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23322594 ↗
  • MalfunctionOct 1, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23189750 ↗
  • MalfunctionSep 16, 2025Fracture; Fracture

    IT WAS REPORTED THAT AN IMPACTOR WAS FOUND BROKEN BY SPD BEFORE STERILIZING. ATTEMPTS FOR ADDITIONAL INFORMATION HAVE BEEN MADE AND ALL HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23062233 ↗
  • MalfunctionJul 1, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL SURGERY, THE IMPACTOR BROKE WHEN THE SURGEON GAVE THE FINAL BLOW DURING IMPACTION OF THE DEFINITIVE FEMORAL COMPONENT. THERE WAS NO SURGICAL DELAY. THE SURGICAL TECHNIQUE WAS UTILIZED. THE SURGERY WAS COMPLETED WITH ANOTHER DEVICE. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22371711 ↗
  • MalfunctionJun 30, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE IMPACTOR BROKE DURING IMPACTION. THE SURGICAL TECHNIQUE WAS UTILIZED. THERE WAS NO SURGICAL DELAY. NO FOREIGN BODIES WERE RETAINED. THE SURGERY WAS COMPLETED WITH ANOTHER DEVICE. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22356703 ↗
  • MalfunctionMar 7, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE FEMORAL IMPACTOR BROKE IN SURGERY. THE SURGICAL TECHNIQUE WAS UTILIZED. THERE WAS NO SURGICAL DELAY. NO FOREIGN BODIES WERE RETAINED. THE SURGERY WAS COMPLETED WITH A SECOND DEVICE. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21547137 ↗
  • MalfunctionJan 28, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21255379 ↗
  • MalfunctionAug 22, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE WAS FRACTURED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20030103 ↗
  • MalfunctionAug 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE IMPACTOR HEAD FRACTURED DURING REMOVAL OF A TRIAL PROSTHESIS AS PART OF A KNEE PROCEDURE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19892730 ↗
  • MalfunctionJun 17, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19552225 ↗
  • MalfunctionApr 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT WHILE USING THE FEMORAL IMPACTOR, THE FEMORAL IMPACTOR BLOCK BROKE IN HALF. THE SURGICAL TECHNIQUE WAS UTILIZED. THERE WAS NO SURGICAL DELAY. NO FOREIGN BODIES WERE RETAINED. THERE WAS NO PATIENT HARM. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19020569 ↗
  • MalfunctionOct 16, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL TOTAL KNEE PROCEDURE, THE HEAD OF THE IMPACTING INSTRUMENT FRACTURED DURING FEMORAL IMPLANTATION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH A NEW DEVICE. THERE WAS NO PATIENT IMPACT OR ADVERSE EVENTS AS A RESULT OF THE MALFUNCTION. DILIGENCE IS IN PROCESS FOR THIS EVENT; TO DATE NO FURTHER INFORMATION HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17938968 ↗
  • MalfunctionJun 14, 2023Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17127144 ↗
  • MalfunctionNov 8, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING THE IMPACTING OF THE FEMUR, THE BLACK FEMORAL IMPACTOR BLOCK BROKE. NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15752637 ↗
  • MalfunctionSep 16, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING KNEE ARTHROPLASTY, THE FEMORAL IMPACTOR FRACTURED WHEN IT WAS IMPACTED WITH A HAMMER. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15436979 ↗
  • MalfunctionMar 3, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING KNEE ARTHROPLASTY A CRACK WAS IDENTIFIED IN THE TIBIAL ARTICULAR SURFACE PROVISIONAL PRIOR TO TRIALING AND THE FEMORAL IMPACTOR FRACTURED IN HALF UPON IMPACTION WITH THE FEMUR. ADDITIONAL INSTRUMENTS WERE USED TO COMPLETE THE PROCEDURE. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13661297 ↗
  • MalfunctionMay 28, 2021Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11900785 ↗
  • MalfunctionOct 28, 2020Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Consequences Or Impact To Patient