Skip to content

Phantom Hindfoot TTC/TC Nail System P99-110-4630Recalls & reported problems

Paragon 28, Inc. · Drill, 4.6 x 300mm, Cannulated, 3/16" Sq. Connection · Class 1

All models of Phantom Hindfoot TTC/TC Nail System (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 30, 2021Class IITerminated

    A thin wall condition between the inner cannulation and the flutes of the drill which potentially could cause a fracture through the wall of the drill to the cannulation beneath.

    Z-2488-2021 · Root cause: Process change control · Recall details · FDA record ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page