Skip to content

Grappler R3FLEX Stabilization System P82-001-1237-SKRecalls & reported problems

Paragon 28, Inc. · Grappler R3FLEX Single Implant Kit · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889795128140), received between Feb 6, 2026 and Feb 6, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Fracture1
  • Positioning Failure1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 6, 2026Positioning Failure; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ATTEMPTED IMPLANT PLACEMENT UTILIZING TIBIAL AND FIBULAR TUNNELS, THE IMPLANT COULD NOT BE ADVANCED AND INTRA-OPERATIVE DIFFICULTIES OCCURRED WITH THE INSERTER AND SUTURE. FOLLOWING DRILLING AS PER THE SURGICAL TECHNIQUE, THE IMPLANT WAS ADVANCED THROUGH THE FIBULAR TUNNEL AND THE TIBIAL TUNNEL WAS SOUGHT. AFTER INITIAL RESISTANCE, A MALLET WAS USED PER TECHNIQUE; FLUOROSCOPIC IMAGING SHOWED THE IMPLANT HAD NOT ADVANCED. THE SURGEON ATTEMPTED TO REMOVE THE IMPLANT THROUGH THE FIBULAR TUNNEL TO REDRILL AND VERIFY TRAJECTORY BUT WAS UNABLE TO RETRIEVE IT. A SUBSEQUENT ATTEMPT TO LOCATE THE TIBIAL TUNNEL WAS MADE, AND AFTER APPLYING SIGNIFICANT FORCE THE IMPLANT ADVANCED SLIGHTLY, THEN THE INSERTER TIP DETACHED FROM THE GRAY HANDLE. WHEN THE SURGEON LIGHTLY TUGGED THE GRAY HANDLE TO DEPLOY THE SOCK, THE SUTURE BROKE AND EXITED THROUGH THE FIBULAR TUNNEL WHICH RESULTED IN THE SOCK BEING RETAINED BY THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24276778 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →