Skip to content

Norwegian, Rapid Infuser RI-2, 230V, 750 ml/min 903-00039A-NRecalls & reported problems

Belmont Instrument Corporation · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00896128002909), received between Sep 13, 2019 and Sep 13, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 13, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    OUR DISTRIBUTOR RECEIVED A COMPLAINT FROM THE USER FACILITY AND PROVIDED THE FOLLOWING REPORT: "WHILE USING OUR BELMONT, USERS DISCOVERED THAT THE EQUIPMENT WAS RUNNING ON BATTERY POWER AND THAT THE OUTLET ON THE POWER OUTLET WAS WITHOUT VOLTAGE. (THE INDICATOR LIGHT ON THE PLUG WAS OFF). THEY MOVED THE APPLIANCE CORD TO ANOTHER OUTLET. THE SECURITY OF THIS WITHDRAWAL WENT IMMEDIATELY. THE BELMONT UNIT GOT TRIED ON A THIRD COURSE WHERE THE COURSE FUSE ALSO WENT BEFORE THE EQUIPMENT WAS TAKEN OUT OF SERVICE. OTHER EQUIPMENT THAT WAS IN USE ALSO LOST TENSION, SO IT BECAME A SITUATION THAT WAS EXPERIENCED AND IN PART BECAME CRITICAL. WE LOST MONITORING ON THE MX 700 BUT USED THE X2 ON BATTERY, AS WELL AS ALL INFUSION PUMPS WENT ON BATTERY. DUE TO THE PRESENCE OF TECHNICAL PERSONNEL, THEY WERE QUICKLY CONNECTED TO THE THREE COURSE FUSES SO THAT THE OPERATION COULD CONTINUE WITHOUT HARM TO THE PATIENT."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9048718 ↗

Other Pump, Infusion devices

All Pump, Infusion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page