Skip to content

Sechrist Industries, Inc. 3300HR-01-0Recalls & reported problems

Sechrist Industries, Inc. · HBO CHMBR,ASSY,33IN,RVRS,ENGLISH,WHITE,CLASSIC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00899660002055), received between May 14, 2025 and May 14, 2025 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Cramp(s) /Muscle Spasm(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER REPORTED TO SECHRIST THAT SEVERAL PATIENTS HAVE COMPLAINED OF SEVERE CRAMPING AFTER COMING OUT OF THE CHAMBER. CUSTOMER DID NOT REPORT ANY ISSUES WITH THE CHAMBER. THEY REQUESTED SECHRIST COME OUT AND EVALUATE THE CHAMBER ANYWAY. THEY WERE ALSO GOING TO CONTACT THE SUPPLIER OF THEIR MEDICAL GAS FOR A PURITY REPORT.

    Patient problems: Cramp(s) /Muscle Spasm(s)

    MAUDE report 22021872 ↗

Other Chamber, Hyperbaric devices

All Chamber, Hyperbaric devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page