Skip to content

NA 7205325Recalls & reported problems

Smith & Nephew, Inc. · 5.5 MM ABRADER STRAIGHT BURRS, POWER/EP-1, BLACK, 8000 MAXIMUM RPM, PACKAGED 6 PER BOX, STERILE · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03596010251534), received between Sep 21, 2017 and Aug 25, 2020 — classified as 8 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Material Fragmentation6
  • Flaked4
  • Material Separation1
  • Metal Shedding Debris1
  • Peeled/Delaminated1

Patient problems

  • No Known Impact Or Consequence To Patient6
  • No Information4
  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 25, 2020Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE; THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH LESS THAN 30 MINUTES DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10448645 ↗
  • InjuryAug 24, 2020Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE; THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH LESS THAN 30 MINUTES DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10445601 ↗
  • InjuryAug 24, 2020Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE; THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH LESS THAN 30 MINUTES DELAY USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10445538 ↗
  • InjuryAug 24, 2020Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE; THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH LESS THAN 30 MINUTES DELAY USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10445485 ↗
  • InjuryAug 18, 2020Flaked; Material Fragmentation; Flaked; Material Fragmentation

    IT WAS REPORTED THAT DURING HIP ARTHROSCOPY PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE. IT IS UNKNOWN IF THERE WAS A DELAY AND HOW THE PROCEDURE WAS COMPLETED. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Information

    MAUDE report 10423592 ↗
  • InjuryAug 18, 2020Flaked; Flaked

    IT WAS REPORTED THAT DURING HIP ARTHROSCOPY PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE. IT IS UNKNOWN IF THERE WAS A DELAY AND HOW THE PROCEDURE WAS COMPLETED. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Information

    MAUDE report 10423588 ↗
  • InjuryAug 18, 2020Flaked; Flaked

    IT WAS REPORTED THAT DURING HIP ARTHROSCOPY PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE. IT IS UNKNOWN IF THERE WAS A DELAY AND HOW THE PROCEDURE WAS COMPLETED. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Information

    MAUDE report 10423582 ↗
  • InjuryAug 18, 2020Flaked; Material Fragmentation; Flaked; Material Fragmentation

    IT WAS REPORTED THAT DURING HIP ARTHROSCOPY PROCEDURE, THE BLADE WAS SHOWING METAL DEBRIS IN THE JOINT WHEN IT IS USED AND RUBBING ON THE INSIDE SHAFT OF THE BLADE. IT IS UNKNOWN IF THERE WAS A DELAY AND HOW THE PROCEDURE WAS COMPLETED. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Information

    MAUDE report 10423541 ↗
  • MalfunctionNov 28, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT DURING THE PROCEDURE, WHILE THE BLADE WAS BEING USED IN THE PATIENT, PLASTIC SHAVINGS OF THE CONNECTOR FELL OFF INTO THE JOINT. THE PLASTIC SHAVINGS WERE TRIED TO BE REMOVED BY FORCEPS, BUT COULD NOT COMPLETELY BE TAKEN OUT FROM THE PATIENT. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE AND NO PATIENT INJURIES WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: Foreign Body In Patient

    MAUDE report 9390505 ↗
  • MalfunctionFeb 13, 2018Peeled/Delaminated

    IT WAS REPORTED THAT THE DEVICE IS SHEDDING. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7268058 ↗
  • MalfunctionSep 21, 2017Metal Shedding Debris

    IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY THE BURR WAS SHEDDING, SUCTION WAS USED. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6883871 ↗

FDA clearance and classification

The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 7205325

Has the NA 7205325 been recalled?

No FDA recall record in our data names this exact device identifier (DI 03596010251534), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 7205325?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Sep 21, 2017 and Aug 25, 2020. In those reports, the most common specific device problems are “Material Fragmentation” (6), “Flaked” (4) and “Material Separation” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 7205325?

None of the reports are classified as deaths, 8 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 7205325 FDA approved?

Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.

Who makes the NA 7205325, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 7205325). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page