Skip to content

NA 7205335Recalls & reported problems

Smith & Nephew, Inc. · 4.5 MM LONG CURVED FULL RADIUS CONCAVE SHAVER BLADES, 18 CM WORKING LENGTH, PACKAGED 3 PER BOX, STERILE · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03596010251626), received between May 15, 2018 and Jul 30, 2024 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Corroded1
  • Crack1
  • Device Markings/Labelling Problem1
  • Failure to Cut1
  • Fracture1
  • Improper or Incorrect Procedure or Method1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient4

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 30, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT, DURING SURGERY, THE FULL RADIUS BLADE WAS DELAMINATING, A LAYER OF METAL CAME OFF. THE METAL MATERIAL WAS SUCKED PUT OF THE PATIENT USING A DYONICS SHAVER DEVICE. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE INSTEAD. THERE WAS A DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED. PATIENT STATUS IS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19861160 ↗
  • MalfunctionMar 11, 2024Break; Nonstandard Device; Break; Nonstandard Device

    IT WAS REPORTED THAT, DURING SET UP, WHEN THE FULL RADIUS BLADE PACKAGE WAS OPENED, IT WAS FOUND THAT THE PRODUCT BLADE WAS NOT EXPOSED. THERE WAS A BACK-UP DEVICE AVAILABLE AND NO DELAY WAS REPORTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18879415 ↗
  • MalfunctionNov 6, 2023Corroded; Corroded

    IT WAS REPORTED THAT DURING SET UP OR INSPECTION IN AN ACL, THE NEWLY PURCHASED CURVED CONCAVE FULL RADIUS WAS RUSTY. THE CLINICIAN REFUSES TO USE THEM. THERE WAS A S+N BACK-UP DEVICE AVAILABLE. NON-SIGNIFICANT DELAY WAS REPORTED, AND NO PATIENT COMPLICATIONS WERE REPORTED. AS PER THE PRELIMINARY RESULTS OF THE INVESTIGATION, THE INNER SHAFT OF DEVICE ONE HAS BEEN FRACTURED PROXIMAL TO THE CURVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18079077 ↗
  • MalfunctionNov 3, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE CURVED FULL RADIUS INNER SHAFT WAS PUSHED ALL THE WAY IN AND WAS JAMMED IN THE DISTAL TIP OF THE OUTER SHAFT. ALSO, THE INNER SHAFT WAS FRACTURED PROXIMAL TO THE CURVE. THE DEFICIENCY WAS OBSERVED WHILE PERFORMING THE INVESTIGATION FOR THE DEVICE; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18065985 ↗
  • MalfunctionSep 11, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, THE FULL RADIUS BLADE HAD WEAR AND REMNANTS OF FOREIGN BODIES OF METAL FROM THE BLADE, AND SMALL DEBRIS IN THE JOINT. THE SURGEON LOCKED AT THE TIP OF THE BLADE AND NOTICED WEAR ON THE INSIDE OF THE BLADE. THE PROCEDURE WAS FINISHED WITH THE SAME DEVICE. NO DELAY OR FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17725356 ↗
  • MalfunctionMay 3, 2023Break; Crack; Material Twisted/Bent; Mechanical Jam; Break; Crack; Material Twisted/Bent; Mechanical Jam

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE FULL RADIUS BLADE STALLED AND THEN BROKE. NO BROKEN PIECES FELL INTO THE PATIENT'S JOINT. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16861730 ↗
  • InjuryMar 11, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING SURGERY OF HIP ARTHROSCOPY USING FULL RADIUS BL CONCAVE EP-1 (BX 3), THE BLADE INNER CORE WAS FOUND WEAR AND CHIP. THE PROCEDURE FINISHED WITH A SMITH AND NEPHEW BACKUP DEVICE. SURGICAL DELAY LESS THAN OR EQUAL TO 30MINS. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11467876 ↗
  • MalfunctionJul 4, 2019Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT DURING PROCEDURE, IT WAS FOUND THAT LABEL STATES ITEM IS A CONCAVE, BUT IT IS ACTUALLY A CONVEX BLADE. THE DEVICE WAS USED ON THE PATIENT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY AND NO PATIENT INJURIES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8762800 ↗
  • MalfunctionJun 10, 2019No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY THE DEVICE HAD THE RIGHT LABELLING OUTSIDE THE BOX AND PACKAGE BUT WHEN THE DEVICE WAS OPENED, THE MATERIAL WAS NOT THE ONE THAT THEY USE. IT WAS A DIFFERENT PART OF WHAT THE CUSTOMER ORDERED. THE DEVICE WAS USED BY ERROR. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE SUCCESSFULLY. DELAY OR PATIENT INJURIES WERE NOT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8683107 ↗
  • MalfunctionMay 30, 2019Improper or Incorrect Procedure or Method; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT DURING A SURGERY, THE DEVICE HAD THE RIGHT LABELLING OUTSIDE THE BOX AND PACKAGE BUT WHEN THE DEVICE WAS OPENED, THE MATERIAL WAS NOT THE ONES THAT THEY USE. IT WAS A DIFFERENT PART OF WHAT THE CUSTOMER ORDERED. THE DEVICE WAS USED BY ERROR. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE SUCCESSFULLY. DELAY OR PATIENT INJURIES WERE NOT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8652638 ↗
  • MalfunctionMay 15, 2018Break; Failure to Cut

    IT WAS REPORTED THAT THE BLADE WAS NOT CUTTING TISSUE, IT WAS REMOVED FROM THE PATIENT TO CHECK WHAT HAPPENS, AND THEN, IT WAS OBSERVED THAT INTERNAL PART OF THE BLADE WAS BROKEN. THE BLADE BROKE INSIDE THE PATIENT, ALL PIECES WERE REMOVED. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7516671 ↗

FDA clearance and classification

The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 7205335

Has the NA 7205335 been recalled?

No FDA recall record in our data names this exact device identifier (DI 03596010251626), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 7205335?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between May 15, 2018 and Jul 30, 2024. In those reports, the most common specific device problems are “Break” (3), “Corroded” (1) and “Crack” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 7205335?

None of the reports are classified as deaths, 1 as an injury and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 7205335 FDA approved?

Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.

Who makes the NA 7205335, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 7205335). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page