NA 7205345Recalls & reported problems
Smith & Nephew, Inc. · 4.5 MM INCISOR PLUS STRAIGHT BLADES, POWER/EP-1, VIOLET, PACKAGED 6 PER BOX, STERILE · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 03596010251688), received between Feb 8, 2018 and Mar 8, 2026 — classified as 1 injury, 18 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Flaked6
- Material Fragmentation3
- Material Twisted/Bent2
- Delivered as Unsterile Product1
- Fracture1
- Insufficient Device Problem Information1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Foreign Body In Patient3
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 8, 2026Break
IT WAS REPORTED THAT, DURING AN LCP PROCEDURE, AN INCISOR SHAVER BLADE BROKE WHEN SURGEON BEGAN TO SHAVE THE PATIENT'S CARTILAGE, AND IT BECAME VERY HARD. NO PIECES OF THE DEVICE FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED WITHOUT ANY DELAY, USING A SIMILAR S+N BACK UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24542185 ↗MalfunctionJan 9, 2026Break; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING A KNEE ARTHROSCOPY, THE INNER PART OF ONE (1) BOXED INCISOR BLADE BROKE, PREVENTING IT FROM OSCILLATING AND BECOMING COMPLETELY STUCK. THE BLADE WAS TAKEN OUT OF THE PATIENT INTACT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24023693 ↗MalfunctionDec 23, 2025Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ARTHROSCOPY AND BILATERAL OSTEOCHONDROPLASTY PROCEDURE, WHILE USING THE INCISOR PLUS BLADE ON A KNEE WITH SIGNIFICANT FIBROSIS, THE BLADE BECAME BENT AND BROKE WHEN IT WAS STRAIGHTENED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23877627 ↗MalfunctionSep 8, 2025Positioning Failure; Material Deformation; Positioning Failure; Material Deformation
IT WAS REPORTED THAT, DURING AN UNSPECIFIED PROCEDURE, THE SEAL OF (B)(4) UNITS OF THE BOXED INCISOR PLUS BL WAS TURNED AND COULDN¿T BE REPLACED, THEREFORE, THEY NO LONGER FIT INTO THE SHAVER. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE INSTEAD. THERE WAS A DELAY GREATER THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22992145 ↗MalfunctionAug 25, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING AN ANKLE ARTHROSCOPY, THE TIP OR END OF THE INCISOR BLADE SNAPPED OFF. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. IT IS UNKNOWN HOW WAS THE SURGERY COMPLETED. THERE WAS A DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22883947 ↗MalfunctionAug 21, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANKLE ARTHROSCOPY, THE TIP END OF ONE (1) INCISOR PLUS BLADE SNAPPED OFF. THE BROKEN PIECES WERE SUCCESSFULLY REMOVED FROM THE PATIENT. IT IS UNKNOWN IF THERE WAS A DELAY. THE PROCEDURE WAS RESUMED, USING A S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22861020 ↗MalfunctionFeb 11, 2025Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, THE INCISOR PLUS BLADE BROKE BECAUSE OF RIGID FIBROSIS IN THE PATIENT'S KNEE. THE PIECES WERE REMOVED FROM INSIDE THE PATIENT WITH TWEEZERS. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21354689 ↗MalfunctionJan 14, 2025Flaked; Flaked
IT WAS REPORTED THAT DURING A SYNOVECTOMY OPERATION, WHILE THE INSTRUMENT WAS INSIDE THE PATIENT, WHEN ONE OF THE BOXED INCISOR PLUS WAS BEING USED, THE SURGEON SUDDENLY OBSERVED METALLIC POWDER ON THE SCREEN AND IMMEDIATELY SUCTION FOR CLEANING, THE METALLIC POWDER WAS REMOVED FROM THE PATIENT WITH SUCTION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21148183 ↗MalfunctionNov 20, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THERE WERE METAL PIECES COMING OFF FROM THE INCISOR PLUS BLADE INTO THE PATIENT'S KNEE. IT IS UNKNOWN IF THERE WAS A BACK UP DEVICE AVAILABLE OR IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20734996 ↗MalfunctionOct 20, 2024Break; Delivered as Unsterile Product; Material Split, Cut or Torn; Break; Delivered as Unsterile Product; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE INCISOR PLUS PACKAGING WAS CRUSHED AND BROKEN. THE SURGEON MENTIONED THAT IT WAS NO LONGER STERILE. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20490356 ↗MalfunctionAug 6, 2024Material Fragmentation; Suction Problem; Material Fragmentation; Suction Problem
IT WAS REPORTED THAT DURING AN ELBOW DEBRIDEMENT, THE BLADE PRODUCED SLIGHT IRON FILINGS AND SOON BECAME CLOGGED. AFTER UNSUCCESSFUL ATTEMPTS AT CLEANING AND DREDGING, A DOUBLE-TOOTH BLADES WERE REPLACED AND USED AS A BACK UP DEVICE TO COMPLETE THE OPERATION. THE SHRED FRAGMENTS WERE REMOVED FROM THE PATIENT USING SUCTION. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19922342 ↗MalfunctionApr 16, 2024Flaked; Material Integrity Problem; Mechanical Jam; Insufficient Device Problem Information; Flaked; Material Integrity Problem; Mechanical Jam; Insufficient Device Problem Information
IT WAS REPORTED THAT THREE (3) INCISOR PLUS BL FROM THE SAME PACKAGE WERE NO LONGER FUNCTIONAL AFTER A SHORT TIME. AFTER THE BLADE WAS CONNECTED TO THE HANDPIECE IT DID NOT WORK. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY OR IF THERE WAS PATIENT INVOLVEMENT; HOWEVER, THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES, USING A SMITH AND NEPHEW BACKUP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19121919 ↗InjuryDec 11, 2020Break; Break
IT WAS REPORTED THAT DURING A SURGERY, WHEN THE INCISOR BIADE ENTERED INTO THE PATIENT BODY, SOMETHING BRIGHT FELL OFF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10991263 ↗MalfunctionNov 26, 2019Flaked; Flaked
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY AND MENISECTOMY WHEN THE SURGEON WAS USING THE 4.5MM BLADE, METAL SHAVINGS WERE OBSERVED IN THE KNEE; THESE WERE REMOVED BY WASHING THEM OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR PATIENT INJURIES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 9373966 ↗MalfunctionNov 26, 2019Flaked; Flaked
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY AND MENISECTOMY, WHEN THE SURGEON WAS USING THE 4.5MM BLADE, METAL SHAVINGS WERE OBSERVED IN THE KNEE; THESE WERE REMOVED BY WASHING THEM OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR PATIENT INJURIES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 9373957 ↗MalfunctionAug 10, 2019Flaked; Flaked
IT WAS REPORTED THAT DURING A SHOULDER CUFF REPAIR SURGERY, IT WAS FOUND THAT THE PLANER HEAD HAD DEBRIS FALLING OFF. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8881555 ↗MalfunctionJun 17, 2019Flaked; Flaked
IT WAS REPORTED THAT DURING KNEE ARTHROSCOPY, FRAGMENTS OF THE SHAVER WERE LEFT WITHIN THE PATIENT WHICH WERE UNABLE TO BE REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 8705883 ↗MalfunctionAug 31, 2018Material Fragmentation
IT WAS REPORTED THAT DURING THE PROCEDURE THE TIP RELEASED FRAGMENTS. THE FRAGMENTS WERE REMOVED WITH THE CLAMP TWEEZER. NO PATIENT INJURY WAS REPORTED. BACK-UP DEVICE WAS USED TO COMPLETE THE SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7839633 ↗MalfunctionFeb 8, 2018Break
IT WAS REPORTED THAT THE DSAVER BLADE BROKEN IN ROTATION IN THE JOINT DURING ASK. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7251892 ↗
FDA clearance and classification
The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 7205345
Has the NA 7205345 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010251688), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 7205345?
The FDA's MAUDE database lists 19 adverse event reports for this device, received between Feb 8, 2018 and Mar 8, 2026. In those reports, the most common specific device problems are “Break” (8), “Flaked” (6) and “Material Fragmentation” (3); the most common patient problem is “Foreign Body In Patient” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 7205345?
None of the reports are classified as deaths, 1 as an injury and 18 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 7205345 FDA approved?
Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.
Who makes the NA 7205345, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 7205345). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SAW, POWERED, AND ACCESSORIES devices
- DYONICS POWERMAX ELITE 72200616Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES4,288 reports
- NA 72200616SSmith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES486 reports
- NA 72201500Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES136 reports
- PLATINUM 72205321Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES117 reports
- Zimmer Biomet™ 89-8509-450-60Zimmer Surgical SA · Saw, Powered, And Accessories83 reports
- NA 72200617Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES77 reports
- NA 72202530Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES53 reports
- DYONICS INCISOR 72203013Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES46 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page