NA 7210976Recalls & reported problems
Smith & Nephew, Inc. · 4.5 MM INCISOR PLUS ELITE STRAIGHT DISPOSABLE BLADES, SLATE,PACKAGED 6 PER BOX, STERILE · Class 1
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 03596010532626), received between Jan 11, 2018 and Jan 23, 2026 — classified as 4 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Flaked3
- Material Opacification2
- Material Separation2
- Delivered as Unsterile Product1
- Detachment of Device or Device Component1
- Material Deformation1
- Material Disintegration1
Patient problems
- No Known Impact Or Consequence To Patient5
- Foreign Body In Patient4
- No Clinical Signs, Symptoms or Conditions3
- No Information2
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryJan 23, 2026Flaked
IT WAS REPORTED THAT DURING AN ANKLE JOINT PROCEDURE, THE INCISOR PLUS ELITE 4.5 MM BLADE DISCHARGED A LARGE AMOUNT OF METAL DEBRIS INTO THE JOINT CAVITY. REPEATED IRRIGATION WAS INEFFECTIVE IN REMOVING THE DEBRIS, WHICH REMAINED INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED USING THE SAME DEVICE, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 24161512 ↗MalfunctionMay 20, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE INCISOR PLUS ELITE BLADE CAME DAMAGED BY THE SHIPPER. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22057108 ↗MalfunctionMay 20, 2024Break; Detachment of Device or Device Component; Material Deformation; Break; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY OF THE LEFT ANKLE, THE INCISOR ELITE BLADE LOST ITS SECOND JAW IN THE PATIENT'S JOINT. THE JAW WAS RECOVERED BUT THE OPERATION WAS LENGTHENED, PROLONGED ANESTHESIA TIME. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED OR THE EXTENT OF THE DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355439 ↗InjuryOct 26, 2023Break; Break
IT WAS REPORTED THAT DURING A CUFF/SHOULDER PROCEDURE, THE INCISOR BLADE THORNS BROKE AND WERE LEFT INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NON-SIGNIFICANT DELAY USING A S+N BACK UP DEVICE NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 18016389 ↗InjuryOct 6, 2023Break; Flaked; Break; Flaked
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, SMALL PIECES OF THE INCISOR PLUS ELITE WENT INTO THE PATIENT´S KNEE. THE VERY SMALL FRAGMENTS WERE LEFT INSIDE THE BODY. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A NON-SIGNIFICANT SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 17890106 ↗InjuryFeb 20, 2023Flaked; Flaked
IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE, THE SURGEON WAS COMPLAINING ABOUT METAL DUST/DEBRIS COMING FROM THE INCISOR ELITE BLADE. THE JOINT WAS WASH OUT AND ALL THE DEBRIS WAS REMOVED. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16405615 ↗MalfunctionNov 5, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY PROCEDURE, THE BLADE WAS USED AND MADE LOTS OF METAL DEBRIS IN THE JOINT. PIECES WERE SUCCESSFULLY REMOVED FROM THE PATIENT WITH GRASPER AND SHAVER SUCTION. BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO SIGNIFICANT DELAY AND NO PATIENT COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10797368 ↗MalfunctionNov 5, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY PROCEDURE, THE BLADE WAS USED AND MADE LOTS OF METAL DEBRIS IN THE JOINT. PIECES WERE SUCCESSFULLY REMOVED FROM THE PATIENT WITH GRASPER AND SHAVER SUCTION. BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO SIGNIFICANT DELAY AND NO PATIENT COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10797287 ↗MalfunctionJul 30, 2019Material Opacification; Material Opacification
IT WAS REPORTED THAT, DURING AN ACL PLASTY SURGERY, WHILE USING THE MOTOR DRIVE UNIT TO DEBRIDE THE NOTCH, IT WAS NOTICED THAT THE 4.5MM INCISOR PLUS ELITE BLADE WAS RELEASING METALLIC FLAKE-SHAPED PARTICLES IN THE JOINT. THE SURGEON ¿WASHED¿ A LOT TO MAKE THE FLAKES GOING OUT OF THE JOINT; THEN, HE USED AS WELL A GRASP TO GET OUT THE FLAKES THAT COULD NOT BE VACUUMED. A 3.5MM INCISOR PLUS ELITE WAS USED INSTEAD TO COMPLETE THE PROCEDURE. THE SURGERY WAS NOT SIGNIFICANTLY DELAYED. THE PATIENT WAS NOT INJURED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8843495 ↗MalfunctionJul 30, 2019Material Opacification; Material Opacification
IT WAS REPORTED THAT, DURING AN ACL PLASTY SURGERY, WHILE USING THE MOTOR DRIVE UNIT TO DEBRIDE THE NOTCH, IT WAS NOTICED THAT THE 4.5MM INCISOR PLUS ELITE BLADE WAS RELEASING METALLIC FLAKE-SHAPED PARTICLES IN THE JOINT. THE SURGEON ¿WASHED¿ A LOT TO MAKE THE FLAKES GOING OUT OF THE JOINT; THEN, HE USED AS WELL A GRASP TO GET OUT THE FLAKES THAT COULD NOT BE VACUUMED. A 3.5MM INCISOR PLUS ELITE WAS USED INSTEAD TO COMPLETE THE PROCEDURE. THE SURGERY WAS NOT SIGNIFICANTLY DELAYED. THE PATIENT WAS NOT INJURED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8843494 ↗MalfunctionJan 10, 2019Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE SHAVER BLADE TIP GOT BROKEN DURING PROCEDURE. IT IS UNKNOWN IF THERE WAS DELAY OR BACK UP AVAILABLE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8235475 ↗MalfunctionAug 22, 2018Material Disintegration
IT WAS REPORTED THAT 90 SECONDS AFTER STARTING MENISCAL RESECTION, METAL DEBRIS OCCURRED IN THE JOINT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7807067 ↗MalfunctionJan 30, 2018Metal Shedding Debris
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY AND AFTER FEW MINUTES OF USE SOME METALLIC PARTICLES CAME OFF FROM THE DEVICE. THEY MANAGED TO SUCK UP THE PARTICLES TO LEAVE THE SITE CLEAN. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7229476 ↗MalfunctionJan 11, 2018Material Fragmentation
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY AFTER A FEW MINUTES OF USE, SOME METALLIC PARTICLES CAME OFF FROM THE DEVICE. THEY MANAGED TO SUCK UP THE PARTICLES TO LEAVE THE SITE CLEAN. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 7182034 ↗
Other SAW, POWERED, AND ACCESSORIES devices
- DYONICS POWERMAX ELITE 72200616Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES4,290 reports
- NA 72200616SSmith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES486 reports
- NA 72201500Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES136 reports
- PLATINUM 72205321Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES118 reports
- Zimmer Biomet™ 89-8509-450-60Zimmer Surgical SA · Saw, Powered, And Accessories83 reports
- NA 72200617Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES77 reports
- NA 72202530Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES53 reports
- DYONICS INCISOR 72203013Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES46 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.