NA 72200724Recalls & reported problems
Smith & Nephew, Inc. · 4.0 MM ELITE ACROMIONIZER STRAIGHT DISPOSABLE BURRS, MAUVE, 10000 MAXIMUM RPM, PACKAGED 6 PER BOX, STERILE · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Acromionizer disposable burr is a powered surgical tool for shaping bone in arthroscopy. Reports mostly describe burrs making strange noises, overheating, melting or breaking during shoulder procedures, often delaying surgery while a backup device was used. There are 25 reports that name its identifier, with no deaths and 4 injuries mentioned. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 03596010578020), received between Nov 23, 2018 and Aug 3, 2023 — classified as 4 injuries, 21 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Material Fragmentation9
- Failure to Cut8
- Material Deformation6
- Overheating of Device3
- Noise, Audible2
- Break1
- Defective Device1
- Manufacturing, Packaging or Shipping Problem1
Patient problems
- No Known Impact Or Consequence To Patient15
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2023Material Deformation; Noise, Audible; Material Deformation; Noise, Audible
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY, FIVE (5) DYONICS ACROMIONIZER BURRS WERE MAKING A STRANGE NOISE AND THE TERMINAL FUSED. SURGERY WAS COMPLETED AFTER A SIGNIFICANT DELAY (>30 MIN) USING A BACK-UP DEVICE INSTEAD. NO FURTHER COMPLICATIONS WERE REPORTED. PATIENT'S STATUS IS OKAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17454747 ↗MalfunctionJul 12, 2023Melted; Melted
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, TWO (2) ACROMIONIZER BURR FAILED. THE ACROMIONIZER BURRS HAD A STRANGE NOISE, AFTER THAT THEY MELTED AND BROKE IN THE PART WHERE THEY ARE CONNECTED TO THE MDU. ONE DEVICE STOPPED BEFORE BREAKAGE, BUT IT IS UNKNOWN WHICH ONE. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY GREATER THAN 30 MINUTES USING THE SAME DEVICES. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17309730 ↗MalfunctionMar 6, 2023Noise, Audible; Noise, Audible
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, WHEN TWO (2) DYONICS BLADES WERE ATTACHED TO A SHAVER HANDPIECE AND ROTATED THEY MADE A STRANGE NOISE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A 45 MINUTES DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16484726 ↗MalfunctionMay 23, 2022Overheating of Device; Material Deformation; Overheating of Device; Material Deformation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY TWO BLADES MADE A STRANGE NOISE. IT OCCURRED AFTER ABOUT 10 SECONDS OF ROTATION. THE PROCEDURE WAS COMPLETED USING AN UNKNOWN DEVICE. IT IS UNKNOWN IF THERE WAS A DELAY. NO PATIENT COMPLICATIONS WERE REPORTED. OVERHEATING WAS REPORTED ON DEVICE WITH THE LOT NUMBER 50943904.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14470601 ↗MalfunctionAug 26, 2021Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, WHEN OPENING THE PACKAGE OF THE DYONICS ACROMIONIZER BURR IT WAS FOUND THAT THE STERILIZATION TRAY ACCIDENTALLY CONTAINED AN ABRADER BURR. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12370336 ↗InjuryJan 5, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A SHOULDER SURGERY THE DYONICS ACROMIONIZER BURR 4MM INNER AND OUTER SHEATHS MELDED TOGETHER. THE PROCEDURE WAS COMPLETED WITHOUT A SIGNIFICANT DELAY USING A 5.5 MM BURR WITH THE SAME HANDPIECE AS A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11120963 ↗InjuryJan 5, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A SHOULDER SURGERY THE DYONICS ACROMIONIZER BURR 4MM INNER AND OUTER SHEATHS MELDED TOGETHER. THE PROCEDURE WAS COMPLETED WITHOUT A SIGNIFICANT DELAY USING A 5.5 MM BURR WITH THE SAME HANDPIECE AS A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11120901 ↗InjuryJan 5, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A SHOULDER SURGERY THE DYONICS ACROMIONIZER BURR 4MM INNER AND OUTER SHEATHS MELDED TOGETHER. THE PROCEDURE WAS COMPLETED WITHOUT A SIGNIFICANT DELAY USING A 5.5 MM BURR WITH THE SAME HANDPIECE AS A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11120812 ↗InjuryJan 5, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A SHOULDER SURGERY THE DYONICS ACROMIONIZER BURR 4MM INNER AND OUTER SHEATHS MELDED TOGETHER. THE PROCEDURE WAS COMPLETED WITHOUT A SIGNIFICANT DELAY USING A 5.5 MM BURR WITH THE SAME HANDPIECE AS A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11120705 ↗MalfunctionOct 7, 2020Break; Overheating of Device; Break; Overheating of Device
IT WAS REPORTED THAT DURING A BONE BLOCK SURGERY, WHILE USING THE BURR AND HUB, THEY SNAPPED OFF INTO TWO PIECES AFTER BEING USED FOR UNDER 1 MINUTE, THE DEVICE DID NOT BREAK INSIDE THE PATIENT AND THERE WAS NO SHEDDING OR FLAKING DUE TO THIS MALFUNCTION. THE PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE WITH NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED. PRELIMINARY RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT OVERHEATED DUE TO AN INADEQUATE IRRIGATION FLOW WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10643606 ↗MalfunctionJun 24, 2020Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING ARTHROSCOPY, THE DYONICS BURR COULD NOT BE ROTATED WHEN THE DEVICE WAS USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A S+N BACK-UP. NO PATIENT INJURIES OR DELAY WERE REPORTED. BASED ON THE RESULTS OF INVESTIGATION THE BURR OVERHEATED AND HAD MELTED COMPONENTS THIS MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10193036 ↗MalfunctionDec 4, 2018Mechanical Problem; Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONECUTTER TO TAKE SOME OF THE BONE. THE BONECUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8129319 ↗MalfunctionDec 4, 2018Material Disintegration
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONECUTTER TO TAKE SOME OF THE BONE. THE BONECUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8129019 ↗MalfunctionDec 4, 2018Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONECUTTER TO TAKE SOME OF THE BONE. THE BONECUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8128897 ↗MalfunctionNov 23, 2018Material Fragmentation; Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100087 ↗MalfunctionNov 23, 2018Material Fragmentation
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100084 ↗MalfunctionNov 23, 2018Material Fragmentation; Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100082 ↗MalfunctionNov 23, 2018Material Fragmentation
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100081 ↗MalfunctionNov 23, 2018Peeled/Delaminated; Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100071 ↗MalfunctionNov 23, 2018Defective Device
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8100069 ↗MalfunctionNov 23, 2018Material Fragmentation; Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100065 ↗MalfunctionNov 23, 2018Material Fragmentation; Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100063 ↗MalfunctionNov 23, 2018Material Fragmentation; Failure to Cut
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100013 ↗MalfunctionNov 23, 2018Material Fragmentation
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100006 ↗MalfunctionNov 23, 2018Material Fragmentation
IT WAS REPORTED THAT IT TAKES A LOT OF EFFORT FOR THE BONE CUTTER TO TAKE SOME OF THE BONE. THE BONE CUTTER BLADE PRODUCED SMALL CHIPS OF METAL IN THE JOINT. THEY WERE REMOVED BY FLUSHING OUT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8100001 ↗
FDA clearance and classification
The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 72200724
Has the NA 72200724 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010578020), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 72200724?
The FDA's MAUDE database lists 25 adverse event reports for this device, received between Nov 23, 2018 and Aug 3, 2023. In those reports, the most common specific device problems are “Material Fragmentation” (9), “Failure to Cut” (8) and “Material Deformation” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 72200724?
None of the reports are classified as deaths, 4 as injuries and 21 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 72200724 FDA approved?
Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.
Who makes the NA 72200724, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 72200724). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SAW, POWERED, AND ACCESSORIES devices
- DYONICS POWERMAX ELITE 72200616Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES4,288 reports
- NA 72200616SSmith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES486 reports
- NA 72201500Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES136 reports
- PLATINUM 72205321Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES117 reports
- Zimmer Biomet™ 89-8509-450-60Zimmer Surgical SA · Saw, Powered, And Accessories83 reports
- NA 72200617Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES77 reports
- NA 72202530Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES53 reports
- DYONICS INCISOR 72203013Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES46 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page