Skip to content

NA 72200728Recalls & reported problems

Smith & Nephew, Inc. · 4.0 MM ELITE STONECUTTER STRAIGHT DISPOSABLE BURRS, MAROON, 10000 MAXIMUM RPM, PACKAGED 6 PER BOX, STERILE · Class 1

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010578068), received between Nov 16, 2018 and May 9, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device4
  • Flaked2
  • Material Separation2
  • Material Fragmentation1
  • Melted1
  • Noise, Audible1
  • Physical Resistance/Sticking1
  • Temperature Problem1

Patient problems

  • No Known Impact Or Consequence To Patient6
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2025Flaked; Flaked

    IT WAS REPORTED THAT DURING A PROCEDURE, ONE OF THE DYONICS BURR PRODUCT COLLAR SHREDDED WHEN PUT ON MAXIMUM POWER. THE PROCEDURE WAS COMPLETED WITHOUT THE PRODUCT AND AN UNKNOWN AMOUNT OF TIME OF SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21994158 ↗
  • MalfunctionMay 20, 2024Flaked; Flaked

    IT WAS REPORTED THAT DURING AN ANKLE ARTHROSCOPY, THE TIP OF THE DYONICS STONECUTTER FELL OFF. METAL SHAVINGS WERE NOTED IN THE PATIENT VIA ARTHROSCOPE. ALL METAL SHAVINGS WERE SUCTIONED OUT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19351930 ↗
  • MalfunctionMay 23, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DISTAL CLAVICLE EXCISION, THE DYONICS STONECUTTER BURR SHREDDED A BIT AND CREATED SOME METAL SHAVINGS IN THE SHOULDER. ALL THE METAL SHAVINGS WERE CLEANED UP WITH A SHAVER. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16988868 ↗
  • MalfunctionOct 22, 2019Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A PROCEDURE, THE BLADE MADE NOISE WHEN THE DEVICE IS ROTATING. BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO DELAY AND NO PATIENT INJURIES WERE REPORTED. RESULTS OF THE INVESTIGATION HAVE CONCLUDED THAT THIS UNIT OVERHEATED WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9218863 ↗
  • MalfunctionJun 19, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT AFTER CONNECTING THE BURR ON THE HANDPIECE, THE BURR STOPPED TURNING AND THE HANDPIECE SETS TO ALARM STATE. REMOVING THE BURR, PLASTICS HAD BEEN DAMAGED. THE SURGEON CHANGED THE HANDPIECE AND THE INCIDENT REPEATED AGAIN. THEN THE SURGEON USED ANOTHER BURR (OTHER PRODUCT REFERENCE WITH BIGGER DIAMETER AND IT WAS OK.) NO PLASTICS PIECES WERE LEFT INTO THE PATIENT. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8714163 ↗
  • MalfunctionJun 18, 2019Material Separation; Material Separation; Physical Resistance/Sticking

    IT WAS REPORTED THAT AFTER CONNECTING THE BURR ON THE HANDPIECE, THE BURR STOPPED TURNING AND THE HANDPIECE SETS TO ALARM STATE. REMOVING THE BURR, PLASTICS HAD BEEN DAMAGED. THE SURGEON CHANGED THE HANDPIECE AND THE INCIDENT REPEATED AGAIN. THEN THE SURGEON USED ANOTHER BURR (OTHER PRODUCT REFERENCE WITH BIGGER DIAMETER AND IT WAS OK.) NO PLASTICS PIECES WERE LEFT INTO THE PATIENT. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8709609 ↗
  • MalfunctionDec 4, 2018Melted; Overheating of Device; Noise, Audible

    IT WAS REPORTED THAT THE BLADE MADE A NOISE WHEN THE DEVICE IS ROTATING. NO PATIENT INJURY REPORTED. BACK-UP DEVICE WAS USED TO COMPLETE THE SURGERY. ACCORDING TO INVESTIGATION RESULTS, THE PRODUCT HAS BECOME OVER HEATED AND THE SLUFF CHAMBER AND ADAPTER BODY HAVE BEGUN TO MELT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8129876 ↗
  • MalfunctionNov 16, 2018Overheating of Device; Temperature Problem

    IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8080917 ↗
  • MalfunctionNov 16, 2018Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY AND NO PATIENT INJURIES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8080874 ↗

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.