Skip to content

NA 7210909Recalls & reported problems

Smith & Nephew, Inc. · 5.5 MM FULL RADIUS ELITE STRAIGHT DISPOSABLE BLADES, OLIVE,PACKAGED 6 PER BOX, STERILE · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03596010582737), received between Feb 14, 2019 and May 16, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Flaked” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Flaked3
  • Break2
  • Material Twisted/Bent2
  • Insufficient Device Problem Information1
  • Material Deformation1
  • Mechanical Jam1
  • Mechanical Problem1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 16, 2025Break; Material Deformation; Material Twisted/Bent; Scratched Material; Insufficient Device Problem Information; Physical Resistance/Sticking; Break; Material Deformation; Material Twisted/Bent; Scratched Material; Insufficient Device Problem Information; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A MENISCECTOMY, THE DISPOSABLE BLADE STOPPED WORKING, AND WHEN IT WAS REMOVED FROM THE HANDPIECE, THE TIP WAS GETTING STUCK WHEN TURNING. THE PROCEDURE WAS COMPLETED WITHOUT ANY SURGICAL DELAY USING A S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22039725 ↗
  • MalfunctionMar 19, 2025Break; Break

    IT WAS REPORTED THAT DURING A CUFF PROCEDURE, THE SURGEON BROKE THE FULL RADIOS BLADE AND CLAIMED IT WAS OF POOR QUALITY. THE PIECES WERE REMOVED. IT IS UNKNOWN IF THERE WAS A BACK UP DEVICE AVAILABLE. NO DELAY WAS REPORTED AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21649010 ↗
  • MalfunctionSep 30, 2024Flaked; Flaked

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, METAL DEBRIS WERE DEPOSITED IN THE KNEE DURING USE OF THE KNIFE BLADE 5.5 FULL RADIUS ELITE DISPOSABLE. THE DEBRIS WAS RETRIEVED BY SUCTION. IT IS UNKNOWN IF THERE WAS A BACK-UP DEVICE AVAILABLE OR IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20339093 ↗
  • MalfunctionAug 5, 2024Flaked; Material Twisted/Bent; Mechanical Jam; Flaked; Material Twisted/Bent; Mechanical Jam

    IT WAS REPORTED THAT DURING SHOULDER ARTHROSCOPY, THE USE OF THE BLADE 5.5 FULL RADIUS LOCKED THE HANDPIECE AND WHITE POWDER CAME OUT WHEN THE BLADE WAS FORCE TO BE USED. IT BROKE SEVERAL TIMES DURING THE SURGERY. THE PROCEDURE WAS A COMPLETED WITHOUT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19906928 ↗
  • MalfunctionDec 18, 2023Flaked; Flaked

    IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPY, THE 5.5 MM BLADE ELITE HAD A STRONG METAL ABRASION. THE ROOT CAUSE OF THE FAILURE IS UNCLEAR. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18348770 ↗
  • MalfunctionFeb 14, 2019Mechanical Problem

    IT WAS REPORTED THAT THE SHAVER WAS INSERTED INTO THE PATIENT AND IT DOES NOT MOVE. WHEN FINALLY MOVES, THE SHAVER TEETH WERE PUSHED OUTSIDE. . THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8342077 ↗

FDA clearance and classification

The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 7210909

Has the NA 7210909 been recalled?

No FDA recall record in our data names this exact device identifier (DI 03596010582737), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 7210909?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 14, 2019 and May 16, 2025. In those reports, the most common specific device problems are “Flaked” (3), “Break” (2) and “Material Twisted/Bent” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 7210909?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 7210909 FDA approved?

Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.

Who makes the NA 7210909, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 7210909). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page