NA 7210980Recalls & reported problems

Smith & Nephew, Inc. · 3.5 MM CONCAVE, FULL RADIUS, POWER/EP-1, 17 MM LENGTH, BEIGE, PACKAGED 6 PER BOX, STERILE · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 03596010595775), received between May 30, 2018 and Sep 3, 2024 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Break” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 5 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break11
  • Detachment of Device or Device Component2
  • Fracture2
  • Mechanical Jam2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2024Break; Break

    IT WAS REPORTED THAT, DURING A PROCEDURE, A PIECE OF THE FULL RADIUS BLADE BROKE OFF INTO THE PATIENT. THE PIECE WAS RETRIEVED FROM THE PATIENT WITHOUT INCIDENT. THERE WAS A DELAY OF LESS THAN 30 MINUTES REPORTED, HOWEVER, IT IS UNKNOWN IF THERE WAS A BACK-UP DEVICE AVAILABLE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20131548 ↗
  • MalfunctionSep 2, 2024Fracture; Mechanical Jam; Fracture; Mechanical Jam

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY MEDIAL MENISCECTOMY, THE INNER LUMEN OF TWO (2) CURVED FULL RADIUS DID NOT ROTATE IN OSCILLATION, FORWARD OR REVERSE. THE PROCEDURE WAS COMPLETED USING A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A 2-MIN DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20122255 ↗
  • MalfunctionAug 26, 2024Fracture; Mechanical Jam; Fracture; Mechanical Jam

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY MEDIAL MENISCECTOMY, THE INNER LUMEN OF THE CURVED FULL RADIUS BLADE DID NOT ROTATE IN OSCILLATION, FORWARD OR REVERSE. THE PROCEDURE WAS COMPLETED USING A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A 2-MIN SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20064346 ↗
  • MalfunctionMay 31, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, A BLADE INNER PART DETACHED FROM THE TIP AND IT STOPPED TURNING, NOTHING FELL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING A S+N BACK UP DEVICE. THERE WAS A DELAY OF 2-3 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19439278 ↗
  • MalfunctionMay 31, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, TWO (2) BLADE INNER PART DETACHED FROM THE TIP AND IT STOPPED TURNING, NOTHING FELL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING A S+N BACK UP DEVICE. THERE WAS A DELAY OF 2-3 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19439181 ↗
  • MalfunctionApr 23, 2024Break; Break

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, THE SHAVER GOT CLOGGED, TECH WENT TO UNCLOG IT AND NOTICED THE TIP ON THE INSIDE OF THE FULL RADIUS BLADE HAD BROKEN OFF INSIDE THE SHAVER SHEET. NO PIECES WERE LEFT IN THE PATIENT, THEY WERE NOT IN THE PATIENT FREELY. ANOTHER BLADE WAS OPENED AND THE SAME THING HAPPENED. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19168676 ↗
  • MalfunctionMar 26, 2024Break; Break

    IT WAS REPORTED THAT THE 3.5 CURVED BLADE BROKE. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT AND IF THERE WAS A BACK-UP DEVICE AVAILABLE OR IF THERE WAS A SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18982353 ↗
  • MalfunctionFeb 16, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR ARTHROSCOPY, THE INNER LUMEN OF THE 3.5 CURVED BLADE BROKE WHILE OUTSIDE THE JOINT SPACE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A DELAY OF 2 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18721576 ↗
  • MalfunctionFeb 16, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR ARTHROSCOPY, THE INNER LUMEN OF THE 3.5 CURVED BLADE BROKE WHILE OUTSIDE THE JOINT SPACE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A DELAY OF 2 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18721553 ↗
  • MalfunctionJan 30, 2024Break; Break

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY TWO (2) BLADE CONCAVE FULL RADIUS BROKE AT THE CURVE. INSIDE THE PATIENT, THERE WAS NO DEBRIS IN THE KNEE. THEY WERE OSCILLATING FINE AND THE TIP JUST STOPPED BECAUSE THE INNER LUMEN WAS BROKEN. THE INNER-LUMEN OF THE SHAVERS BROKE AT THE BEND OF THE SHAVERS, THE OUTER LUMEN WERE INTACT. NO METAL SHAVINGS IN THE KNEES. IT IS UNKNOWN IF THERE WAS A BACK UP DEVICE AVAILABLE OR IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18610368 ↗
  • MalfunctionJan 24, 2024Break; Break

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY A BLADE CONCAVE FULL RADIUS BROKE AT THE CURVE, INSIDE THE PATIENT, THERE WAS NO DEBRIS IN THE KNEE. IT WAS OSCILLATING FINE AND THE TIP JUST STOPPED BECAUSE THE INNER LUMEN WAS BROKEN. THE INNER-LUMEN OF THE SHAVER BROKE AT THE BEND OF THE SHAVER, THE OUTER LUMEN WAS INTACT. NO METAL SHAVINGS IN THE KNEES. IT IS UNKNOWN IF THERE WAS A BACK UP DEVICE AVAILABLE OR IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18572547 ↗
  • MalfunctionJan 19, 2024Break; Break

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY TWO (2) BLADE CONCAVE FULL RADIUS BROKE AT THE CURVE, INSIDE THE PATIENT. THEY WERE OSCILLATING FINE AND THE TIP JUST STOPPED BECAUSE THE INNER LUMEN WAS BROKEN. THE INNER-LUMEN OF THE SHAVERS BROKE AT THE BEND, THE OUTER LUMEN WERE INTACT. NO METAL SHAVINGS IN THE KNEES. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18544458 ↗
  • MalfunctionJan 17, 2024Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TIP OF THE BLADE BROKE AND FELL OF INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18526352 ↗
  • MalfunctionJan 11, 2024Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BLADE TIP BROKE. THE PROCEDURE WAS COMPLETED USING A S+N BACK UP DEVICE. THERE WAS NO SURGICAL DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18499085 ↗
  • MalfunctionMay 30, 2018Break

    IT WAS REPORTED THAT THE TOP OF THE BLADE WAS CHIPPED OFF WHILST PERFORMING A KNEE ARTHROSCOPY. NO PATIENT INJURY OR SIGNIFICANT TIME DELAY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7555069 ↗

FDA clearance and classification

The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 7210980

Has the NA 7210980 been recalled?

No FDA recall record in our data names this exact device identifier (DI 03596010595775), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 7210980?

The FDA's MAUDE database lists 15 adverse event reports for this device, received between May 30, 2018 and Sep 3, 2024. In those reports, the most common specific device problems are “Break” (11), “Detachment of Device or Device Component” (2) and “Fracture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 7210980?

None of the reports are classified as deaths, 0 as injuries and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 7210980 FDA approved?

Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.

Who makes the NA 7210980, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 7210980). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →