ENDOBUTTON 72200147Recalls & reported problems
Smith & Nephew, Inc. · ENDOBUTTON CL ULTRA 20MM · Class 2 · Implant
All models of ENDOBUTTON (21)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 03596010597267), received between Aug 8, 2017 and Mar 26, 2026 — classified as 5 injuries, 15 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Adverse Event Without Identified Device or Use Problem3
- Component Missing3
- Delivered as Unsterile Product1
- Device Damaged Prior to Receipt by User1
- Device Damaged Prior to Use1
- Device Dislodged or Dislocated1
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Known Impact Or Consequence To Patient5
- Device Embedded In Tissue or Plaque1
- Fever1
- Foreign Body In Patient1
- Impaired Healing1
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2026Break
IT WAS REPORTED THAT, DURING A PROCEDURE, THE ENDOBUTTON BROKE. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT SN BACK UP. IT IS UNKNOWN IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24696123 ↗MalfunctionAug 18, 2025Component Missing; Component Missing
IT WAS REPORTED THAT DURING A PROCEDURE THE ENDOBUTTON ULTRA 20MM BOX WAS OPENED AND THERE WAS NO IMPLANT INSIDE THE BOX, CONTAINING ONLY STICKERS AND A BOOKLET, THE PROCEDURE WAS COMPLETED USING A SMITH AND NEPHEW BACK-UP DEVICE, THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22825010 ↗MalfunctionAug 11, 2025Component Missing; Component Missing
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHEN THE BOX OF THE 20MM ENDOBUTTON ULTRA WAS OPENED, IT WAS NOTICED THAT THERE WAS NO IMPLANT INSIDE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22770769 ↗MalfunctionAug 11, 2025Component Missing; Component Missing
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHEN THE BOX OF THE 20MM ENDOBUTTON ULTRA WAS OPENED, IT WAS NOTICED THAT THERE WAS NO IMPLANT INSIDE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22770688 ↗InjuryMar 5, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE ULTRA ENDOBUTTON TRACTION SUTURE BROKE, THE SUTURE WAS LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETED USING A S+N BACK UP DEVICE. THERE WAS A DELAY OF LESS THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 21526010 ↗InjuryJan 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, SIX WEEKS AFTER A LEFT KNEE ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY + ARTHROSCOPIC MENISCUS SHAPING SURGERY + ARTHROSCOPIC KNEE SYNOVECTOMY SURGERY + TENDON EXCISION SURGERY + JOINT THERAPEUTIC SUBSTANCE INJECTION, WHICH WAS PERFORMED ON (B)(6) 2024 AND IN WHICH A 20 MM ENDOBUTTON CL ULTRA WAS INSERTED, THE PATIENT HAD POOR WOUND HEALING. ON (B)(6) 2024, UNDER GENERAL ANESTHESIA, A DEBRIDEMENT AND VSD DRAINAGE WERE PERFORMED FOR POOR WOUND HEALING AFTER THE PRIMARY SURGERY OF THE LEFT ANTERIOR CRUCIATE LIGAMENT. ON (B)(6) 2024, APPROXIMATELY, THE PATIENT EXPERIENCED RECURRENT FEVER WITH A MAXIMUM TEMPERATURE OF 38.5 DEGREES CELSIUS AND THE BLOOD TEST RESULTS SHOWED 80.5% NEUTROPHILS AND 141.89MG/L C-REACTIVE PROTEIN. THE PATIENT WAS TREATED WITH VANCOMYCIN 1.0G Q12H. ON (B)(6) 2024, UNDER GENERAL ANESTHESIA, THE PATIENT UNDERWENT ARTHROSCOPIC CLEANING OF THE LEFT KNEE SURGERY WITH POOR WOUND HEALING. DURING THIS REVISION SURGERY, PURULENT FLUID ACCUMULATION AND SYNOVIAL INFECTION WERE OBSERVED IN THE KNEE JOINT, AND THE ANTERIOR CRUCIATE LIGAMENT GRAFT FAILED. KNEE JOINT SYNOVIAL CLEANING WAS PERFORMED, AND A DRAINAGE TUBE WAS PLACED. THE ANCHOR REMA
Patient problems: Fever; Impaired Healing; Synovitis
MAUDE report 21134449 ↗MalfunctionMar 21, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE STERILE FOIL OF THE ENDOBUTTON WAS NOT WELL AFFIXED, THE IMPLANT WAS NON-STERILE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY. IT UNKNOWN IF THERE WAS A SMITH AND NEPHEW BACK UP DEVICE AVAILABLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18952177 ↗MalfunctionFeb 19, 2024Packaging Problem; Packaging Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, WHEN OPENING THE PACKAGING OF THE ENDOBUTTON ULTRA, THE NURSE NOTICED THAT THE PACKAGING HAD A HOLE IN IT. THE SAME PROBLEM WAS FOUND ON A SECOND PACKAGING. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18731387 ↗MalfunctionMay 3, 2023Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT DURING SET UP FOR AN ACL RECONSTRUCTION SURGERY, ONCE THE PACKAGE OF THE ENDOBUTTON DEVICE WAS OPENED, IT WAS FOUND THE ASEPTIC POUCH WAS DAMAGED. A BACK UP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16863891 ↗InjuryNov 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON REGISTRY REVIEW 2021 NEW ZEALAND ACL REGISTRY IMPLANT REPORT, AN UNKNOWN AMOUNT OF PATIENTS REQUIRED A REVISION SURGERY AFTER AN ACL PROCEDURE USING ENDOBUTTON DEVICES. PATIENTS OUTCOME IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 15848647 ↗MalfunctionAug 24, 2022Device Damaged Prior to Receipt by User; Device Damaged Prior to Receipt by User
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHEN THE CARTON OF THE ENDOBUTTON WAS OPENED, IT WAS FOUND THAT THE STERILE PACKAGE HAD BEEN DAMAGED AND COULD NOT BE USED. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS NO DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15286688 ↗InjuryJul 27, 2022Break; Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION SURGERY USING ENDOBUTTON, THE WHITE SUTURE BROKE UNDER THE STRAIN, THE CLOSED-LOOP OF THE ENDOBUTTON WAS CUT BECAUSE THE PLATES COULD NOT BE REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SURGICAL DELAY OF 30 MINUTES USING A BACK-UP DEVICE. THE PLATE WAS REMOVED FROM THE PATIENT AND AN ADDITIONAL BONE HOLE WAS REQUIRED TO COMPLETE THE SURGERY. NO FURTHER COMPLICATIONS WERE REPORTED. CURRENT HEALTH STATUS IS GOOD, PATIENT HAS BEEN DISCHARGED.
Patient problems: Perforation; Rupture
MAUDE report 15117867 ↗MalfunctionNov 11, 2020Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION, WHEN OPENING THE BOX OF THE ENDOBUTTON, IT WAS NOTICED THAT THERE WAS NOT IMPLANT OR STICKY LABELS IN THE BOX. ONLY THE INSTRUCTIONS FOR USE WAS IN THE BOX. THE PROCEDURE WAS COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10825484 ↗MalfunctionJan 13, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A PCL PROCEDURE, WHEN PULLING THE GRAFT, LOOP OF THE ENDOBUTTON CL GOT RUPTURED. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. A DELAY GREATER THAN 30 MINUTES WAS REPORTED. NO PATIENT INJURIES WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9581000 ↗InjuryJun 14, 2019Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A KNEE PROCEDURE, PATIENT EXPERIENCED A PAIN, SWELLING AND MICRO BLOODY EXUDATION IN THE OPERATION AREA. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY OR IF A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. PATIENT WAS REPORTED TO BE IN GOOD CONDITION.
Patient problems: Pain; Swelling
MAUDE report 8701050 ↗MalfunctionJan 3, 2019Break
IT WAS REPORTED THAT DURING THE SURGERY WHILE IMPLANTING THE TUNNEL, THE CLUSTER OF THE LOOP BROKE. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8214553 ↗MalfunctionOct 25, 2018Expiration Date Error
IT WAS REPORTED THAT AN EXPIRED ENDOBUTTON CL WAS IMPLANTED ON THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8005199 ↗MalfunctionMay 28, 2018Material Frayed
IT WAS REPORTED THAT DURING PROCEDURE THE SURGEON FELT THAT THE CONDITION OF GRAFT WAS NOT GOOD COMPARED WITH USUAL. WHEN THE GRAFT WAS PULLED UP INTO PREPARED BONE SITE, ENDOBUTTON CL WAS CHECKED OUTSIDE FEMORAL BONE SITE AND IT WAS FRAYED. THEN GRAFT WAS PULLED UP FOR FIXING. FOR FINAL CHECK OF FIXING, WHEN SUTURE WAS PULLED UP, GRAFT WAS CUT AND CAME OUT FROM BONE SITE. THEREFORE, ENDOBUTTON CL WAS GOT BACK AND FOUND THAT PART OF CL DAMAGED AND CUT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7547796 ↗MalfunctionMay 20, 2018Break
IT WAS REPORTED THAT THE SUTURE BROKE WHEN IT WAS PULLED THROUGH THE FEMORAL TUNNEL. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7528650 ↗MalfunctionAug 8, 2017Break; Difficult to Remove
IT WAS REPORTED THAT THE GUIDE WIRE BROKE DURING THE PROCEDURE. THIS WAS NOT NOTICED UNTIL THE END OF THE PROCEDURE. SURGERY TIME WAS EXTENDED IN AN UNSUCCESSFUL ATTEMPT TO RETRIEVE THE BROKEN PIECE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6774126 ↗
Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,686 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,398 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE932 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE721 reports
- RIGIDLOOP 232447Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE633 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue624 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE609 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE460 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page