Skip to content

NA 72201519Recalls & reported problems

Smith & Nephew, Inc. · DYONICS 3.5 MM ABRADER BURRS, 7 CM LENGTH, AQUA, PACKAGED 6 PER BOX, STERILE · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03596010621641), received between Aug 15, 2018 and Nov 17, 2025 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component2
  • Device Fell1
  • Material Deformation1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 17, 2025Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT, DURING AN ARTHRODESIS OF THE ANKLE PROCEDURE, THE ABRADER BLADE WAS BROKEN. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT'S BODY. THE PROCEDURE WAS RESUMED AFTER A 10-MINUTE SURGICAL DELAY. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED, AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23569485 ↗
  • MalfunctionOct 23, 2024Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TIP OF THE INNER BLADE OF THE ABRADER 3.5 WAS BROKEN. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT'S BODY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF 15 MINUTES. IT IS UNKNOWN IF THERE WAS A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20517177 ↗
  • InjuryOct 11, 2022Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE INNER BLADE OF THE 3.5MM ABRADER BROKE. ALL THE PIECES WERE REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15581738 ↗
  • InjuryJun 11, 2021Break; Break

    IT WAS REPORTED THAT THE DURING ANKLE ARTHROSCOPIC PROCEDURE, INSIDE THE PATIENT, THE ABRADER 3.5 MAG-MINI DISP. BLADE WAS BROKEN WHEN SHAVING THE TALUS SUBCHONDRAL BONE. THE PROCEDURE WAS FINISHED WITH A SMIT AND NEPHEW BACKUP DEVICE. 10 MINUTES DELAY, NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11987634 ↗
  • MalfunctionJan 22, 2019Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE PROCEDURE FOR RESECTION OF BONE SPURS, THE TIP OF THE BLADE WERE CAME OUT. THE PIECE WAS REMOVED. NO PATIENT INJURIES AND COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8267323 ↗
  • MalfunctionAug 15, 2018Material Fragmentation; Detachment of Device or Device Component; Device Fell

    IT WAS REPORTED THAT PIECES OF METAL WERE FLYING OFF OF THEM WHILE BEING USED IN A CASE. THE PIECES FELT OFF INTO THE PATIENT'S ANATOMY, ALL THE PIECES WERE REMOVED FROM THE PATIENT. BACK-UP DEVICE WAS AVAILABLE. AS PER CUSTOMER RESPONSE THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7787277 ↗

FDA clearance and classification

The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 72201519

Has the NA 72201519 been recalled?

No FDA recall record in our data names this exact device identifier (DI 03596010621641), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 72201519?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Aug 15, 2018 and Nov 17, 2025. In those reports, the most common specific device problems are “Break” (4), “Detachment of Device or Device Component” (2) and “Device Fell” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 72201519?

None of the reports are classified as deaths, 2 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 72201519 FDA approved?

Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.

Who makes the NA 72201519, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 72201519). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page