NA 72201519Recalls & reported problems
Smith & Nephew, Inc. · DYONICS 3.5 MM ABRADER BURRS, 7 CM LENGTH, AQUA, PACKAGED 6 PER BOX, STERILE · Class 1
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03596010621641), received between Aug 15, 2018 and Nov 17, 2025 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Detachment of Device or Device Component2
- Device Fell1
- Material Deformation1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 17, 2025Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT, DURING AN ARTHRODESIS OF THE ANKLE PROCEDURE, THE ABRADER BLADE WAS BROKEN. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT'S BODY. THE PROCEDURE WAS RESUMED AFTER A 10-MINUTE SURGICAL DELAY. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23569485 ↗MalfunctionOct 23, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TIP OF THE INNER BLADE OF THE ABRADER 3.5 WAS BROKEN. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT'S BODY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF 15 MINUTES. IT IS UNKNOWN IF THERE WAS A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20517177 ↗InjuryOct 11, 2022Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE INNER BLADE OF THE 3.5MM ABRADER BROKE. ALL THE PIECES WERE REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15581738 ↗InjuryJun 11, 2021Break; Break
IT WAS REPORTED THAT THE DURING ANKLE ARTHROSCOPIC PROCEDURE, INSIDE THE PATIENT, THE ABRADER 3.5 MAG-MINI DISP. BLADE WAS BROKEN WHEN SHAVING THE TALUS SUBCHONDRAL BONE. THE PROCEDURE WAS FINISHED WITH A SMIT AND NEPHEW BACKUP DEVICE. 10 MINUTES DELAY, NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11987634 ↗MalfunctionJan 22, 2019Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE PROCEDURE FOR RESECTION OF BONE SPURS, THE TIP OF THE BLADE WERE CAME OUT. THE PIECE WAS REMOVED. NO PATIENT INJURIES AND COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8267323 ↗MalfunctionAug 15, 2018Material Fragmentation; Detachment of Device or Device Component; Device Fell
IT WAS REPORTED THAT PIECES OF METAL WERE FLYING OFF OF THEM WHILE BEING USED IN A CASE. THE PIECES FELT OFF INTO THE PATIENT'S ANATOMY, ALL THE PIECES WERE REMOVED FROM THE PATIENT. BACK-UP DEVICE WAS AVAILABLE. AS PER CUSTOMER RESPONSE THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7787277 ↗
Other SAW, POWERED, AND ACCESSORIES devices
- DYONICS POWERMAX ELITE 72200616Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES4,290 reports
- NA 72200616SSmith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES486 reports
- NA 72201500Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES136 reports
- PLATINUM 72205321Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES118 reports
- Zimmer Biomet™ 89-8509-450-60Zimmer Surgical SA · Saw, Powered, And Accessories83 reports
- NA 72200617Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES77 reports
- NA 72202530Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES53 reports
- DYONICS INCISOR 72203013Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES46 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.