Skip to content

BIOSURE 72202273Recalls & reported problems

Smith & Nephew, Inc. · BIOSURE PK PEEK OPTIMA INTERFERENCE SCREW 9 X 30 MM · Class 2

All models of BIOSURE (68)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03596010622082), received between Oct 13, 2017 and Dec 29, 2023 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Adverse Event Without Identified Device or Use Problem1
  • Component Falling1
  • Difficult to Advance1
  • Failure to Fire1
  • Insufficient Device Problem Information1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Known Impact Or Consequence To Patient3
  • Cyst(s)1
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2023Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BIOSURE BREAK IN THE FIRST CONTACT WITH THE PATIENT BONE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THE BACK UP DEVICE WAS USED IN THE ORIGINAL BONE HOLE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18419464 ↗
  • MalfunctionNov 23, 2023Break; Failure to Fire; Break; Failure to Fire

    IT WAS REPORTED THE DURING AN ARTHROSCOPY, THE BIOSURE PK SCREW WAS STUCK IN THE SHAFT. THE PROCEDURE WAS COMPLETED WITH A DELAY GREATER THAN 30 MINUTES USING A S+N BACK UP DEVICE IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18202868 ↗
  • MalfunctionJun 7, 2023Break; Break

    IT WAS REPORTED THAT DURING AN ACL, A BIOSURE PK SCREW BROKE DURING INSERTION; THE BROKEN PIECES WERE REMOVED FROM THE JOINT. SURGERY RESUMED AFTER A NON-SIGNIFICANT DELAY WITH A BACK-UP DEVICE USED IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED. NO FURTHER DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17083665 ↗
  • InjuryNov 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON REGISTRY REVIEW 2021 (B)(6) ACL REGISTRY IMPLANT REPORT, AN UNKNOWN AMOUNT OF PATIENTS REQUIRED A REVISION SURGERY AFTER AN ACL PROCEDURE USING BIOSURE PK SCREW DEVICES. IT IS UNKNOWN HOW WERE THE EVENTS TREATED/RESOLVED. PATIENTS OUTCOME IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 15818288 ↗
  • MalfunctionDec 24, 2020Break; Break

    IT WAS REPORTED THAT DURING AN ACL SURGERY THE BIOSURE HAD A SLIGHT DIVET IN THE HEAD, IT SEEMS TO BE A DETACHED FRAGMENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACKUP DEVICE. NO PATIENT INJURIES OR COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11072245 ↗
  • MalfunctionDec 24, 2020Break; Break

    IT WAS REPORTED THAT DURING AN ACL SURGERY THE BIOSURE HAD A SLIGHT DIVET IN THE HEAD, IT SEEMS TO BE A DETACHED FRAGMENT BUT THE SURGEON DECIDED TO IMPLANT IT ANYWAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING THE SAME DEVICE. NO PATIENT INJURIES OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11072244 ↗
  • MalfunctionDec 24, 2020Break; Break

    IT WAS REPORTED THAT DURING AN ACL SURGERY THE BIOSURE HAD A SLIGHT DIVET IN THE HEAD WHICH SEEMS TO BE A DETACHED FRAGMENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACKUP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11072229 ↗
  • MalfunctionSep 14, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT, DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION, WHEN THE DOCTOR ATTEMPTED TO USE A "9X30MM BIOSURE PEEK SCREW" FOR FIXATION IN THE TIBIA, IT WAS NOTICED WHEN REMOVING THE DEVICE FROM THE PACKAGING, THAT THERE WAS A SMALL PART OF THE THREAD THAT APPEARED TO BE PEELING OFF. THEREFORE, DR. THOUGHT THIS MAY CAUSE DAMAGE TO THE INTEGRITY OF THE SCREW. IN CONSEQUENCE, THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING ANOTHER "9X30MM BIOSURE PEEK SCREW". NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10532455 ↗
  • MalfunctionMay 24, 2019Break

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE ITEM BROKE. THE PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE WITH NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8639856 ↗
  • InjuryNov 17, 2017Insufficient Device Problem Information

    IT WAS REPORTED THAT PATIENT HAD AN ACL RECONSTRUCTION. PATIENT CAME BACK COMPLAINING OF PAIN AROUND THE TIBIAL FIXATION SITE. MRI SHOWS LARGE CYSTS CONTAINING AROUND 40-50ML OF A JELLY LIKE SUBSTANCE. A SECOND SURGERY WAS REQUIRED YEARS AFTER THE INITIAL ACL TO REMOVE THE SCREWS.

    Patient problems: Cyst(s); Pain

    MAUDE report 7041554 ↗
  • MalfunctionOct 13, 2017Break; Component Falling; Difficult to Advance

    IT WAS REPORTED THAT ONE PATIENT WAS HAVING A STANDARD ACL RECONSTRUCTION USING HAMSTRING TENDONS SECURED WITH ENDOBUTTON (FEMUR) AND BIOSURE PK. THE ALLOGRAFT WAS SIZED AS 9MM AND A 9MM HOLE WAS DRILLED IN THE TIBIA. AFTER THE GRAFT WAS INSTITUTED A 1.2MM GUIDEWIRE WAS INSERTED AND THE CHOSEN 9 X 30MM BIOSURE PK SCREW WAS INSERTED. IT APPEARED NOT EASILY INSERTED AND HAD NO BITE IN THE TISSUE. ON REMOVAL OF THE SCREW IT WAS NOTED THAT THE TIP HAD BROKEN OFF. ANOTHER 9 X 30MM BIOSURE PK SCREW WAS CHOSEN AND IT WAS INSERTED WITHOUT ANY PROBLEMS. NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6949857 ↗

Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page