Mobi-C® Plug & Fit US MB3556Recalls & reported problems
LDR Spine Usa, Inc. · Total Cervical Disc Replacement System: Mobi-C Cervical Disc Prosthesis · Class 3 · Implant
All models of Mobi-C Plug & Fit US (13)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03662663000079), received between Jun 26, 2019 and Sep 3, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Disconnection2
- Device Handling Problem1
- Failure to Disconnect1
- Fracture1
- Loose or Intermittent Connection1
- Malposition of Device1
- Migration1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
- Peripheral Nervous Injury1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2025Fracture; Migration; Fracture; Migration
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED TO ADDRESS RADICULOPATHY CAUSED BY A POST-OP MOBI-C CORE FRACTURE AT C5/C6. THE SUPERIOR ENDPLATE HAD ALSO SHIFTED FORWARD TO WHERE IT WAS 2-3MM PROUD OF THE VERTEBRAL BODY. THE IMPLANT WAS REMOVED AND CONVERTED TO ACDF USING COMPETITIVE PRODUCT.
Patient problems: Peripheral Nervous Injury
MAUDE report 22956704 ↗MalfunctionJun 6, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A MOBI-C IMPLANT BOTTOM ENDPLATE DETACHED FROM THE PEEK WHILE IMPLANTING AND IMPACTING. ANOTHER MOBI-C IMPLANT WAS USED TO COMPLETE THE PROCEDURE AND THERE WAS NO REPORTED PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19477460 ↗MalfunctionJul 29, 2019Disconnection; Failure to Disconnect; Disconnection; Failure to Disconnect
IT WAS REPORTED THAT DURING SURGERY, UPON INSERTING THE IMPLANT, THERE WAS DIFFICULTY RELEASING THE IMPLANT FROM THE CARTRIDGE. THE PEEK CARTRIDGE WAS REMOVED AND IT WAS FOUND THAT THE TOP ENDPLATE WAS TWISTED. THE IMPLANT WAS REMOVED AND A NEW IMPLANT WAS USED IN ITS PLACE. THERE WAS NO PATIENT OR SURGICAL EFFECT REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8835336 ↗MalfunctionJul 16, 2019Disconnection; Malposition of Device; Detachment of Device or Device Component; Disconnection; Malposition of Device; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE PROCEDURE THE IMPLANT CAME OFF ITS RETENTION FEATURE DURING REPOSITIONING BY THE SURGEON THE IMPLANT WAS REPLACED WITH A ALTERNATE IMPLANT TO COMPLETE THE CASE. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8792502 ↗MalfunctionJun 26, 2019Loose or Intermittent Connection; Device Handling Problem; Loose or Intermittent Connection; Device Handling Problem
IT WAS REPORTED THAT DURING SURGERY, THE MOBI-C IMPLANT WAS LOOSE ON THE PEEK CARTRIDGE HOLDING IT. AS SUCH, A DIFFERENT IMPLANT WAS USED IN ITS PLACE. NO SURGICAL IMPACT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8735429 ↗
Other Prosthesis, Intervertebral Disc devices
- Prodisc C SK PDSM5Centinel Spine, LLC · Prosthesis, Intervertebral Disc2 recalls4 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall12 reports
- Prodisc C 09.820.055SCentinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall9 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall4 reports
- Prodisc C SK PDSXL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page