Skip to content

Mobi-C MB3795Recalls & reported problems

LDR Spine Usa, Inc. · MOBI-C IMPLANT M « STANDARD » 17x19 H5 US · Class 3 · Implant

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03662663002042), received between Feb 3, 2022 and Jul 26, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Migration1
  • Patient-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1
  • Osteolysis1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE INSERTION OF A MOBI-C IMPLANT, THE PEEK CARTRIDGE DISASSEMBLED AND RESULTED IN EARLY DEPLOYMENT OF THE DEVICE. RATHER THAN REBUILD AND REATTACH, THEY DECIDED TO USE ANOTHER IMPLANT TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19840842 ↗
  • InjuryOct 19, 2023Migration; Migration

    IT WAS REPORTED A REVISION SURGERY WAS PERFORMED TO ADDRESS A MOBI-C IMPLANT THAT MIGRATED POST-OPERATIVELY. ONE MONTH AFTER SURGERY, THE LOWER PLATE WAS PROLAPSED INTO THE SPINAL CANAL. THE REVISION WAS PERFORMED TO REPLACE THE MOBI-C IMPLANT WITH AN ACDF DEVICE EVEN THOUGH THE PATIENT WAS ASYMPTOMATIC.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17970076 ↗
  • MalfunctionJan 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING INSERTION OF A MOBI-C IMPLANT, THE STOP PENETRATED PAST THE VERTEBRAL BODY WHICH LED TO THE IMPLANT BEING TOO POSTERIOR. THE SURGEON TRIED TO PULL THE IMPLANT OUT AND THE IMPLANT DISENGAGED. THERE WAS A 10-MINUTE DELAY AND NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16102206 ↗
  • MalfunctionJun 13, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT A PATIENT DEVELOPED OSTEOLYSIS AFTER A LEVEL MOBI-C PROCEDURE. THE SURGEON IS CURRENTLY MONITORING THIS PATIENT AND NO PLANS FOR REVISION HAVE BEEN IDENTIFIED AT THIS TIME. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: Osteolysis

    MAUDE report 14676924 ↗
  • MalfunctionFeb 3, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE INFERIOR ENDPLATE DETACHED FROM THE CARTRIDGE, CAUSING THE MOBI-C IMPLANT TO ROTATE INTO THE IMPROPER PLACEMENT. THE SURGEON REMOVED AND REPLACED THE IMPLANT WITHOUT A DELAY OR PATIENT IMPACTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13445244 ↗

Other Prosthesis, Intervertebral Disc devices

All Prosthesis, Intervertebral Disc devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.