Aequalis Reversed II MWD148Recalls & reported problems
Tornier · Glenoid Sphere Extractor Screw · Class 1
All models of AEQUALIS REVERSED II (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03700386972010), received between Sep 27, 2016 and Sep 19, 2022 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation2
- Break1
- Fracture1
- Material Deformation1
- Material Twisted/Bent1
- Mechanical Jam1
Patient problems
- Patient Problem/Medical Problem2
- Device Embedded In Tissue or Plaque1
- Failure of Implant1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 19, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE TIP OF THE MWB148 INSTRUMENT BROKE OFF DURING USE. THE REPLACEMENT WAS AVAILABLE. ADDITIONALLY, A 5 MINUTE DELAY WAS REPORTED. SURGERY WAS COMPLETED SUCCESSFULLY. NO ADDITIONAL INFORMATION REPORTED.
Patient problems: Insufficient Information
MAUDE report 15451456 ↗MalfunctionMar 10, 2022Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING A RSA REVISION GLENOSPHERE REMOVAL THE PHYSICIAN NOTICED THAT THE GLENOSPHERE DID NOT DISASSOCIATE WHEN HE ATTEMPTED TO UNTHREAD THE INSTRUMENT. WHEN IT CAME OUT THE INNER PIECE WAS BENT AT APPROXIMATELY A 45 DEGREE ANGLE. HE CORRECTED IT WITH THE RATCHETING CLAMP FROM THE EXTRACTION SET. HE THEN ASKED IF WE HAD ANOTHER DEVICE, AFTER WHICH HE TRIED USING THE MWB216 AND MWD148 COMBINATION. THE GLENOSPHERE DISASSOCIATED, BUT AFTER THE PROCEDURE THOSE TWO PIECES WERE STUCK TOGETHER AND THE MWJ130 INNER PORTION WAS NO LONGER TRACKING PROPERLY. NO DELAYS, OR COMPLICATIONS RESULTED FROM THE INSTRUMENTS BECOMING DAMAGED. HOWEVER, THEY DID NEEDED TO BE REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13721419 ↗MalfunctionAug 19, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE SURGEON WAS ATTEMPTING TO TAKE THE SIZE 36 GLENOSPHERE OFF AND UPSIZE DUE TO STABILITY ISSUES THE PATIENT WAS HAVING. AFTER BACKING THE SET SCREW OUT AS FAR AS POSSIBLE THE SCREW CAME TO A POSITIVE STOP AND WOULD NO LONGER SPIN. THE SURGEON USED BOTH THE REVERSE II AND PERFORM REVERSE EXTRACTORS TO ATTEMPT TO REMOVE THE SPHERE BUT HE ENDED UP BREAKING THE FEET OF THE PERFORM REVERSE EXTRACTOR AND BENDING THE REVERSE II EXTRACTOR ATTEMPTING TO REMOVE THE SPHERE. THE SURGEON TRIED TO RE-ADVANCE THE SCREW DOWN JUST TO RESET EVERYTHING BUT WAS UNABLE TO DO SO AS THE SPHERE SCREW WAS STUCK AND WOULD NOT ADVANCE CLOCKWISE NOR COUNTERCLOCKWISE. WE NEEDED TO GO TO A 42 SPHERE TO LATERALIZE THE PATIENT FOR STABILITY BUT WE WERE UNABLE TO DO SO. THE 36 SPHERE IS STILL IN THE PATIENT.
Patient problems: Failure of Implant
MAUDE report 12344856 ↗MalfunctionFeb 8, 2019Material Fragmentation; Material Fragmentation
THE THREE FOLLOWING INSTRUMENTS ARE CROSS THREADED AND CANNOT BE CONTINUED TO USE. THE TIP OF THE GLENOID SPHERE EXTRACTOR SCREW (MWD148) BROKE OFF INSIDE THE WOUND AND IT TOOK 1.5 HOURS TO RETRIEVE IT. THE GLENOID SPHERE EXTRACTOR SCREW (MWD148) ALSO BECAME CROSS THREADED IN THE GLENOID SPHERE EXTRACTOR 42 MM (MWB218).
Patient problems: Patient Problem/Medical Problem
MAUDE report 8320595 ↗InjurySep 27, 2016Material Fragmentation; Material Deformation; Material Fragmentation; Material Deformation
THE TIP OF THE SCREW BROKE OFF AND THEN WAS NOT USED TO REMOVE THE SPHERE. WHEN THE SPHERE WAS EXTRACTED FROM THE PATIENT WE LOOKED INSIDE THE SPHERE FOR THE TIP AND COULD NOT FIND IT. THE SPHERE DID "FLY" OUT OF THE PATIENT, SO THE TIP COULD HAVE FLOWN AND LANDED SOMEWHERE IN THE ROOM. WE DID LOOK ON THE FLOOR AS WELL. GIVEN THIS, THERE IS STILL A CHANCE THAT THE TIP IS INSIDE THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque; Patient Problem/Medical Problem
MAUDE report 5979278 ↗
Other EXTRACTOR devices
- ATTUNE 2545-00-138Depuy (Ireland) · EXTRACTOR392 reports
- PINNACLE 2217-50-001Depuy (Ireland) · EXTRACTOR243 reports
- RECLAIM 2975-00-635Depuy (Ireland) · EXTRACTOR157 reports
- ATTUNE 2545-00-060Depuy (Ireland) · EXTRACTOR153 reports
- ESC PINNACLE 2218-00-010Depuy (Ireland) · EXTRACTOR151 reports
- NA 2125-00-600Depuy Orthopaedics, Inc. · EXTRACTOR130 reports
- P.F.C. 86-5226Depuy (Ireland) · EXTRACTOR80 reports
- NA 03.010.000Synthes GmbH · EXTRACTOR74 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page