POWERLED TM ARD568415010ARecalls & reported problems
Maquet Sas · PWD55+SF HD K3 · Class 2
All models of POWERLED TM (280)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03700712402532), received between Jun 28, 2023 and Feb 5, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K070442); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 8, 2023Open, Classified
A potential for a light system to fall in the operating room.
Z-0763-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗
Most reported problems
Device problems
- Peeled/Delaminated3
- Detachment of Device or Device Component2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2026Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. IT WAS STATED THE PAINT WAS CHIPPING ON FORK. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24267513 ↗MalfunctionJan 15, 2026Peeled/Delaminated
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24078003 ↗MalfunctionJan 12, 2026Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. IT WAS STATED THE PAINT WAS CHIPPING FROM FORK DUE TO IMPACT. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24035351 ↗MalfunctionSep 19, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ON 14TH SEPTEMBER, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED 500. IT WAS STATED THE COVER FROM THE JOINTS HAD SPRUNG OUT AND HIT A PATIENT'S WRIST IN THE MIDDLE OF SURGERY. THERE WAS NO MEDICAL INTERVENTION REQUIRED, NO INJURY OR PERMANENT DAMAGE TO THE BODY STRUCTURE REPORTED, NO DELAYS IN SURGERY. SPRING ARM COVER CHANNEL WAS WORN DUE TO INCORRECT POSITION OF THE DUST COVER. DAMAGE OF THE SPRING ARM COVERS INDICATED THAT THE DUST COVERS WERE NOT FITTED CORRECTLY AND THIS MIGHT BE A REASON WHY THEY COULD POP OUT AND FALL OFF. IT HAS BEEN BROUGHT TO GETINGE TECHNICIAN ATTENTION THAT NIGHT SHIFT HOSPITAL ENGINEERS HAVE BEEN PUTTING THEM BACK IN THEMSELVES PRIOR TO THE INCIDENT. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY IN CASE OF REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17771652 ↗MalfunctionJun 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ON 15TH JUNE, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. BASED ON PHOTOGRAPHIC EVIDENCE THE COVERS WERE MISSING FROM SPRING ARM. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17219555 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070442. One submission often covers many models, so check the product and lot details in each record.
- Z-3224-2026 · Aug 7, 2026 · Open, ClassifiedPowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD568474114A, 2. ARD567501286;
- Z-1990-2015 · May 29, 2015 · TerminatedThe PowerLED / X-TEN / VOLISTA Surgical Light Systems are surgical light systems intended to be used to provide visible illumination of the surgical area or the patient. Ceiling su
Other Light, surgical, ceiling mounted devices
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls6 reports
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls2 reports
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls2 reports
- POWERLED TM ARD568474114AMaquet Sas · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record