POWERLED TM ARD568446710ARecalls & reported problems
Maquet Sas · PWD73+DF HD R K3 · Class 2
All models of POWERLED TM (280)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03700712402648), received between Sep 13, 2023 and Apr 28, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K070442); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 8, 2023Open, Classified
A potential for a light system to fall in the operating room.
Z-0763-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Peeled/Delaminated3
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2025Peeled/Delaminated; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWER LED. IT WAS STATED THE PAINT WAS MISSING ON THE DEVICE. THE DESIGNATED COMPLAINT UNIT EMPLOYEE CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THAT PAINT WAS PEELING FROM FORK AND HEADLIGHT COVER. IT WAS ESTABLISHED THE EQUIPMENT HAS COLLIDED MANY TIMES. THERE WAS NO INJURY, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PAINT PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21917141 ↗MalfunctionApr 25, 2025Peeled/Delaminated; Peeled/Delaminated
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. IT WAS STATED THE PAINT WAS MISSING ON THE DEVICE. THE DESIGNATED COMPLAINT UNIT EMPLOYEE CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE THAT PAINT WAS PEELING FROM FORK AND HEADLIGHT COVER. IT WAS ESTABLISHED THE EQUIPMENT HAS COLLIDED MANY TIMES. THERE WAS NO INJURY, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PAINT PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21909292 ↗MalfunctionApr 26, 2024Crack; Peeled/Delaminated; Detachment of Device or Device Component; Crack; Peeled/Delaminated; Detachment of Device or Device Component
ON 23RD APRIL, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. AS IT WAS STATED, VARIOUS PAINT CHIPS WERE PRESENT ON THE DEVICE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION IN CASE OF REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19189296 ↗MalfunctionNov 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ON 7TH SEPTEMBER, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. IT WAS STATED THE SCREWS FROM THE SPRING ARM WERE MISSING. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18217057 ↗MalfunctionSep 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - POWERLED. DURING PREVENTIVE MAINTENANCE IT WAS FOUND THE SCREWS WERE MISSING FROM SPRING ARM. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17734702 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070442. One submission often covers many models, so check the product and lot details in each record.
- Z-3224-2026 · Aug 7, 2026 · Open, ClassifiedPowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD568474114A, 2. ARD567501286;
- Z-1990-2015 · May 29, 2015 · TerminatedThe PowerLED / X-TEN / VOLISTA Surgical Light Systems are surgical light systems intended to be used to provide visible illumination of the surgical area or the patient. Ceiling su
Other Light, surgical, ceiling mounted devices
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls6 reports
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls2 reports
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls2 reports
- POWERLED TM ARD568474114AMaquet Sas · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
- HarmonyAIR A-Series Lighting SystemSteris Corporation · Light, surgical, ceiling mounted2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page