Skip to content

Humeral Cup Stability 104-1009Recalls & reported problems

Fx Solutions · HUMERAL CUP STABILITY PE/TA6V Ø40/+9 · Class 2 · Implant

All models of Humeral Cup Stability (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03701037309469), received between Apr 5, 2022 and Dec 12, 2025 — classified as 6 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Appropriate Device Problem Term/Code Not Available1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Joint Dislocation3
  • Inadequate Osseointegration1
  • Medical device site infection1
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 12, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    REVISION OCCURRED APPROXIMATELY 8 DAYS AFTER THE PRIOR REVISION SURGERY, WHICH OCCURRED ON (B)(6) 2025 AND WAS RECORDED IN PER 25-805. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO USER ERROR. ACCORDING TO THE DISTRIBUTOR, THE SURGEON DID NOT FOLLOW THE SURGICAL TECHNIQUE: HE DID NOT LINE UP THE HUMERAL CUP IMPACTOR AND IMPACTED FROM AN OBLIQUE ANGLE, RESULTING IN THE DISSOCIATION OF THE HUMERAL SPACER FROM THE STEM. A 40 MM + 9 HUMERAL STABILITY CUP AND 9 MM SPACER WERE EXPLANTED. THESE ITEMS WERE REPLACED WITH IDENTICAL COMPONENTS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23791417 ↗
  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OCCURRED APPROXIMATELY 14 DAYS AFTER THE PRIMARY SURGERY, WHICH OCCURRED ON (B)(6) 2025. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO DISLOCATION OF UNKNOWN CAUSE. A 40 MM + 9 HUMERAL STABILITY CUP WAS EXPLANTED. THIS ITEM WAS REPLACED WITH A 40 MM + 6 HUMERAL STABILITY CUP AND A 9 MM SPACER.

    Patient problems: Joint Dislocation

    MAUDE report 23314056 ↗
  • InjuryJun 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OCCURRED APPROXIMATELY 3 WEEKS AFTER THE PRIMARY SURGERY, WHICH OCCURRED ON 22-MAY-2025. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO DISLOCATION OF UNKNOWN CAUSE. A 40 + 9 MM HUMERAL STABILITY CUP WAS EXPLANTED. THIS WAS REPLACED WITH A 40 MM + 3 HUMERAL STABILITY CUP, AND A 9 MM SPACER.

    Patient problems: Joint Dislocation

    MAUDE report 22346503 ↗
  • InjuryJun 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OCCURRED APPROXIMATELY 4 MONTHS AFTER THE PRIMARY SURGERY WHICH OCCURRED ON (B)(6) 2024. NO FALL OR OTHER TRAUMA REPORTED. REVISION OCCURRED DUE TO DISLOCATION OF UNKNOWN CAUSE. 40 MM + 9 HUMERAL STABILITY CUP EXPLANTED AND REPLACED WITH A 40 MM + 9 HUMERAL STABILITY CUP AND 2 OF 9 MM SPACERS.

    Patient problems: Joint Dislocation

    MAUDE report 19637596 ↗
  • InjuryApr 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO AN INFECTION ON (B)(6) 2024. THE IMPLANTATION DATE WAS ON (B)(6) 2024. A CUP AND A GLENOSPHERE WERE EXPLANTED. A CUP AND A GLENOSPHERE WERE IMPLANTED.

    Patient problems: Medical device site infection

    MAUDE report 19211821 ↗
  • InjuryApr 5, 2022Appropriate Device Problem Term/Code Not Available

    THE PATIENT WAS REVISED ON (B)(6) 2022 DUE TO INFECTION. APPROXIMATELY 27 DAYS AFTER THE FIRST SURGERY. THE SURGEON EXPLANTED 1 ECCENTRIC GLENOSPHERE, 2 LOCKING SCREW, 2 STANDARD SCREW, 2 CORTICAL SCREW ANS 1 HUMERAL CUP STABILITY Ø40+9. THE SURGEON IMPLANTED 1 HELEMET HEAD AND 1 OFFSET TAPER ADAPTER.

    Patient problems: Unspecified Infection

    MAUDE report 14010466 ↗

Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices

All Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices →