MENIX DUO Meniscal suture system, 2 anchorsRecalls & reported problems

S.B.M · MENIX® DUO system consists of a single-use launcher which deploys a nonabsorbable implant composed of two anchors connected by a suture. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03760304680526), received between Mar 10, 2026 and Apr 29, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Twisted/Bent1

Patient problems

  • Insufficient Information2
  • Foreign Body In Patient1
  • Rupture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Material Twisted/Bent

    (B)(4). INCIDENT OCCURED IN FRANCE. COMPLAINT DESCRIPTION: OBSERVATION IN 1 PHOTO: DEFORMED NEEDLE TIPS - MESSAGE: "SHARP TIP NOT REMOVED".

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 25030342 ↗
  • MalfunctionMar 10, 2026Break

    (B)(4) INCIDENT OCCURED IN ARGENTINA. COMPLAINT DESCRIPTION: EVENT DESCRIPTION: "THE PEEK IMPLANTS DID NOT DEPLOY PROPERLY. IN THE FIRST CASE, THE REPAIR APPEARED TO BE WELL FIXED; HOWEVER, WHEN TYING THE KNOT, THE SUTURE BROKE. A SECOND DEVICE WAS USED, BUT THE IMPLANTS FAILED TO PROPERLY DEPLOY BEHIND THE CAPSULE. A THIRD DEVICE WAS THEN USED, AND THE SAME ISSUE OCCURRED. A TOTAL OF THREE DEVICES WERE INVOLVED: TWO UNITS FROM BATCH 254186 AND ONE UNIT FROM BATCH 254178. ALL THREE ISSUES OCCURRED DURING THE SAME SURGERY. THE SURGERY DURATION WAS APPROXIMATELY 40 MINUTES. THE PHYSICIAN CONSIDERED IT A DIFFICULT CASE TO REPAIR. THE JOINT WAS DEFINED AS CHALLENGING DUE TO THE LIMITED WORKING SPACE AND THE COMPLEXITY OF THE LESION. AS THE MENISCAL REPAIR COULD NOT BE SUCCESSFULLY COMPLETED USING SUTURES, THE PROCEDURE WAS FINALIZED WITH A PARTIAL MENISCECTOMY".

    Patient problems: Insufficient Information; Rupture

    MAUDE report 24556006 ↗