Accu-Chek Spirit Cartridge 05206073001Recalls & reported problems

Roche Diabetes Care GmbH · Accu-Chek Spirit 3.15ml Cart. Sys. 25pc · Class 2

All models of Accu-Chek Spirit Cartridge (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is an Accu-Chek Spirit insulin cartridge used with an insulin pump. The 30 reports mostly describe insulin leaking out of the cartridge — into the pump's cartridge compartment, at the piston rod connection, or between the cartridge and the infusion tubing — sometimes across several cartridges from the same package, and in some cases blood sugar rose as a result. No deaths and no injuries were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 04015630882908), received between Sep 5, 2019 and Jul 23, 2024 — classified as 30 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (30 reports).

1 other recall cover devices cleared under the same FDA submission (K042887); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 8 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak30

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • Hyperglycemia8
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE. THE INSULIN LEAK OCCURRED WITH 8 CARTRIDGES FROM THE SAME PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19808034 ↗
  • MalfunctionJul 11, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19725417 ↗
  • MalfunctionJul 11, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE. THE INSULIN LEAK APPEARED AT THE CONNECTION WITH THE PISTON ROD. THIS ISSUE OCCURRED WITH TWO INSULIN CARTRIDGES FROM THE SAME PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19717875 ↗
  • MalfunctionJul 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN BETWEEN THE CONNECTION OF THE CARTRIDGE AND THE INFUSION TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19704623 ↗
  • MalfunctionJul 8, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE RESULTING IN ELEVATED BLOOD GLUCOSE LEVELS.

    Patient problems: Hyperglycemia

    MAUDE report 19689420 ↗
  • MalfunctionJul 3, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE RESULTING IN ELEVATED BLOOD GLUCOSE LEVELS.

    Patient problems: Hyperglycemia

    MAUDE report 19665501 ↗
  • MalfunctionJul 3, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN BETWEEN THE ADAPTER AND INFUSION TUBING RESULTING IN ELEVATED BLOOD GLUCOSE LEVELS. THE PATIENT'S BLOOD GLUCOSE LEVEL INCREASED TO 300 MG/DL. THE INSULIN LEAK OCCURRED WITH CARTRIDGES FROM THE SAME PACKAGE. IT IS UNKNOWN HOW MANY CARTRIDGES WERE DEFECTIVE. THE PATIENT THINKS THE INSULIN CARTRIDGE WAS THE CAUSE OF THE LEAKAGE.

    Patient problems: Hyperglycemia

    MAUDE report 19665471 ↗
  • MalfunctionJul 3, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE RESULTING IN ELEVATED BLOOD GLUCOSE LEVELS. THE ISSUE HAS OCCURRED FOR SEVERAL WEEKS. IT IS UNKNOWN HOW MANY CARTRIDGES WERE DEFECTIVE.

    Patient problems: Hyperglycemia

    MAUDE report 19665467 ↗
  • MalfunctionJun 7, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT INSULIN WAS LEAKING OUT OF THE STOPPER AT THE BACK OF THE INSULIN CARTRIDGE. THE INSULIN LEAK OCCURRED WITH 5 CARTRIDGES FROM THE SAME LOT NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19485162 ↗
  • MalfunctionJun 7, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE RESULTING IN ELEVATED BLOOD GLUCOSE LEVELS.

    Patient problems: Hyperglycemia

    MAUDE report 19483163 ↗
  • MalfunctionJun 4, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE RESULTING IN ELEVATED BLOOD GLUCOSE LEVELS. THE ISSUE OCCURRED WITH TWO CARTRIDGES.

    Patient problems: Hyperglycemia

    MAUDE report 19459845 ↗
  • MalfunctionApr 25, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19180395 ↗
  • MalfunctionApr 16, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN DUE TO A CRACK ON THE CARTRIDGE 2 TO 3 WEEKS AGO. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19116063 ↗
  • MalfunctionApr 8, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19067000 ↗
  • MalfunctionApr 8, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19066885 ↗
  • MalfunctionApr 4, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19043250 ↗
  • MalfunctionMar 26, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18977141 ↗
  • MalfunctionMar 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18960548 ↗
  • MalfunctionJan 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18569095 ↗
  • MalfunctionDec 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18350670 ↗
  • MalfunctionDec 6, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE. THIS ISSUE OCCURRED WITH FIVE CARTRIDGES OF THE SAME LOT NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18274637 ↗
  • MalfunctionNov 15, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE REPEATEDLY LEADING TO ELEVATED BLOOD GLUCOSE LEVELS.

    Patient problems: Hyperglycemia

    MAUDE report 18140034 ↗
  • MalfunctionOct 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE. THE ISSUE ALSO OCCURRED WITH CARTRIDGE FROM ANOTHER PACK WITH THE SAME LOT NUMBER.

    Patient problems: Hyperglycemia

    MAUDE report 17956060 ↗
  • MalfunctionOct 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17930646 ↗
  • MalfunctionOct 12, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INSULIN CARTRIDGE LEAKED INSULIN INTO THE CARTRIDGE COMPARTMENT OF THE INFUSION DEVICE. THE CALLER ALLEGED THAT THIS OCCURRED WITH TWO CARTRIDGES FROM THE SAME BOX/LOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17923132 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042887. One submission often covers many models, so check the product and lot details in each record.

  • Z-1734-2015 · Apr 22, 2015 · Terminated
    AccuChek Spirit insulin infusion pump.

Other Pump, Infusion, Insulin devices

All Pump, Infusion, Insulin devices →