Skip to content

Cholesterol Gen.2 04718917190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630922062), received between Jul 1, 2024 and Apr 27, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

1 other recall cover devices cleared under the same FDA submission (K031824); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 27, 2026Non Reproducible Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH CHOLESTEROL GEN.2 ON A COBAS C 111 ANALYZER. THE SAMPLE INITIALLY RESULTED IN A CHOLESTEROL VALUE OF 408.19 MG/DL. THE SAMPLE WAS REPEATED, RESULTING IN CHOLESTEROL VALUES OF 203.71 MG/DL AND 201.57 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25001605 ↗
  • MalfunctionJul 1, 2024High Test Results; High Test Results

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH CHOLESTEROL GEN.2 (CHOL2) ASSAY ON A COBAS C111 ANALYZER. INITIAL RESULT: 385 MG/DL. THE PHYSICIAN QUESTIONED THE INITIAL RESULT AND REQUESTED A REPEAT OF THE ORIGINAL SAMPLE. ON 10-JUN-2024: REPEAT RESULT: 213 MG/DL. THE REPEAT RESULT WAS CONSIDERED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19644150 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K031824. One submission often covers many models, so check the product and lot details in each record.

  • Z-1921-2015 · May 7, 2015 · Terminated
    Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems,

Other ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL devices

All ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page