Cholesterol Gen.2 05168538190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630924967), received between Sep 29, 2022 and Sep 19, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K031824); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2
  • No Device Output1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Low Test Results; Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE CHOLESTEROL GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 8000 C702 MODULE. THE INITIAL RESULT WAS 0.25 MMOL/L. THE SAMPLE WAS REPEATED AND THE RESULT WAS 3.9 MMOL/L. THE SAMPLE WAS TESTED 10 TIMES AND THE AVERAGE RESULT WAS 3.973 MMOL/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23097623 ↗
  • MalfunctionFeb 26, 2024Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE TRIGL TRIGLYCERIDES AND CHOL2 CHOLESTEROL GEN.2 RESULTS FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 8000 COBAS C 701 MODULE. THIS MDR IS FOR THE CHOL2 ASSAY. PLEASE REFER TO MDR WITH A1. PATIENT IDENTIFIER (B)(6) FOR THE TRIGL ASSAY. THE REPORTED COMPLAINED THAT THE PATIENT SAMPLE HAD CHOL2 AND TRIGL RESULTS CLOSE TO ZERO BUT THE HDL-C AND LDL-C TOTALED TO OVER 100 MG/DL (HDL = 52 MG/DL; LDL = 53 MG/DL). THE PHYSICIAN DID NOT BELIEVE THE RESULTS BECAUSE THEY COULD NOT BE CORRECT AND WERE INCOMPATIBLE WITH LIFE. THE REPORTER STATED THAT THERE IS AN INTERFERENCE DUE TO THE THERAPY OF PARACETAMOL INTOXICATION INDICATED IN THE FOUR ASSAYS MENTIONED AND ALSO ANOTHER INTERFERENCE FOR THE CHOL2 AND TRIGL ASSAYS. ON (B)(6) 2024: THE INITIAL CHOL2 RESULT WAS <10 MG/DL. THE FIRST REPEAT RESULT WAS <10 MG/DL. ON (B)(6) 2024: THE SECOND REPEAT RESULT WAS <10 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18776411 ↗
  • MalfunctionSep 29, 2022No Device Output; No Device Output

    THE INITIAL REPORTER RECEIVED QUESTIONABLE CHOL2 CHOLESTEROL GEN.2, ELECSYS TOTAL PSA IMMUNOASSAY, AND IRON2 IRON GEN.2 RESULTS FROM 1 PATIENT ON A COBAS 8000 COBAS C 701 MODULE. THE CUSTOMER WAS UNABLE TO OBTAIN PLAUSIBLE RESULTS FROM 1 PATIENT'S SAMPLES. THE DATES OF TESTING WERE (B)(6) 2022. ALL SAMPLES GAVE A RESULT OF "???" FOR ALL TESTS EXCEPT FOR IRON ON (B)(6) 2022. THE RESULT WAS REQUESTED BUT NOT PROVIDED. IT IS SUSPECTED THAT THERE MAY BE AN INTERFERING FACTOR IN THE PATIENT SAMPLE. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE ANALYZER SERIAL NUMBER IS (B)(4). THIS MEDWATCH WILL COVER CHOLESTEROL. REFER TO MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE PSA RESULTS AND MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE IRON RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15503359 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K031824. One submission often covers many models, so check the product and lot details in each record.

  • Z-1921-2015 · May 7, 2015 · Terminated
    Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems,