Cholesterol Gen.2 05168538190Recalls & reported problems
Roche Diagnostics GmbH · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630924967), received between Sep 29, 2022 and Sep 19, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K031824); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results2
- No Device Output1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 19, 2025Low Test Results; Low Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CHOLESTEROL GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 8000 C702 MODULE. THE INITIAL RESULT WAS 0.25 MMOL/L. THE SAMPLE WAS REPEATED AND THE RESULT WAS 3.9 MMOL/L. THE SAMPLE WAS TESTED 10 TIMES AND THE AVERAGE RESULT WAS 3.973 MMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23097623 ↗MalfunctionFeb 26, 2024Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE TRIGL TRIGLYCERIDES AND CHOL2 CHOLESTEROL GEN.2 RESULTS FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 8000 COBAS C 701 MODULE. THIS MDR IS FOR THE CHOL2 ASSAY. PLEASE REFER TO MDR WITH A1. PATIENT IDENTIFIER (B)(6) FOR THE TRIGL ASSAY. THE REPORTED COMPLAINED THAT THE PATIENT SAMPLE HAD CHOL2 AND TRIGL RESULTS CLOSE TO ZERO BUT THE HDL-C AND LDL-C TOTALED TO OVER 100 MG/DL (HDL = 52 MG/DL; LDL = 53 MG/DL). THE PHYSICIAN DID NOT BELIEVE THE RESULTS BECAUSE THEY COULD NOT BE CORRECT AND WERE INCOMPATIBLE WITH LIFE. THE REPORTER STATED THAT THERE IS AN INTERFERENCE DUE TO THE THERAPY OF PARACETAMOL INTOXICATION INDICATED IN THE FOUR ASSAYS MENTIONED AND ALSO ANOTHER INTERFERENCE FOR THE CHOL2 AND TRIGL ASSAYS. ON (B)(6) 2024: THE INITIAL CHOL2 RESULT WAS <10 MG/DL. THE FIRST REPEAT RESULT WAS <10 MG/DL. ON (B)(6) 2024: THE SECOND REPEAT RESULT WAS <10 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18776411 ↗MalfunctionSep 29, 2022No Device Output; No Device Output
THE INITIAL REPORTER RECEIVED QUESTIONABLE CHOL2 CHOLESTEROL GEN.2, ELECSYS TOTAL PSA IMMUNOASSAY, AND IRON2 IRON GEN.2 RESULTS FROM 1 PATIENT ON A COBAS 8000 COBAS C 701 MODULE. THE CUSTOMER WAS UNABLE TO OBTAIN PLAUSIBLE RESULTS FROM 1 PATIENT'S SAMPLES. THE DATES OF TESTING WERE (B)(6) 2022. ALL SAMPLES GAVE A RESULT OF "???" FOR ALL TESTS EXCEPT FOR IRON ON (B)(6) 2022. THE RESULT WAS REQUESTED BUT NOT PROVIDED. IT IS SUSPECTED THAT THERE MAY BE AN INTERFERING FACTOR IN THE PATIENT SAMPLE. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE ANALYZER SERIAL NUMBER IS (B)(4). THIS MEDWATCH WILL COVER CHOLESTEROL. REFER TO MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE PSA RESULTS AND MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE IRON RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15503359 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K031824. One submission often covers many models, so check the product and lot details in each record.
- Z-1921-2015 · May 7, 2015 · TerminatedCholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems,
Other ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL devices
- PTS Panels® CHOL+Glu Test Strips 1765Polymer Technology Systems, Inc. · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL2 recalls3 reports
- CardioChek 5700411Henry Schein, Inc. · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL1 recall5 reports
- Alere Cholestech LDX® 10-987Alere San Diego, Inc. · Enzymatic Esterase--Oxidase, Cholesterol1 recall0 reports
- Synchron 467825Beckman Coulter, Inc. · Enzymatic Esterase--Oxidase, Cholesterol1 recall0 reports
- Cholesterol 08057443190Roche Diagnostics GmbH · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL21 reports
- Cholesterol Gen.2 3039773190Roche Diagnostics GmbH · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL11 reports
- Cholesterol Gen.2 04718917190Roche Diagnostics GmbH · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL2 reports
- Dimension Vista® Flex® reagent cartridge CHOL 10445143Siemens Healthcare Diagnostics Inc. · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page