Bilirubin Direct Gen.2 (Doumas standardization) 05589061190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630927531), received between May 27, 2025 and Jan 30, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results3
- Non Reproducible Results3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 30, 2026High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE BILIRUBIN DIRECT GEN.2 RESULTS FROM THE COBAS 6000 C 501 ANALYZER. THE INITIAL RESULT WAS 0.5 MG/DL. THE REPEAT RESULT WAS 0.15 MG/DL AND WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24223662 ↗MalfunctionAug 25, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE BILIRUBIN DIRECT GEN.2 RESULTS FOR 2 PATIENT SAMPLES ON A COBAS 4000 C311 STAND ALONE SYSTEM. ON (B)(6) 2025, SAMPLE 1 HAD AN INITIAL DIRECT BILIRUBIN RESULT OF 1.605 MG/DL. THE SAMPLE WAS REPEATED, AND THE RESULT WAS 1.639 MG/DL. THE DOCTOR QUESTIONED THE RESULT. THE REPEAT RESULT FROM AN EXTERNAL LABORATORY THAT USES AN ABBOTT ANALYZER WAS 0.46 MG/DL. THIS REPEAT RESULT WAS DEEMED CORRECT. ON (B)(6) 2025, SAMPLE 2 HAD AN INITIAL DIRECT BILIRUBIN RESULT OF 1.969 MG/DL. THE REPEAT RESULT FROM THE EXTERNAL LABORATORY WAS 0.56 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22883103 ↗MalfunctionMay 27, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE BILIRUBIN DIRECT GEN.2 RESULT FROM THE COBAS 4000 C311 STAND ALONE SYSTEM. THE INITIAL RESULT WAS 1.025 MG/DL. THE PEDIATRICIAN REQUESTED A NEW BLOOD SAMPLE TO BE ANALYZED AT ANOTHER LABORATORY USING AN ABBOTT ANALYZER. THE RESULT FOR THE NEW SAMPLE WAS 0.50 MG/DL TWICE. ON (B)(6) 2025, THE INITIAL SAMPLE WAS RETESTED ON THE COBAS C311 ANALYZER, AND THE RESULT WAS 1.158 MG/DL. THE SAME INITIAL SAMPLE WAS SENT TO ANOTHER LABORATORY THAT USES AN ABBOTT ANALYZER, AND THE RESULT WAS 0.52 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22087186 ↗
Other DIAZO COLORIMETRY, BILIRUBIN devices
- Dimension® Flex® reagent cartridge TBI 10444957Siemens Healthcare Diagnostics Inc. · DIAZO COLORIMETRY, BILIRUBIN2 recalls4 reports
- Dimension Vista® Flex® reagent cartridge TBIL 10445146Siemens Healthcare Diagnostics Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall7 reports
- VITROS 8383051Ortho-Clinical Diagnostics, Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall5 reports
- VITROS 1612365Ortho-Clinical Diagnostics, Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall1 report
- ARCHITECT 6L45-21Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN46 reports
- ARCHITECT 6L45-22Abbott GmbH · DIAZO COLORIMETRY, BILIRUBIN30 reports
- ARCHITECT 6L45-41Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN18 reports
- Bilirubin Direct Gen.2 08056951190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN17 reports