Bilirubin Total Gen3 5795419190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Bilirubin Total Gen3 (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04015630928903), received between Jan 28, 2019 and Jul 24, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K131544); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • False Negative Result1
  • Non Reproducible Results1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 24, 2024Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings

    WE RECEIVED AN ALLEGATION ABOUT A SUSPECTED DRUG INTERFERENCE FOR 1 PATIENT SERUM SAMPLE TESTED WITH BIL-T GEN.3 (BILT3) ASSAY ON A COBAS 8000 COBAS C701 MODULE. THE CUSTOMER MENTIONED THAT THE SAMPLE WAS ABNORMALLY DARK IN COLOR AND THEY WERE CONCERNED THAT THE DARK COLOR OF THE SERUM MIGHT BE INTERFERING WITH THE TEST. THE BILT3 RESULT WAS 10.2. THE UNIT OF MEASUREMENT WAS NOT PROVIDED. THE RESULT WAS REPORTED OUTSIDE THE LABORATORY AND THE SAMPLE WAS NOT REPEATED. THE CUSTOMER ALSO MENTIONED THAT THE SERUM INDEX (SI) WAS NORMAL: HEMOLYSIS: 20. ICTERUS: 4. LIPEMIA: 49.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19820342 ↗
  • MalfunctionMay 2, 2024False Negative Result; Non Reproducible Results; False Negative Result; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE BILIRUBIN TOTAL GEN. 3 RESULTS FROM THE COBAS 8000 COBAS C 701 MODULE. THE INITIAL RESULT WAS 2.55 MG/L AND THE REPEAT RESULT ON (B)(6) 2024 WAS 16.63 MG/L. THE REPEAT RESULT WAS BELIEVED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19231591 ↗
  • MalfunctionMar 8, 2024Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE TOTAL BILIRUBIN (BIL-T GEN 3) RESULTS FROM THE COBAS 8000 C702 MODULE. THE RESULT WAS 9.9 MG/DL ON ANOTHER ANALYZER AND THE RESULT ON THIS ANALYZER WAS 0.5 MG/DL. THE CUSTOMER ALSO PROVIDED A RESULT OF 10.4 MG/DL. THE CUSTOMER REPORTED A RESULT OF 10.2 MG/DL THAT WAS BELIEVED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18865253 ↗
  • MalfunctionApr 24, 2023Low Test Results; Low Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED QUESTIONABLE RESULTS FOR ONE PATIENT SAMPLE TESTED WITH BILT3 BILIRUBIN TOTAL GEN.3 ON COBAS 8000 C (701) MODULE SERIAL NUMBER (B)(6) , A COBAS 6000 C (501) MODULE (SERIAL NUMBER UNKNOWN), AND AN UNKNOWN ROCHE ANALYZER (SERIAL NUMBER UNKNOWN). THE PATIENT SAMPLE WAS NOTED TO BE BROWN IN COLOR AND THE REPORTER ALLEGES IT MAY CONTAIN AN INTERFERING FACTOR AGAINST A COMPONENT OF THE TOTAL BILIRUBIN ASSAY. THE REPORTER ALSO NOTED THAT THE SAME ISSUE HAS OCCURRED WITH THIS PATIENT BEFORE. THE PLASMA TUBE OF THE PATIENT SAMPLE INITIALLY RESULTED IN A TOTAL BILIRUBIN VALUE OF 0.7 MG/DL WHEN TESTED ON THE C 701 ANALYZER. THE QUESTIONABLE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE PLASMA AND SERUM TUBES OF THE PATIENT SAMPLE WERE REPEATED ON THE C 501 ANALYZER AND ALL TOTAL BILIRUBIN RESULTS WERE 0.7 MG/DL. THE SERUM SAMPLE WAS THEN SENT TO A SECOND SITE WHERE IT WAS TESTED ON AN UNKNOWN ROCHE ANALYZER, RESULTING IN A TOTAL BILIRUBIN VALUE OF 0.7 MG/DL. THE SERUM SAMPLE WAS SENT TO A THIRD SITE WHERE IT WAS TESTED ON A BECKMAN COULTER AU680 ANALYZER, RESULTING IN A TOTAL BILIRUBIN VALUE OF 3.6 MG/DL. THE SERUM SAMPLE WAS ALSO SENT TO A FOURT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16797144 ↗
  • MalfunctionJan 28, 2019Low Test Results; Low Test Results

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS. ERRONEOUS LOW RESULTS WERE GENERATED BY THE COBAS 8000 C 702 MODULE. THE EVENTS INVOLVED A TOTAL OF 1 PATIENT WITH THE FOLLOWING: ONE ERRONEOUS RESULT FOR BILT3 BILIRUBIN TOTAL GEN.3. THE PATIENT'S AGE WAS REQUESTED, BUT NOT PROVIDED. THE PATIENT'S WEIGHT WAS REQUESTED, BUT NOT PROVIDED. THERE PATIENT'S GENDER WAS REQUESTED, BUT NOT PROVIDED. THE PATIENT'S RACE WAS REQUESTED, BUT NOT PROVIDED. THE PATIENT'S ETHNICITY WAS REQUESTED, BUT NOT PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8281240 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131544. One submission often covers many models, so check the product and lot details in each record.

  • Z-0153-2020 · Sep 11, 2019 · Terminated
    Cobas c Bilirubin Total Gen.3 Catalog Number: 05795419190 - Product Usage: In vitro test for the quantitative determination of the total bilirubin in serum and plasma of adults an
  • Z-1145-2019 · Dec 17, 2018 · Terminated
    Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160 ***added as of 5/10/19***05795397190, 05795648190 Product Usage: In vitro te

Other DIAZO COLORIMETRY, BILIRUBIN devices

All DIAZO COLORIMETRY, BILIRUBIN devices →