Creatine Kinase 7190794190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
All models of Creatine Kinase (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04015630940806), received between Jul 20, 2022 and Sep 22, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “High Test Results” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Aug 15, 2024Class IIOpen, Classified
Creatine Kinase (CK) Assay Abnormal calibrations on the cobas c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.
Z-3187-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- High Test Results4
- Low Test Results4
- Non Reproducible Results4
- Incorrect, Inadequate or Imprecise Result or Readings3
- Improper or Incorrect Procedure or Method1
- Incorrect Measurement1
- Material Integrity Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Incorrect, Inadequate or Imprecise Result or Readings
There was an allegation of questionable Creatine Kinase results for 12 patient samples on a cobas 4000 c311 stand alone system. As a representative example, three patient results are provided. Patient 1's initial result was 680 U/L.The repeat result from a different c311 was 66 U/L.Patient 2's initial result was 1008 U/L.The repeat result from a different c311 was 97 U/L.Patient 3's initial result was 1890 U/L. The repeat result from a different c311 was 197 U/L.The repeat results were deemed to be correct.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50856428 ↗MalfunctionJul 16, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH CREATINE KINASE (CK) ASSAY ON A COBAS 6000 C501 MODULE. INITIAL RESULT: 78 U/L. THE PHYSICIAN QUESTIONED THE INITIAL RESULT AS IT DID NOT MATCH THE PATIENT'S PREVIOUS RESULT OF 132,220 U/L. THE PHYSICIAN EXPECTED THE INITIAL RESULT TO BE >100,000 U/L. THE SAMPLE WAS THEN REPEATED AT MULTIPLE DILUTIONS (BOTH WITH AUTO-DILUTION BY THE INSTRUMENT AND MANUAL DILUTION), AND THE REPEAT RESULTS WERE >100,000 U/L. ONE OF THE REPEAT RESULTS WAS 163,985 U/L (ACCOMPANIED BY A DATA FLAG). THE FINAL REPEAT RESULT WAS 181,100 U/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22521887 ↗MalfunctionNov 8, 2024High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CKMBL CREATINE KINASE-MB RESULTS FOR 1 PATIENT ON A COBAS 6000 C501 MODULE. SAMPLE 1: THE INITIAL RESULT WAS 8712 U/L. THIS RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE SAMPLE WAS REPEATED WITH A 1:10 DILUTION ON ANOTHER MODULE AND THE RESULT WAS 877 U/L. THE SAMPLE WAS REPEATED ON ANOTHER MODULE AND THE RESULTS WERE: 7249 U/L WITH A DATA FLAG THAT INVALIDATED THE RESULT, 7348 U/L WITH A DATA FLAG THAT INVALIDATED THE RESULT, 7124 U/L WITH A DATA FLAG THAT INVALIDATED THE RESULT, 7108 U/L WITH A DATA FLAG THAT INVALIDATED THE RESULT, 8280 U/L WITH A DATA FLAG THAT INVALIDATED THE RESULT, AND 8374 U/L WITH A DATA FLAG THAT INVALIDATED THE RESULT. SAMPLE 2: A NEW SAMPLE WAS TESTED ON ANOTHER MODULE AND THE RESULT WAS 54 U/L THE SAMPLE WAS REPEATED WITH A 1:10 DILUTION ON ANOTHER MODULE AND THE RESULT WAS 7 WITH A DATA FLAG THAT INVALIDATED THE RESULT. SAMPLE 3: A NEW SAMPLE WAS TESTED ON ANOTHER MODULE AND THE RESULT WAS 61 U/L. THE RESULTS OF "55-60 U/L" WERE CONSIDERED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20645570 ↗MalfunctionOct 4, 2024Use of Device Problem; Improper or Incorrect Procedure or Method; Non Reproducible Results; Use of Device Problem; Improper or Incorrect Procedure or Method; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CK CREATINE KINASE RESULTS FOR 1 PATIENT ON A COBAS 6000 C (501) MODULE. THE CUSTOMER RECEIVED TWO SERUM TUBES FROM THE SAME PATIENT COLLECTION FOR TWO DIFFERENT TEST PANELS THAT BOTH INCLUDED CK. THE DIFFERENCE IN THE CK VALUES BETWEEN THE TUBES COLLECTED AT THE SAME TIME PROMPTED THEM TO QUESTION THE RESULTS. THE CK RESULT FROM THE FIRST PATIENT SAMPLE WAS 141 U/L. THE CK RESULT FROM THE SECOND PATIENT SAMPLE WAS 70 U/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20370521 ↗MalfunctionSep 5, 2024Incorrect, Inadequate or Imprecise Result or Readings; Material Integrity Problem; Incorrect, Inadequate or Imprecise Result or Readings; Material Integrity Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE CKMBL CREATINE KINASE-MB RESULTS FROM THE COBAS INTEGRA 400 PLUS W/O ISE ANALYZER. THE INITIAL RESULT WAS 6655 UL/L. THE REPEAT RESULT WAS 6569 UL/L AND AFTER CALIBRATION, THE REPEAT RESULT WAS 82 UI/L. ON (B)(6) 2024, THE REPEAT RESULT WAS 7360 UI/L. AFTER ANOTHER CALIBRATION AND CLEANING OF THE INNER AND OUTER WATER TANK, THE REPEAT RESULT WAS 5784 UI/L. FOR THIS TEST, THE CUSTOMER USED A DIFFERENT EDTA SAMPLE FROM THE SAME VENIPUNCTURE. THE RESULT OF 82 UI/L WAS REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20148726 ↗MalfunctionSep 12, 2023High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CREATININE KINASE (CK) RESULTS FOR ONE PATIENT FROM TWO COBAS 501 ANALYZERS. DIFFERENT SAMPLES FROM THE SAME CHILD, TAKEN A FEW HOURS APART, HAD DIFFERENCES IN CK VALUES OF THOUSANDS OF IU. AT 16.00 PM, SAMPLE 1 TESTED ON ANALYZER SERIAL NUMBER (B)(6) HAD A RESULT OF GREATER THAN 5000 IU. THE REPEAT RESULT WITH A DILUTION WAS 5800 IU. THE SAME SAMPLE WAS TESTED WITH A DILUTION ON ANALYZER SERIAL NUMBER (B)(6) AND THE RESULT WAS 6000 IU. SAMPLE 2 WAS DRAWN AT 22.00 PM AND WAS TESTED ON ANALYZER SERIAL NUMBER (B)(6) WITH A RESULT OF 70 IU. THIS SAMPLE WAS THEN TESTED ON ANALYZER SERIAL NUMBER (B)(6) WITH A RESULT OF 68 IU. ON (B)(6) 2023, THE CUSTOMER RETESTED THE SAME SAMPLES. SAMPLE 1 RESULT FROM ANALYZER SERIAL NUMBER (B)(6) WITH A DILUTION WAS 5894 IU. SAMPLE 1 RESULT FROM ANALYZER SERIAL NUMBER (B)(6) WITH A DILUTION WAS 6000 IU. THE RESULT FOR SAMPLE 2 USING ANALYZER SERIAL NUMBER (B)(6) WAS 70 IU. THE RESULT FOR SAMPLE 2 USING ANALYZER SERIAL NUMBER (B)(6) WAS 68 IU. THE RESULT FROM A THIRD SAMPLE DRAWN FROM THE PATIENT AND TESTED ON ANALYZER SERIAL NUMBER (B)(6) WAS 67 IU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17731157 ↗MalfunctionJun 6, 2023High Test Results; High Test Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE ISE INDIRECT NA, K AND CI FOR GEN.2 RESULTS FROM AN UNSPECIFIED NUMBER OF PATIENT SAMPLES TESTED ON THE COBAS 6000 C 501 (UL) V. THE INITIAL RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER STATED THAT ON (B)(6) 2023 AT AROUND NOON, THEY HAD DIFFICULTY CALIBRATING THE NA ASSAY. THEY THEN CHANGED THE "REAGENTS" AND WERE ABLE TO PERFORM A SUCCESSFUL CALIBRATION AND QC. THEY RAN PATIENT SAMPLES THAT NIGHT WHEN A DOCTOR CALLED QUESTIONING A NA RESULT THAT WAS TOO HIGH IN COMPARISON WITH THE POINT OF CARE (POC) RESULT. THE REPORTER THEN RERAN THE PATIENT SAMPLE ON THE OTHER MODULE AND A RESULT DISCREPANCY WAS NOTED. THEY THEN RERAN OTHER PATIENT SAMPLES ON THEIR TWO OTHER C501 MODULES AND NOTED MULTIPLE DISCREPANT NA, AND POSSIBLY A FEW K AND CL RESULTS. THE REPORTER WAS ABLE TO PROVIDE TWO EXAMPLES OF DISCREPANT NA RESULTS. PATIENT SAMPLE 1: THE INITIAL RESULT FROM THE MODULE WAS 148 MMOL/L. THE REPEAT RESULT FROM ONE OF THE TWO OTHER C501 MODULES WAS 140 MMOL/L. PATIENT SAMPLE 1: THE INITIAL RESULT FROM THE MODULE WAS 147 MMOL/L. THE REPEAT RESULT FROM ONE OF THE TWO OTHER C 501 MODULES WAS 139 MMOL/L. THE REPORTER STATED THAT TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17073105 ↗MalfunctionApr 25, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S PLASMA SAMPLE TESTED WITH CREATINE KINASE (CK) ASSAY ON A COBAS 6000 C501 (UL) V SERIAL NUMBER (B)(6). THE INITIAL RESULT WAS 19 U/L. THE REPEAT RESULTS WERE 8076 U/L WITH A DATA FLAG AND 9785 U/L. THE RESULTS WERE OBTAINED FROM A SINGLE SAMPLE. THE QUESTIONABLE RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE PHYSICIAN QUESTIONED THE RESULT AND THE SAMPLE WAS REPEATED. THE REPEAT RESULT DEEMED TO BE THE CORRECT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16809455 ↗MalfunctionApr 12, 2023Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE CREATININE KINASE (CK) RESULT FROM THE COBAS 6000 C501 MODULE SERIAL NUMBER (B)(4). IN THE AFTERNOON AT 18:04, THE INITIAL RESULT FOR SAMPLE 1 WAS 733 U/L (WITH NO DATA FLAG) AND WAS REPORTED OUTSIDE OF THE LABORATORY. LATER IN THE EVENING AROUND 22:00, SAMPLE 2 WAS DRAWN FROM THE PATIENT AND THE CK RESULT WAS 2505 U/L WITH A DATA FLAG. WITH A 1:50 DILUTION, THE RESULT WAS 34941 U/L. SAMPLE 1 WAS THEN REPEATED AT 22:45 WITH A 1:50 DILUTION AND THE RESULT WAS 37337 U/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16726441 ↗MalfunctionDec 12, 2022Incorrect Measurement; Low Test Results; Incorrect Measurement; Low Test Results
THE INITIAL REPORTER STATED THEY RECEIVED QUESTIONABLE RESULTS FOR ONE PATIENT SAMPLE TESTED WITH CK CREATINE KINASE ON TWO COBAS 6000 C (501) MODULE ANALYZERS. NO INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER SUSPECTED THE SAMPLE CONTAINS AN INTERFERING FACTOR AGAINST A COMPONENT OF THE CK ASSAY. THE SAMPLE WAS TESTED ON THE FIRST C 501 ANALYZER, RESULTING IN A VALUE OF < 6 U/L WITH A DATA FLAG. THE SAMPLE WAS TESTED TWICE ON THE SECOND C 501 ANALYZER, EACH TIME RESULTING IN A VALUE OF < 6 U/L WITH A DATA FLAG. A SERIAL NUMBER OF (B)(4) WAS PROVIDED FOR ONE OF THE TWO C 501 ANALYZERS. IT IS NOT KNOWN IF THIS WAS THE FIRST ANALYZER OR THE SECOND ANALYZER USED FOR TESTING. THE SERIAL NUMBER OF THE OTHER C 501 ANALYZER WAS REQUESTED, BUT NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15959879 ↗MalfunctionJul 20, 2022High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CK CREATINE KINASE RESULTS FROM COBAS 6000 C 501 MODULE SERIAL NUMBER (B)(4) DUE TO INTERFERENCE. SAMPLE 1 INITIAL RESULT WAS 500 U/L. THE REPEAT RESULT WITH A DILUTION WAS >57000 U/L. AS THIS DID NOT MATCH THE CLINICAL PICTURE FOR THE PATIENT, A NEW SAMPLE WAS DRAWN. SAMPLE 2 INITIAL RESULT WAS 7 U/L WITH A DATA FLAG. THE REPEAT RESULT WITH A REDUCED SAMPLE VOLUME WAS 47072 U/L WITH A DATA FLAG. ON ANOTHER ANALYZER, THE INITIAL RESULT WAS 1339 U/L WITH A DATA FLAG. THE REPEAT RESULT WITH A REDUCED SAMPLE VOLUME WAS 46386 U/L WITH A DATA FLAG. ADDITIONAL RESULTS FROM (B)(6) 2022 OF 1222 U/L WITH A DATA FLAG AND 46255 U/L WERE PROVIDED. FROM (B)(6) 2022, RESULTS OF 2971 U/L WITH A DATA FLAG AND 32794 U/L WERE PROVIDED. IT IS UNCLEAR FROM WHICH SAMPLE THESE RESULTS WERE GENERATED. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15070073 ↗
Other DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES devices
- VITROS 8001133Ortho-Clinical Diagnostics, Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES1 recall10 reports
- VITROS 8058232Ortho-Clinical Diagnostics, Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES1 recall9 reports
- ADVIA Centaur® Systems, ACS:180® Systems CKMB Calibrator 10311570Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES1 recall0 reports
- Atellica IM CKMB Calibrator 10995532Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES1 recall0 reports
- Atellica IM CKMB 10995531Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES2 reports
- Atellica IM CKMB 10995530Siemens Healthcare Diagnostics Inc. · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES2 reports
- Creatine Kinase 08057460190Roche Diagnostics GmbH · DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page