Creatine Kinase 08057460190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Creatine Kinase (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336121122), received between Apr 11, 2026 and Apr 11, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

1 other recall cover devices cleared under the same FDA submission (K160570); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE CREATINE KINASE RESULTS FOR 3 PATIENT SAMPLES ON A COBAS C 503 ANALYTICAL UNIT. THE CUSTOMER REPEATED PREVIOUS PATIENT SAMPLES ON A NEW REAGENT LOT AS THEY WERE HAVING ISSUES WITH QC RECOVERING HIGH. SAMPLE 1 HAD AN INITIAL RESULT OF 79.0 U/L. THE SAMPLE WAS REPEATED ON ANOTHER ANALYZER AND THE RESULT WAS 67.5 U/L. ON (B)(6) 2026, THE SAMPLE WAS REPEATED ON THE INITIAL ANALYZER WITH A NEW REAGENT LOT AND THE RESULT WAS 59.5 U/L. SAMPLE 2 HAD AN INITIAL RESULT OF 62.0 U/L. THE SAMPLE WAS REPEATED ON ANOTHER ANALYZER AND THE RESULT WAS 48.0 U/L. ON (B)(6) 2026, THE SAMPLE WAS REPEATED ON THE INITIAL ANALYZER WITH A NEW REAGENT LOT AND THE RESULT WAS 47.6 U/L. SAMPLE 3 HAD AN INITIAL RESULT OF 15075.0 U/L. ON (B)(6) 2026, THE SAMPLE WAS REPEATED WITH A NEW REAGENT LOT AND THE RESULT WAS 9683 U/L. THE REPEAT RESULTS FROM THE NEW REAGENT LOT WERE DEEMED CORRECT. CLARIFICATION ON WHETHER THE INITIAL RESULTS WERE PRODUCED ON (B)(6)2026 WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24851932 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160570. One submission often covers many models, so check the product and lot details in each record.

  • Z-3187-2024 · Aug 15, 2024 · Open, Classified
    Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative determination of creatine kinase (CK) in human serum and plasma. Cata

Other DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES devices