VITROS 8001133Recalls & reported problems

Ortho-Clinical Diagnostics, Inc. · VITROS Chemistry Products CKMB Slides · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10758750004201), received between May 18, 2023 and Feb 19, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (10 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 6 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 11, 2020Open, Classified

    VITROS Chemistry Products CKMB Slides Potential for Falsely Elevated Results Due to CK-MM Interference

    Z-0770-2021 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Low Test Results10

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2026Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM A VITROS ISOENZYME PERFORMANCE VERIFIER (IPV) PROCESSED USING VITROS CHEMISTRY PRODUCTS CKMB SLIDES LOT 4922-0272-8370 ON A VITROS 350 CHEMISTRY SYSTEM. VITROS IPV I LOT Y2755 VITROS CKMB RESULTS OF 9.3, 13.8, 11.1, 7.8, 12.7, 10.7, 12.1, 9.7, 10.6, 12.3, 6.3, 15.0, 11.1, 15.1, 8.7, 8.7 AND 15.2 U/L VS AN EXPECTED RESULT OF 21.5 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER-THAN-EXPECTED RESULTS WERE FROM A NON-PATIENT CONTROL FLUID AND WERE NOT REPORTED FROM THE LABORATORY. THE CUSTOMER STATED THAT NO PATIENT SAMPLE RESULTS HAD BEEN PROCESSED DURING THE TIMEFRAME OF THE EVENT. THERE HAVE BEEN NO ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24392383 ↗
  • MalfunctionSep 19, 2024Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM FOUR DIFFERENT CORRELATION PATIENT SAMPLES USING VITROS CKMB SLIDE LOT 4917-0268-4441 ON A VITROS 4600 CHEMISTRY SYSTEM, WHEN COMPARED TO RESULTS OBTAINED USING AN ALTERNATE VITROS CKMB SLIDE LOT. CORRELATION PATIENT SAMPLE 2 RESULT OF 16, 14 AND 15 AND U/L VS. THE EXPECTED RESULT OF 31 U/L CORRELATION PATIENT SAMPLE 3 RESULTS OF 14 AND 11 AND 11 U/L VS. THE EXPECTED RESULT OF 30 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER-THAN-EXPECTED VITROS CKMB RESULTS WERE OBTAINED DURING A PATIENT SAMPLE CORRELATION AND WERE NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER SIX OF SIX MDRS FOR THIS EVENT. SIX 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SIX DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20258737 ↗
  • MalfunctionSep 19, 2024Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM FOUR DIFFERENT CORRELATION PATIENT SAMPLES USING VITROS CKMB SLIDE LOT 4917-0268-4441 ON A VITROS 4600 CHEMISTRY SYSTEM, WHEN COMPARED TO RESULTS OBTAINED USING AN ALTERNATE VITROS CKMB SLIDE LOT. CORRELATION PATIENT SAMPLE 2 RESULT OF 16, 14 AND 15 AND U/L VS. THE EXPECTED RESULT OF 31 U/L CORRELATION PATIENT SAMPLE 3 RESULTS OF 14 AND 11 AND 11 U/L VS. THE EXPECTED RESULT OF 30 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER-THAN-EXPECTED VITROS CKMB RESULTS WERE OBTAINED DURING A PATIENT SAMPLE CORRELATION AND WERE NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER FIVE OF SIX MDRS FOR THIS EVENT. SIX 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SIX DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20257306 ↗
  • MalfunctionSep 19, 2024Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM FOUR DIFFERENT CORRELATION PATIENT SAMPLES USING VITROS CKMB SLIDE LOT 4917-0268-4441 ON A VITROS 4600 CHEMISTRY SYSTEM, WHEN COMPARED TO RESULTS OBTAINED USING AN ALTERNATE VITROS CKMB SLIDE LOT. CORRELATION PATIENT SAMPLE 2 RESULT OF 16, 14 AND 15 AND U/L VS. THE EXPECTED RESULT OF 31 U/L CORRELATION PATIENT SAMPLE 3 RESULTS OF 14 AND 11 AND 11 U/L VS. THE EXPECTED RESULT OF 30 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER-THAN-EXPECTED VITROS CKMB RESULTS WERE OBTAINED DURING A PATIENT SAMPLE CORRELATION AND WERE NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER FOUR OF SIX MDRS FOR THIS EVENT. SIX 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SIX DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20257094 ↗
  • MalfunctionSep 19, 2024Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM FOUR DIFFERENT CORRELATION PATIENT SAMPLES USING VITROS CKMB SLIDE LOT 4917-0268-4441 ON A VITROS 4600 CHEMISTRY SYSTEM, WHEN COMPARED TO RESULTS OBTAINED USING AN ALTERNATE VITROS CKMB SLIDE LOT. CORRELATION PATIENT SAMPLE 2 RESULT OF 16, 14 AND 15 AND U/L VS. THE EXPECTED RESULT OF 31 U/L CORRELATION PATIENT SAMPLE 3 RESULTS OF 14 AND 11 AND 11 U/L VS. THE EXPECTED RESULT OF 30 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER-THAN-EXPECTED VITROS CKMB RESULTS WERE OBTAINED DURING A PATIENT SAMPLE CORRELATION AND WERE NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER THREE OF SIX MDRS FOR THIS EVENT. SIX 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SIX DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20256033 ↗
  • MalfunctionSep 19, 2024Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM FOUR DIFFERENT CORRELATION PATIENT SAMPLES USING VITROS CKMB SLIDE LOT 4917-0268-4441 ON A VITROS 4600 CHEMISTRY SYSTEM, WHEN COMPARED TO RESULTS OBTAINED USING AN ALTERNATE VITROS CKMB SLIDE LOT. CORRELATION PATIENT SAMPLE 2 RESULT OF 16, 14 AND 15 AND U/L VS. THE EXPECTED RESULT OF 31 U/L CORRELATION PATIENT SAMPLE 3 RESULTS OF 14 AND 11 AND 11 U/L VS. THE EXPECTED RESULT OF 30 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER-THAN-EXPECTED VITROS CKMB RESULTS WERE OBTAINED DURING A PATIENT SAMPLE CORRELATION AND WERE NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF SIX MDRS FOR THIS EVENT. SIX 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SIX DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20256005 ↗
  • MalfunctionSep 19, 2024Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM FOUR DIFFERENT CORRELATION PATIENT SAMPLES USING VITROS CKMB SLIDE LOT 4917-0268-4441 ON A VITROS 4600 CHEMISTRY SYSTEM, WHEN COMPARED TO RESULTS OBTAINED USING AN ALTERNATE VITROS CKMB SLIDE LOT. CORRELATION PATIENT SAMPLE 2 RESULT OF 16, 14 AND 15 AND U/L VS. THE EXPECTED RESULT OF 31 U/L CORRELATION PATIENT SAMPLE 3 RESULTS OF 14 AND 11 AND 11 U/L VS. THE EXPECTED RESULT OF 30 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER-THAN-EXPECTED VITROS CKMB RESULTS WERE OBTAINED DURING A PATIENT SAMPLE CORRELATION AND WERE NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF SIX MDRS FOR THIS EVENT. SIX 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SIX DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20255979 ↗
  • MalfunctionMay 18, 2023Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM A RIQAS EXTERNAL QUALITY CONTROL FLUID AND FROM A VITROS ISOENZYME PERFORMANCE VERIFIER (PV) FLUID WHEN RUNNING VITROS CKMB LOT 4903-0248-4599 ON A VITROS 5600 INTEGRATED SYSTEM. VITROS CKMB RIQAS CYCLE 44 SAMPLE 1 RESULT OF 10.0 U/L VERSUS THE EXPECTED RESULT OF 19.098 U/L VITROS ISOENZYME PVI LOT M8776 RESULTS OF 12.82 AND 8.40 U/L VERSUS THE EXPECTED RESULT OF 20.5 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE CUSTOMER MADE NO INDICATION THAT PATIENT RESULTS FROM THE TIME FRAME OF THE EVENT WERE IN QUESTION AND THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER THREE OF THREE MDR¿S FOR THIS EVENT. THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16952112 ↗
  • MalfunctionMay 18, 2023Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM A RIQAS EXTERNAL QUALITY CONTROL FLUID AND FROM A VITROS ISOENZYME PERFORMANCE VERIFIER (PV) FLUID WHEN RUNNING VITROS CKMB LOT 4903-0248-4599 ON A VITROS 5600 INTEGRATED SYSTEM. VITROS CKMB RIQAS CYCLE 44 SAMPLE 1 RESULT OF 10.0 U/L VERSUS THE EXPECTED RESULT OF 19.098 U/L VITROS ISOENZYME PVI LOT M8776 RESULTS OF 12.82 AND 8.40 U/L VERSUS THE EXPECTED RESULT OF 20.5 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE CUSTOMER MADE NO INDICATION THAT PATIENT RESULTS FROM THE TIME FRAME OF THE EVENT WERE IN QUESTION AND THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF THREE MDR¿S FOR THIS EVENT. THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16952073 ↗
  • MalfunctionMay 18, 2023Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS CKMB RESULTS WERE OBTAINED FROM A RIQAS EXTERNAL QUALITY CONTROL FLUID AND FROM A VITROS ISOENZYME PERFORMANCE VERIFIER (PV) FLUID WHEN RUNNING VITROS CKMB LOT 4903-0248-4599 ON A VITROS 5600 INTEGRATED SYSTEM. VITROS CKMB RIQAS CYCLE 44 SAMPLE 1 RESULT OF 10.0 U/L VERSUS THE EXPECTED RESULT OF 19.098 U/L VITROS ISOENZYME PVI LOT M8776 RESULTS OF 12.82 AND 8.40 U/L VERSUS THE EXPECTED RESULT OF 20.5 U/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE CUSTOMER MADE NO INDICATION THAT PATIENT RESULTS FROM THE TIME FRAME OF THE EVENT WERE IN QUESTION AND THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF THREE MDR¿S FOR THIS EVENT. THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16952053 ↗

Other DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES devices

All DIFFERENTIAL RATE KINETIC METHOD, CPK OR ISOENZYMES devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.