GIBECK IPN041364Recalls & reported problems
Teleflex Incorporated · HUMID VENT FILTER PEDI · Class 1
All models of GIBECK (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04026704347841), received between Oct 6, 2023 and Mar 28, 2024 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Defective Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device2
- Obstruction of Flow2
- Blocked Connection1
- Device Sensing Problem1
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 28, 2024Unable to Obtain Readings; Device Sensing Problem; Unable to Obtain Readings; Device Sensing Problem
IT WAS REPORTED THAT: "HEALTHCARE CLINICIAN USED A HUMID-VENT PEDI FILTER ON A PATIENT IN THE DAY CLINIC AND WAS UNABLE TO OBTAIN A CO2 READING ON THE CAPNOGRAPH. CLINICAL CONSEQUENCES: THE FILTER WAS REPLACED AND THE DR WAS THE ABLE TO OBTAIN THE DESIRED READING. ".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18993202 ↗MalfunctionMar 11, 2024Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT : "THE DEVICE WAS DISCOVERED TO HAVE A PARTIALLY BLOOKED PORT. THERE WAS NO CO2 TRACE WHEN EXPECTED, ISOLATED FAULT TO HME, NOTICED SAMPLING PORT LUMEN WAS BLOCKED." ADDITIONAL INFORMATION: THERE WAS NO HEALTH CONSEQUENCES TO THE PATIENT AND NO DESATURATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18874122 ↗MalfunctionFeb 27, 2024Obstruction of Flow; Blocked Connection; Obstruction of Flow; Blocked Connection
IT WAS REPORTED THAT " ABSENT CAPNOGRAPHY TRACE DESPITE BEING ABLE TO VENTILATE THE PATIENT. ON INSPECTION, NO OPENING BETWEEN THE CAPNOGRAPHY PORT AND THE FILTER ".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18786898 ↗InjuryOct 9, 2023Defective Device; Defective Device
IT WAS REPORTED THAT, "BLOCKAGE IN FILTER PREVENTING VENTILATION OF PEDIATRIC PATIENT. SEEMS TO HAVE PLASTIC INSIDE IT CAUSING BLOCKAGE. PATIENT WAS UNABLE TO BE VENTILATED. NO CARBON DIOXIDE READING WAS OBTAINED. ALL ASPECTS OF CIRCUIT WERE EXAMINED. THE FILTER WAS FOUND TO BE THE CAUSE". AT THE TIME OF THIS REPORT THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF ADDITIONAL INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information
MAUDE report 17897320 ↗MalfunctionOct 6, 2023Defective Device; Defective Device
THE REPORT STATES THAT "THE FILTER HAS A PORTAL WHICH IS BLOCKED WITH A PIECE OF PLASTIC THUS OCCLUDING IT AND NOT ALLOWING THE PATIENT'S ANESTHESIA CIRCUIT TO WORK. IT WAS NOTICED PRIOR TO USE ON A PATIENT". NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17887539 ↗
Other CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE) devices
- HUDSON RCI IPN042780Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall3 reports
- HUDSON RCI IPN042781Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall1 report
- DAR 353S19046Covidien LP · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall1 report
- DAR 354/5876Covidien LP · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)25 reports
- AVANOS 93816Avanos Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)24 reports
- HUDSON RCI IPN040673Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)11 reports
- Portex 450952-NLIcu Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)8 reports
- Portex C45031682D-NLIcu Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.