Skip to content

HUDSON RCI IPN044418Recalls & reported problems

Teleflex Incorporated · TRACH-VENT + · Class 1

All models of HUDSON RCI (152)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04026704348176), received between May 16, 2022 and Sep 4, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem2
  • Connection Problem1
  • Contamination /Decontamination Problem1
  • Delivered as Unsterile Product1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Discomfort1
  • Low Oxygen Saturation1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 4, 2024Delivered as Unsterile Product; Contamination /Decontamination Problem; Delivered as Unsterile Product; Contamination /Decontamination Problem

    IT WAS REPORTED THAT"(B)(6) 2024, BEFORE THE NURSE PREPARED TO REPLACE THE PATIENT'S ARTIFICIAL NOSE, SHE FOUND WATER DROPLETS INSIDE THE UNOPENED PACKAGE. THIS WAS DISCARDED IMMEDIATELY AND REPLACED WITH A NEW INTACT PRODUCT TO COMPLETE THE PROCEDURE WITHOUT ADVERSE EFFECT TO THE PATIENT.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20134230 ↗
  • MalfunctionNov 9, 2022Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT "WE ARE USING THESE TRACH-VENT SINCE JULY 2022. WE ARE FACING DIFFICULTY CONNECTING TO THE TRACHEOSTOMY TUBES. ARTIFICIAL NOSES DO NOT CONNECT TO IT (THE ORIFICE IS TOO SMALL)." FURTHER INFORMATION STATES, "THERE WAS NO CONSEQUENCE FOR THE PATIENT, NO DESATURATION. THE DOCTOR DID A DIY SYSTEM TO MAKE SURE THE TRACH-VENT WAS CONNECTING TO THE TUBE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15764230 ↗
  • MalfunctionOct 28, 2022Compatibility Problem; Compatibility Problem

    THE REPORT STATES THAT THE "INCIDENT OCCURRED IN THE RESUSCITATION ROOM ON (B)(6) 2022. THE CONNECTION OF THE TRACK-VENT ON THE CANNULA IS DIFFICULT, THE GAUGES DO NOT MATCH. THIS MAKES THE PROCEDURE PAINFUL FOR THE PATIENT. [THE CLINICIANS] OBSERVED DURING COUGHING EFFORTS THAT THE TRACK-VENT DISCONNECTED FROM THE CANNULA WHICH LED TO EPISODES OF DESATURATION IN THE PATIENT." FURTHER INFORMATION STATES THAT "THE MEDICAL INTERVENTION WAS TO GIVE [OXYGEN] BACK TO THE PATIENT AND MONITOR." THE PATIENT STATUS IS REPORTED AS "FINE."

    Patient problems: Low Oxygen Saturation; Pain

    MAUDE report 15691344 ↗
  • MalfunctionJun 6, 2022Connection Problem; Connection Problem

    IT WAS REPORTED THAT: "THE FILTER WAS CONNECTED TO A TRACHEOTOMY TUBE. THE FILTER DID NOT FIT INTO THE CANNULA. THE ATTACHMENT WAS LOOSE, THE FILTER WAS HANGING DOWN AND THUS OXYGENATION WAS IMPOSSIBLE. IT WAS THEN NECESSARY TO CHANGE THE CANNULA TO CONNECT THE FILTER". ADDITIONAL INFORMATION RECEIVED STATED THAT THE PATIENT DID NOT EXPERIENCE ANY DESATURATION. THE DEVICE WAS CHANGED AND THERE WAS NO CONSEQUENCE FOR THE PATIENT. NO PATIENT HARM OR INJURY REPORTED. PATIENT CONDITION REPORTED AS "FINE".

    Patient problems: Discomfort

    MAUDE report 14605523 ↗
  • InjuryMay 16, 2022Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED "THERE WAS AN EXTRA VALVE INCLUDED IN THE HUMIDIFIER RENDERING THE PRODUCT USELESS AND OBSTRUCTING OUR PATIENTS AIRWAY". DEVICE WAS USED IN THE HOME ENVIRONMENT. ADDITIONAL INFORMATION RECEIVED STATED THAT THE PATIENT HAS AN EXISTING TRACHY AND THE ISSUE WAS CORRECTED BY REMOVING THE DEFECTIVE DEVICE AND REPLACING IT. PATIENT CONDITION REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14407114 ↗

Other CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE) devices

All CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page